| ID | 
Date | 
Author | 
Category | 
Subject | 
| 
  1  
 | 
Tue Oct  6 13:04:36 2020 | 
Jan | misc | AP x-ray sources - documentation | 
Attached are the certificates and documantation of the newly arrived x-ray sources (September 2020), which will be available in the SPARC lab safe at ESR.
Quick overview:
plastic plate:
Am-241 370 kBq
Ba-133 370 kBq
Ba-133  37 kBq
Co-57  370 kBq
Pb-210  37 kBq
foil source:
Am-241  37 kBq
Fe-55  370 kBq  | 
| 
  2  
 | 
Mon Jul 11 14:05:40 2022 | 
Jan Glorius | Detectors | E146 - GEM 3000 | 
The detector GEM-3000 was used during the E146 experiment. 
beam:   208Pb82+ 
energy: 30 MeV/u 
target: H2
angle:  145° (outside)
collimator: brass, d=35mm  
comment: no particular problems with detector
Attached are several spectra with beam and with sources.  | 
| 
  3  
 | 
Mon Jul 11 16:13:34 2022 | 
Jan Glorius | Detectors | E146 - GLP 2500 | 
The detector GLP 2500 was used during the E146 experiment. 
beam:   208Pb82+ 
energy: 30 MeV/u 
target: H2
angle:  90° (outside)
collimator: brass, d=13mm  
comment: Relatively high leakage current of about 7-9 nA. 
Attached are several spectra with beam and with sources.  | 
| 
  4  
 | 
Wed Jul 26 09:11:46 2023 | 
Jan | Detectors | GLP16195-10P | 
Test measurement with the GLP16195-10P detector using 241-Am (foil source) and an ORTEC 671 with the EasyMCA (16195-am241.*).