HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018 E0028 E0038 E0075 E0203_E0018
  masses & isomeres 2025, Page 13 of 17  Not logged in ELOG logo
ID Date Author Categorydown Subject
  81   Wed Apr 2 13:35:17 2025 Michael, Alex, MaeveDAQRSAs data taking
Since there will be no beam for about one hour, the data taking in RSAs has been stopped.
  84   Wed Apr 2 16:28:14 2025 SS, ID, HA etcDAQNTCAP stop/start
NTCAP was stopped and restarted.
  119   Thu Apr 3 11:51:00 2025 CN, IDDAQRSA started for 188Hf ESR tuning
Started both RSAs for the tuning of the ESR to 188Hf. Both pockets are in parking position OUTSIDE.
  124   Thu Apr 3 15:57:16 2025 Shahab, Yue, Zac, EstherDAQData acquisitions running
All data acquisitions (both RSAs and NTCAP) are now running.
  127   Thu Apr 3 17:38:47 2025 Shahab, Zac, Carlo, Esther and many moreDAQRSAs run again!
first file name RSA01....17.38 onwards.
  159   Fri Apr 4 16:59:25 2025 KK BJDAQLassie screen shots
  164   Fri Apr 4 17:59:33 2025 Zac , SidDAQ17:58
DAQ stopped with last file in RSA1-2025.04.04.17.58.49.340.tiq

Now the plan is to change one of the analyzers to narrow band recording for longer from the 245 Schotkky. 
  166   Fri Apr 4 18:38:52 2025 Dima, Zac, Sid, Kanika, Ragan, Shahab, BeatrizDAQnew file name and settings for 245 MHz
new settings:
1.8 seconds recording on 245 MHz on RSA02

starting from file name:

2025.04.04.18.36.32.895


settings are in the attachment
  200   Sat Apr 5 15:25:16 2025 ShahabDAQstop RSA02 looper script
2025-04-05 15:25
Stop the RSA02 looper script to change the settings of analysis
  202   Sat Apr 5 16:24:07 2025 Shahab, Dima, ZacDAQrestarted looper script for RSA02
we removed the wrong entries from the PKL file, and restarted the looper script for RSA02.

important change:

nframes = 84 
lframes = 131072
navg = 4



here are the new parameters:

# Config file for the multiprocessing looper script

[paths]
monitor_dir = "/data.local/G22-00018_00203/data/analyzers/RSA02"
output_dir = "/data.local/G22-00018_00203/results/analyzers/RSA02"
state_file = "/data.local/G22-00018_00203/scripts/e0018_e0203_scripts/processed_files_rsa2.pkl"

[processing]
num_cores = 12
interval_seconds = 0.5
file_ready_seconds = 0.5

[analysis]
nframes = 84 
lframes = 131072
navg = 4
zzmin = 0
zzmax = 1e6
mask = false
dbm = false
todo = ['spectrum', 'spectrogram'] # choose any of or combination 'png', 'spectrum', 'spectrogram'
  205   Sat Apr 5 17:01:30 2025 Ragan, Dima, Zac, ShahabDAQDAQs stopped
NTCAP and analysers are stopped at 17:00.Iris goes inside to move the detectors 
  208   Sat Apr 5 17:45:17 2025 Ragan, Shahab, Dima, ZacDAQrestarted NTCAP
NTCAP restarted 17:43

we have 63% full on bank 2 of the NAS
which means about 11% every day
this gives us roughly another 3 days of recording time
we can always switch to bank 1, but changing that, needs changing the analysis paths. so for now, we continue on bank2

:-)
  210   Sat Apr 5 18:24:09 2025 Dima, Zac, ShahabDAQRSAs stop again, due to misses
RSAs had to stop because the cycle of 1.8 came too often, both analyzers could not catch up with the file transfer.


changed to 1.7 seconds

last file in the 1.8 setting:

RSA01-2025.04.05.18.20.46.005
RSA02-2025.04.05.18.20.43.346


new filenames from the 1.7 s settings:

RSA01-2025.04.05.18.28.12.146.tiq

new analysis settings:

[analysis]
nframes = 80
lframes = 131072
avg = 4




Explanation:

every file has 10625050 samples with 1.7 seconds length
sampling rate: 6.25 MSPS
this means every sample is worth: 1.6e-7 seconds
this means every frame with length of 2^17 samples will have a time length of: 21 ms
finally with an averaging of 4 frames into one, this means that every time frame in the NPZ file will have 80ms length



frequency setting

please see attached pictures.
  214   Sat Apr 5 20:58:37 2025 ShahabDAQremoved temporary NPZ files for additional disk space
Deleted 14 TB of unused spectrum.npz files from APPC, so that there is more disk space.

switched off the production of .tiq_spectrum.npz files to save disk space in the looper script



Explanation:

.tiq_spectrogram.npz —> 3D NPZ files, that can be used for add up and create small 2D "spectrum" file for RionID

.tiq_spectrum.npz —> 2D NPZ files with super high resolution, noone needs them up to now
  237   Sun Apr 6 16:24:36 2025 Wolfram, Shahab, Moriguchi, TakaDAQNTCAP restart
NTCAP restarted with new timestamp.
  276   Mon Apr 7 15:47:03 2025 Shahab, Taka, RuijiuDAQchange NAS and restart both acquisitions
changed storage place to bank 1 of the NAS
restarted the NTCAP with new address
restarted Analyzers with new address
restarted the queqiao script
restarted looper scripts

new files on RSAs start from:

2025-04-07 15:23 onwards

new files on NTCAP have their corresponding time stamp on the directory
  284   Mon Apr 7 17:26:57 2025 Shahab, OliverDAQNew DAQ Settings
new DAQ Settings for the RSAs:

414475588 (192Pb+82   at 214 harmonic)
244.03689

plunger settings: 410 MHz cavity--> 8
245MHz cavity --> 350


recording at 50 MSPS
2^16 points each frame -->  1ms per frame

Average set to: 5 frames --> total of ca. 6.5 ms per average frame

first good NPZ file from this settings on RSAs starts from 17:41
  286   Mon Apr 7 18:17:41 2025 Shahab, OliverDAQupdated DAQ settings
changed 410 resonance frequency back to 213th harmonic
plunger 20

center freq:

412.537 MHz

first good files in this setting are 18:16 onwards.

everything else remains same like last time.
  287   Mon Apr 7 18:37:41 2025 Shahab, OliverDAQupdated DAQ settings
changed 410 resonance frequency back to 212th harmonic
plunger 33

center freq:

410,600 MHz

first good files in this setting are 18:36 onwards.

everything else remains same like last time.
  289   Mon Apr 7 18:52:15 2025 Shahab, OliverDAQupdated DAQ settings
changed 410 resonance frequency back to plunger 8
245 MHz plunger stays at 350


center freq:
410 at 414.474 MHz
245 at 244.000 MHz

first good files in this setting are 18:51 onwards.

everything else remains same like last time.
ELOG V3.1.5-fc6679b