ID |
Date |
Author |
Category |
Subject |
1
|
Fri Jun 28 10:08:25 2019 |
Jan | General | Drawing and 3D-models of the x-ray chamber | Attached is the 2d-drawing and a 3d-model of the current ESR interaction chamber.
Additonally, the drawing of the extension installed at several ports is attached.
Here is a quick overview:
Beam crosses from right to left here. Top is inside the ring, bottom is outside. |
2
|
Mon Feb 24 09:30:48 2020 |
Jan | Status Update | Installed window ports | Updated list of as of Oct 2023 before the beam time in 2024.
angle/pos. | port | current intallation
|
inside
|
35° | CF100 + extension | Be-window, 100 micrometer, dia=+35mm
|
60° | CF63 | detector pocket, 25 micrometer steel window (B. Jurado)
|
90° | CF100 | gate valve for channel plate
|
120° | CF63 | steel-window, 25 micrometer, dia=?
|
150° | CF100 + extension | Be-window, 100 micrometer, dia=+35mm
|
outside
|
60° | CF35 | Channeltron (D. Winters)
|
90° | CF100 + extension | Be-window, 100 micrometer, dia=+35mm
|
120° | CF35 | Quarzglass/fused-silica window, dia=38mm
|
145° | CF100 + extension | Be-window, 100 micrometer, dia=+35mm
|
1st downstream | CF100 | pneumatic drive + detector pocket 25 micron steel window
|
2nd downstream | CF100 | blind
|
bottom
|
1st downstream | CF100 | detector pocket 25 micron steel window
|
2nd downstream | CF100 | blind
|
1st upstream | CF100 | blind
|
top
|
1st downstream | CF100 | detector pocket 25 micron steel window
|
1st upstream | CF100 | blind
|
Updated list of as of 2021 before the beam time in 2022.
angle/pos. | port | current intallation
|
inside
|
35° | CF100 + extension | Be-window, 100 micrometer, dia=+35mm
|
60° | CF63 | detector pocket, 25 micrometer steel window (B. Jurado)
|
90° | CF100 | gate valve for channel plate
|
120° | CF63 | steel-window, 25 micrometer, dia=?
|
150° | CF100 + extension | Be-window, 100 micrometer, dia=+35mm
|
outside
|
60° | CF35 | CaF2 window, dia=?
|
90° | CF100 + extension | Be-window, 100 micrometer, dia=+35mm
|
120° | CF35 | Quarzglass/fused-silica window, dia=38mm
|
145° | CF100 + extension | Be-window, 100 micrometer, dia=+35mm
|
1st downstream | CF100 | blind (23/24: pocket of B.Jurado)
|
2nd downstream | CF100 | blind (23/24: pocket of B.Jurado)
|
bottom
|
1st downstream | CF100 | blind (23/24: pocket of B.Jurado)
|
2nd downstream | CF100 | blind (23/24: pocket of B.Jurado)
|
1st upstream | CF100 | blind
|
top
|
1st downstream | CF100 | blind (23/24: pocket of B.Jurado)
|
1st upstream | CF100 | blind
|
A list of currently installed window flanges at the target chamber, as of the last venting in late 2019.
angle | port | current intallation
|
inside
|
35° | CF100 + extension | Be-window, 100 micrometer, dia=+35mm
|
60° | CF63 | steel-window, 25 micrometer, dia=?
|
90° | CF100 | gate valve for channel plate
|
120° | CF63 | steel-window, 25 micrometer, dia=?
|
150° | CF100 + extension | Be-window, 100 micrometer, dia=+35mm
|
outside
|
60° | CF35 | unknown glass window, dia=?
|
90° | CF100 + extension | Be-window, 100 micrometer, dia=+35mm
|
120° | CF35 | Quarzglass/fused-silica window, dia=38mm
|
145° | CF100 + extension | Be-window, 100 micrometer, dia=+35mm
|
|
3
|
Mon Jul 20 09:04:17 2020 |
Jan Glorius | General | 90 degree detector extension | Attached is the drawing of the flange extension for the 90 degree ports at the target (so-called "Detektorhut").
The depth of the currently installed extensions is 95 mm, which is option "Teil 3 L = 95 mm" in the drawing. |
4
|
Mon Oct 9 15:46:59 2023 |
Jan Glorius | General | fotos of chamber interior | New detector pockets (top/bottom) and a channeltron have been installed in the ESR target chamber.
Fotos of the interior are attached, viewpoint is from the outside flange (zero degree detector position). |
5
|
Tue Oct 10 21:25:28 2023 |
Jan Glorius | General | documentation of installations oct 2023 | A summary of installations in the target line done in october 2023 is given below and in the attachement.
The work was done mainly by Jan Glorius, Davide Racano and Nikita Kotovskiy.
mirror chamber:
- exchanged "south window" [now MgF2]
- exchanged "middle window" [now CaF2]
- installed two NEG pumps, one inside and one outside the ring [before: blind flanges]
interaction chamber:
- installed two detector pockets with stainless steel windows [25 micron], one on top and one on bottom of chamber [before: blind flanges]
- installed pneumatic drive in "zero degree position" [before: blind flange], mounted is a detector pocket with a stainless steel window [25 micron]
- installed channeltron in 60° outside flange CF40 [before: CaF2 window on CF63 adapter]
- opened & closed blind flange on bottom of chamber to check potential collisions of top pocket and moving side pocket
In the attached overview all flanges opened during installation work are labeled. |
|