HOME ESR EXPERIMENTS LABS etc
NucAR lab laser_lab Fritz-Bosch-Lab X-Ray Lab target Beavertail
  NucAR lab  Not logged in ELOG logo
Message ID: 15     Entry time: Thu Jan 26 16:34:43 2023
Author: Jan 
Category: Detectors 
Subject: DSSD W1 3305-09 - UI diagram 
Attached is the UI diagram of the new double-sided silicon strip detector (DSSD). [Micron W1, ID: 3305-9] 
It was taken during the internship of Emily Voigt and Rebanna Klawitter.

Conditons:
- DSSD in dark test chamber
- under atmosphere pressure
- no alpha source or similar

Analysis:
The current is roughly linear with voltage up to values of about 110-120V. After this point the current increases more slowly, kind of a plateau.
The comparison to an older measurement [https://elog.gsi.de/esr/E127/292] with another, identical DSSD is not fully understood. No plateau seems to reached in the range up to 150V. The currents are also significanlty higher.

Conclusion:
Save operation is possible up to 100V. 
Voltages above 100V should only be used in case of ion damage and related performance losses.
Attachment 1: Si_U_I_curve.txt  220 Bytes  | Show | Hide all | Show all
Attachment 2: Praktikum_U_I.pdf  11 kB  | Show | Hide all | Show all
Attachment 3: praktikum.py  962 Bytes  | Show | Hide all | Show all
Attachment 4: W1_3305-9_U_I_diagram.txt  268 Bytes  Uploaded Fri Jan 27 16:17:44 2023  | Hide | Hide all | Show all
U[V]	I[nA]		I[nA]
10		63.5		64.0
20		74.5		75.0
30		81.0		82.0
40		85.0		88.0
50		93.5		94.0
60		103.0		103.5
70		111.5		112.0
80		120.0		120.5
90		128.0		128.0
100		135.5		135.5
110		141.5		141.5
120		146.0		146.0
130		149.0		148.5
140		151.0		151.0
150		153.5		153.5
Attachment 5: UI_diagram_W1_3305-9_2022.png  16 kB  Uploaded Fri Jan 27 16:19:06 2023  | Hide | Hide all | Show all
UI_diagram_W1_3305-9_2022.png
ELOG V3.1.5-fc6679b