Author: Rodolfo 
Category: Laser 
Subject: Quanta Ray - Water leak in the power supply 
After fixing the broken water connector in the laser head a new water leak appeared in the Quanta-Ray power supply.

We found the leak in the water pump flow line (arrow up). We took the water connectors appart, rebuilt the teflon seeling and put the water connectors back again.

Laser is back in operation.
Attachment 1: QRAY_WaterPump.jpg  465 kB  Uploaded Mon Aug 2 12:06:18 2021  | Hide | Hide all