Message ID: 34     Entry time: Thu Mar 24 17:10:33 2022
Author: Carsten 
Category: DAQ 
Subject: DAQ / MBS VUPROM TDC buffer errors 
VUPROM TDC buffer errors potentially due to long TDC common Stop gate (>1ms from div 200).
Signal piped through coinc. unit gicv es shorter sginal (~200ns). This signal used as common stop seems to have cured the TDC buffer errors..