HOME
ESR
EXPERIMENTS
LABS etc
|
NucAR lab
laser_lab
Fritz-Bosch-Lab
X-Ray Lab
target
Beavertail
|
NucAR lab |
Not logged in |
|
|
Message ID: 17
Entry time: Fri Jan 27 10:18:16 2023
|
Author: |
Jan Glorius, Rebanna Klawitter, Emily Voigt |
Category: |
Detectors |
Subject: |
DSSD-Performance-Test |
|
|
Detector: W1 3305-09
Setup:
- on vacuum (p=5e-5mbar)
- dark chamber
- triple alpha source
Daten: Lxg1275:/data.local2/2023_dssd_test/messung001.lmd
The resolution (sigma) is consistently below 0.5% for all strips.
dE/E = sigma/mean (root analysis)
strip dE/E
y1 0.0040
y2 0.0043
y3 0.0037
y4 0.0034
y5 0.0037
y6 0.0035
y7 0.0044
y8 0.0034
y9 0.0037
y10 0.0035
y11 0.0037
y12 0.0033
y13 0.0038
y14 0.0036
y15 0.0046
y16 0.0035
x1 0.0050
x2 0.0041
x3 0.0042
x4 0.0045
x5 0.0042
x6 0.0036
x7 0.0041
x8 0.0041
x9 0.0039
x10 0.0040
x11 0.0038
x12 0.0035
x13 0.0043
x14 0.0038
x15 0.0044
x16 0.0036 |
|
|
|
|
|
|