HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018
  Bound beta decay experiment on Tl-205, Page 9 of 9  Not logged in ELOG logo
Entry  Sun Mar 29 21:02:43 2020, ruijiu, Analysis, decay line of 205Tl beam main_s6_neg8_202003292059.pngmain_s6_neg8_202003292059.root
decay line of 205Tl beam with Ar gas jet target on and S6 slit at -8mm
Ar gas density~1e12 atoms/cm2
Entry  Sun Mar 29 21:39:27 2020, ruijiu, Analysis, Ionization cross section and capture cross section for h-206Pb + Ar 202003292141.png
Ionization cross section = 1320 barn.
Capture cross section = 557+137+44.5=735.5.
sigma_Ion/sigma_cap=1.79
Entry  Mon Mar 30 12:12:25 2020, ruijiu, General, Instruction of NTCAP for shift person Instruction_of_NTCAP_for_shift_person.png
To start a new run with NTCap, we need to retart NTCAP. 
1. Stop labview program. 
2. Start labview program and start a new run.
Entry  Mon Mar 30 13:40:57 2020, ruijiu, Detectors, HV of Inner detector IMG_20200330_133909.jpg
 
Entry  Mon Mar 30 13:41:20 2020, ruijiu, Detectors, HV of Inner detector IMG_20200330_133909.jpg
 
Entry  Mon Mar 30 14:24:46 2020, ruijiu, DAQ, stop taking data. 
1. stop run0272.lmd.
2. stop NTCAP. 
  NTCAp file:
Entry  Mon Mar 30 16:25:05 2020, ruijiu, Detectors, turn down the HV 
turn down the hv and pull out the detector.  

suggest by 
Entry  Mon Mar 30 16:37:53 2020, ruijiu, Analysis,  
 
Entry  Tue Mar 31 15:30:43 2020, ruijiu, Analysis, The beam position on DSSD with different conditions on on the dEE spectrum 6x
I analyzed the first measurement with 206Pb81+ beam, run 250-267. 
Entry  Wed Apr 1 01:02:24 2020, ruijiu, Detectors, the detector cable is disconnected. 
 
Entry  Wed Apr 1 01:05:36 2020, ruijiu, Detectors, detector position fixed at 60 mm VID_20200401_010416.mp4
The detector cable was disconnected . We entered and reconnected the cable.
Entry  Thu Apr 2 12:52:25 2020, ruijiu, General, setup FRS with 2 g/cm2 degrader. 
Yuri sets up FRS with 2 g/cm2 degrader (previous one is 1 g/cm2). The ESR is not changed. The contamination should be less now.
Entry  Thu Apr 9 11:47:28 2020, ruijiu, , backup data IMG_20200409_113155.jpg
2020.4.8  12:00Started copying 
data。
2020.4.9    11:52 Copied 6.3TB 
Entry  Fri Apr 17 18:27:34 2020, ruijiu, Analysis, backup data IMG_20200417_162202.jpg
1.connect USB harddisk  to computer lxg1254 and  copy data to sever.

sync -rtmv --progress --checkSUD  /media/sdgl/data/ litv-exp@sbk0l9:/lustre/ap/litv-exp/2020-
Entry  Mon Mar 30 15:03:58 2020, ruijiu Tino, Yuri, Lasoro, Nik and Uli, Detectors, the gas bottle is empty. IMG_20200330_152735_mh1585575780057.jpg
The gas bottle was empty. Tino replaced 
the previous old bottle with 50 bar gas left. 
Entry  Mon Mar 30 14:48:02 2020, ruijiu and Tino, Detectors, The leakcurrent of SiPad is over the setting limit. Screenshot_20200330-145551.jpgScreenshot_20200330-145544.jpgIMG_20200330_144915.jpg
Because the 
leakcurrent of 
6 Sipads is 
Entry  Tue Feb 25 15:48:31 2020, ruijiu chen, DAQ, Important folders of e121 eperiments: 
1. slow control:
1.1. program file:
litv-exp@ringpi:~/epics/mrcc_ioc/db $ 
Entry  Tue Feb 25 15:49:58 2020, ruijiu chen, DAQ, changing scaling factor. 
How to change scaling factor?
The parameters is in f_user.c file:
1. open the f_user.C with emacs :
Entry  Sat Mar 28 16:12:25 2020, ruijiu, ragan, DAQ, enter ESR and turn on the HV crate. 
 
Entry  Mon Mar 30 07:02:15 2020, tino, Runs, Second measurement with 206Pb81+ beam at I_cooler = 20 mA  
I_cooler = 20 mA

beam: 206Pb81+
ELOG V3.1.5-fc6679b