HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018
  Dielectronic recombination-assisted laser spectroscopy, Page 6 of 6  Not logged in ELOG logo
Entry  Wed Jun 1 08:39:50 2022, Kristian, Konstantin, Calibration, Wavemeter reading 
We recorded files with high wavemeter readout rate and saved it at the atplaser network drive (Q:) - Bismuth - wavemeter folder.

Could be used to calibrate the laser steps. The MBS readout rate is 4 Hz, here 30 Hz, so there are a few more values.
Entry  Wed Jun 1 17:20:04 2022, Rodolfo, Laser, HeNe @ WS7 HeNe_WS7_01-06-2022_1.png
 
Entry  Fri Jun 3 12:01:17 2022, Danyal, Detectors, switched off XUV detector on 1 June ~17:00  
I have switched off the XUV detector on 1 June at about 17:00. 
The high voltages of the MCP (in ESR laser lab) were turned off and inside the ESR cave the power supplies (coils and cathode) and the crate were turned
off. 
Entry  Thu Jun 30 11:12:21 2022, Carsten, Analysis, Updated Go4 Analysis with small Bugfix (PMTOverWavelength SPectra) 208Bi_20220629_fix.tgz
Here attached is a more recentversion of the Go4 Analysis Code with some smaller bugfixes.
For adaequate *.C files please have alook in 208Bi go4 directory on the atplaser account.
Please note, that there is also a (not fully up-to-date) description/manual somewhere here in the Elog.
Entry  Fri Jan 13 17:05:22 2023, Rodolfo, Laser, Testing WS6 WS6_3950_QRayWavelength_13-01-2022.png
Testing the new Wavemeter (HighFinesse, WS6, SN3950) with the QRAY laser light @ 532 nm. Multimode fiber von High Finesse:FC/PC S (Corning, MM, OM2)
Entry  Wed Jan 18 13:02:12 2023, Rodolfo, Laser, Testing WS6 with Cobra WS6_640-639.8_0.005_18-01-2023_11.23.png
Testing the WS6 (S.N. 3950) together with the Cobra. 

Cobra operated with dye DCM in Ethanol + DMSO (Concentration 1:1)
Entry  Tue Jun 6 09:33:00 2023, Rodolfo, Laser, Fiber coupler Cobra_FiberCoupler.jpg
Cobra's new fiber-coupling unit for wavelength reading.
ELOG V3.1.5-fc6679b