Some Default Settings for the XUV
Detector. The copper
mirror should always move fully in.
Possible positions: Between 0mm and 147mm.
Background was
very high for anything higher than 100mm
in the last
experiment. 50mm was optimal. This should
be tested again
with repaired and modified shields.
Good default values for the XUV detector
parameters have
been proven before to be:
Cathode: -5V
Movable shield: 0V
Fixed shield: 0V
Main coil: 8A
Secondary coil: -2.4A
(negative values set in hardware, only
positive software
inputs are possible) |