HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018 E0028 E0038
  Nuclear hyperfine mixing in 229Th89+  Not logged in ELOG logo
Message ID: 4     Entry time: Mon Aug 2 10:54:44 2021
Author: Rodolfo 
Category: Laser 
Subject: Quanta Ray - Water leak in the power supply 
After fixing the broken water connector in the laser head a new water leak appeared in the Quanta-Ray power supply.

We found the leak in the water pump flow line (arrow up). We took the water connectors appart, rebuilt the teflon seeling and put the water connectors back again.

Laser is back in operation.
Attachment 1: QRAY_WaterPump.jpg  465 kB  Uploaded Mon Aug 2 12:06:18 2021  | Show | Show all
ELOG V3.1.5-fc6679b