Entry  Wed Apr 2 13:35:17 2025, Michael, Alex, Maeve, DAQ, RSAs data taking 
Since there will be no beam for about one hour, the data taking in RSAs has been stopped.
Entry  Wed Apr 2 16:28:14 2025, SS, ID, HA etc, DAQ, NTCAP stop/start 
NTCAP was stopped and restarted.
Entry  Thu Apr 3 11:51:00 2025, CN, ID, DAQ, RSA started for 188Hf ESR tuning 
Started both RSAs for the tuning of the ESR to 188Hf. Both pockets are in parking position OUTSIDE.
Entry  Thu Apr 3 15:57:16 2025, Shahab, Yue, Zac, Esther, DAQ, Data acquisitions running 
All data acquisitions (both RSAs and NTCAP) are now running.
Entry  Thu Apr 3 17:38:47 2025, Shahab, Zac, Carlo, Esther and many more, DAQ, RSAs run again! IMG_4972.JPGIMG_4973.JPG
first file name RSA01....17.38 onwards.
Entry  Fri Apr 4 16:59:25 2025, KK BJ, DAQ, Lassie screen shots 2025-04-04_16-57-35-759.png2025-04-04_16-57-28-066.png2025-04-04_16-56-54-385.png
 
Entry  Fri Apr 4 17:59:33 2025, Zac , Sid, DAQ, 17:58 
DAQ stopped with last file in RSA1-2025.04.04.17.58.49.340.tiq

Now the plan is to change one of the analyzers to narrow band recording for longer from the 245 Schotkky. 
Entry  Fri Apr 4 18:38:52 2025, Dima, Zac, Sid, Kanika, Ragan, Shahab, Beatriz, DAQ, new file name and settings for 245 MHz IMG_4981.JPGIMG_4982.JPG
new settings:
1.8 seconds recording on 245 MHz on RSA02

starting from file name:

2025.04.04.18.36.32.895


settings are in the attachment
Entry  Sat Apr 5 15:25:16 2025, Shahab, DAQ, stop RSA02 looper script 
2025-04-05 15:25
Stop the RSA02 looper script to change the settings of analysis
Entry  Sat Apr 5 16:24:07 2025, Shahab, Dima, Zac, DAQ, restarted looper script for RSA02 
we removed the wrong entries from the PKL file, and restarted the looper script for RSA02.

important change:

nframes = 84 
lframes = 131072
navg = 4



here are the new parameters:

# Config file for the multiprocessing looper script

[paths]
monitor_dir = "/data.local/G22-00018_00203/data/analyzers/RSA02"
output_dir = "/data.local/G22-00018_00203/results/analyzers/RSA02"
state_file = "/data.local/G22-00018_00203/scripts/e0018_e0203_scripts/processed_files_rsa2.pkl"

[processing]
num_cores = 12
interval_seconds = 0.5
file_ready_seconds = 0.5

[analysis]
nframes = 84 
lframes = 131072
navg = 4
zzmin = 0
zzmax = 1e6
mask = false
dbm = false
todo = ['spectrum', 'spectrogram'] # choose any of or combination 'png', 'spectrum', 'spectrogram'
Entry  Sat Apr 5 17:01:30 2025, Ragan, Dima, Zac, Shahab, DAQ, DAQs stopped 
NTCAP and analysers are stopped at 17:00.Iris goes inside to move the detectors 
Entry  Sat Apr 5 17:45:17 2025, Ragan, Shahab, Dima, Zac, DAQ, restarted NTCAP 
NTCAP restarted 17:43

we have 63% full on bank 2 of the NAS
which means about 11% every day
this gives us roughly another 3 days of recording time
we can always switch to bank 1, but changing that, needs changing the analysis paths. so for now, we continue on bank2

:-)
Entry  Sat Apr 5 18:24:09 2025, Dima, Zac, Shahab, DAQ, RSAs stop again, due to misses IMG_4991.JPGIMG_4992.JPG
RSAs had to stop because the cycle of 1.8 came too often, both analyzers could not catch up with the file transfer.


changed to 1.7 seconds

last file in the 1.8 setting:

RSA01-2025.04.05.18.20.46.005
RSA02-2025.04.05.18.20.43.346


new filenames from the 1.7 s settings:

RSA01-2025.04.05.18.28.12.146.tiq

new analysis settings:

[analysis]
nframes = 80
lframes = 131072
avg = 4




Explanation:

every file has 10625050 samples with 1.7 seconds length
sampling rate: 6.25 MSPS
this means every sample is worth: 1.6e-7 seconds
this means every frame with length of 2^17 samples will have a time length of: 21 ms
finally with an averaging of 4 frames into one, this means that every time frame in the NPZ file will have 80ms length



frequency setting

please see attached pictures.
Entry  Sat Apr 5 20:58:37 2025, Shahab, DAQ, removed temporary NPZ files for additional disk space 
Deleted 14 TB of unused spectrum.npz files from APPC, so that there is more disk space.

switched off the production of .tiq_spectrum.npz files to save disk space in the looper script



Explanation:

.tiq_spectrogram.npz —> 3D NPZ files, that can be used for add up and create small 2D "spectrum" file for RionID

.tiq_spectrum.npz —> 2D NPZ files with super high resolution, noone needs them up to now
Entry  Sun Apr 6 16:24:36 2025, Wolfram, Shahab, Moriguchi, Taka, DAQ, NTCAP restart 
NTCAP restarted with new timestamp.
Entry  Mon Apr 7 15:47:03 2025, Shahab, Taka, Ruijiu, DAQ, change NAS and restart both acquisitions 
changed storage place to bank 1 of the NAS
restarted the NTCAP with new address
restarted Analyzers with new address
restarted the queqiao script
restarted looper scripts

new files on RSAs start from:

2025-04-07 15:23 onwards

new files on NTCAP have their corresponding time stamp on the directory
Entry  Mon Apr 7 17:26:57 2025, Shahab, Oliver, DAQ, New DAQ Settings 2025-04-07_16-57-16-424.png2025-04-07_17-15-22-069.png
new DAQ Settings for the RSAs:

414475588 (192Pb+82   at 214 harmonic)
244.03689

plunger settings: 410 MHz cavity--> 8
245MHz cavity --> 350


recording at 50 MSPS
2^16 points each frame -->  1ms per frame

Average set to: 5 frames --> total of ca. 6.5 ms per average frame

first good NPZ file from this settings on RSAs starts from 17:41
Entry  Mon Apr 7 18:17:41 2025, Shahab, Oliver, DAQ, updated DAQ settings 
changed 410 resonance frequency back to 213th harmonic
plunger 20

center freq:

412.537 MHz

first good files in this setting are 18:16 onwards.

everything else remains same like last time.
Entry  Mon Apr 7 18:37:41 2025, Shahab, Oliver, DAQ, updated DAQ settings 
changed 410 resonance frequency back to 212th harmonic
plunger 33

center freq:

410,600 MHz

first good files in this setting are 18:36 onwards.

everything else remains same like last time.
Entry  Mon Apr 7 18:52:15 2025, Shahab, Oliver, DAQ, updated DAQ settings 
changed 410 resonance frequency back to plunger 8
245 MHz plunger stays at 350


center freq:
410 at 414.474 MHz
245 at 244.000 MHz

first good files in this setting are 18:51 onwards.

everything else remains same like last time.