HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018 E0028 E0038
  nuclear two-photon decay  Not logged in ELOG logo
Message ID: 9     Entry time: Thu May 6 13:26:56 2021
Author: Helmut Weick 
Category: Detectors 
Subject: TE5 profile grids do not work 
There are several issues.

1. The stepper motor drive does not react from Device-Control, -> check controller in FRS Messhütte
2. They don't get a trigger, other grids in TE-line do -> check raw signals also in FRS Messhütte, ask Strahldiagnose
3. For less noise pockets of grids should be evacuated, but most likely the pump is still off or not even connected -> enter ESR and connect

Check also manual at https://web-docs.gsi.de/~weick/frs/current_grids.html
Below picture of no trigger shown in App. Usually it should say last trigger a few seconds ago and display some noise.
Attachment 1: no_trigger.png  521 kB  Uploaded Thu May 6 14:32:20 2021  | Show | Show all
ELOG V3.1.5-fc6679b