ID |
Date |
Author |
Category |
Subject |
Year |
16
|
Mon Jul 8 14:57:07 2019 |
Yuanming Xing | Detectors | The measurement of the tilting angle | |
As shown in the figure, the detector is tilted by a certain angle which is about 45 degree.
The tilted angle is defined as angle B.
We measured angle A using the vertical laser line (red) and papers with a straight line edge. So we can marked
the line denoted by the laser line when the paper is put on the holder.
In the beginning, the angle A is measured as 43.3±0.5 degree. So angle B= 46.7±0.5 degree
After the adjustment of the detector (to make it in the center of the chamber), the angle A is measured as
43.1±0.5 degree with another paper. So angle B= 46.9±0.5 degree. |
Attachment 1: TiltingAngle.png
|
|
17
|
Tue Jul 9 15:26:59 2019 |
Yuanming Xing | Simulations | Simulated Rutherford scattering distribution on detector | |
Simulated Rutherford scattering distribution on detector
Beam: 111Sn
Luminosity: 3.08 /mb/s
Distance between the detector and central orbit: 2.80cm
Particle Number in the simulation: 5 Million
Beam quality used in MOCADI input file:
4,
0.27386, 0.182573, 0, -4.0, 0
4,
0.131477, 0.383186, 0, 0.0, 0
1,
0.02, 0, 0, 0.0, 0
1. Beam Energy: 6MeV/u
2. Beam Energy: 7MeV/u
3. Beam Energy: 8MeV/u |
Attachment 1: 6MeV.png
|
|
Attachment 2: 7MeV.png
|
|
Attachment 3: 8MeV.png
|
|
353
|
Wed May 19 19:47:14 2021 |
Yuri | General | Beam Halflife at 30 AMeV | 2021 |
|
Attachment 1: 2021-05-19_17-35-00_tcl1032_lassie-monitor.png
|
|
489
|
Thu May 27 08:23:05 2021 |
Yuri, Beatriz | Runs | Run 0148 | 2021 |
No. of stacks: 20
length SC10: 3x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -11.5 mm
open file : e127b_run0148.lmd
beam: 118Te52+
energy: 6 MeV/u
cooler voltage:
purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)
Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +2.0 mm
run start: 08:22:28 27.05.2021
run stop: 09:38:31 27.05.2021
_______________________________________________________________________________
ON Rates
Si_ON_bdt: ~ 3 Hz
Xray_ON_bdt: 100 Hz
BaF_OR: ---
Si voltage: 100.0 V
Si leakage current: 380 nA
SIS particles before: 2.6e9 (one shoot)
ESR current at injection: 4.6e6 (stacked)
ESR particles after decel.: Unreadable
Target ON density: 1.0e14
Beam was off for a few minutes, after coming back the number of particles was reduced for a few minutes and then increased to the values at the beginning of the run. |
490
|
Thu May 27 08:26:15 2021 |
Yuri, Beatriz | General | Optimisation beam position on target and scraper position | 2021 |
Optimised overlap with the target by shifting the beam from -11.5 to 12.8 mm. This lead to an increase of the DSSSD rate of a factor between 2.5 or 3.
Adjusted scraper position by looking at x-ray rate and move it from +1 to +2. |
491
|
Thu May 27 09:39:15 2021 |
Yuri, Beatriz | Runs | Run 0149 | 2021 |
No. of stacks: 20
length SC10: 3x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm
open file : e127b_run0149.lmd
beam: 118Te52+
energy: 6 MeV/u
cooler voltage:
purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)
Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +2.0 mm
run start: 09:38:49 27.05.2021
run stop: 10:09:21 27.05.2021 (stopped for filling det.)
_______________________________________________________________________________
ON Rates
Si_ON_bdt: ~ 3 Hz
Xray_ON_bdt: 100 Hz
BaF_OR: ---
Si voltage: 100.0 V
Si leakage current: 380 nA
SIS particles before: 2.6e9 (one shoot)
ESR current at injection: 2.0e6 (stacked)
ESR particles after decel.: Unreadable
Target ON density: 1.0e14 |
492
|
Thu May 27 10:21:41 2021 |
Yuri, Beatriz | Runs | Run 0150 | 2021 |
No. of stacks: 10
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm
open file : e127b_run0150.lmd
beam: 118Te52+
energy: 6 MeV/u
cooler voltage:
purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)
Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +2.0 mm
run start: 10:21:17 27.05.2021
run stop: 11:35:36 27.05.2021
_______________________________________________________________________________
ON Rates
Si_ON_bdt: ~ 5 Hz
Xray_ON_bdt: 80 Hz
BaF_OR: ---
Si voltage: 100.0 V
Si leakage current: 380 nA
SIS particles before: 2.6e9 (one shoot)
ESR current at injection: 2.0e6 (stacked)
ESR particles after decel.: Unreadable
Target ON density: 1.0e14 |
493
|
Thu May 27 11:36:23 2021 |
Yuri, Beatriz, Jan | Runs | Run 0151 | 2021 |
Power supplies for correction coils were in failure.
No effect seen in schottky after reset and repower, however.
No. of stacks: 10 (switched to 20 shortly)
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm
open file : e127b_run0151.lmd
beam: 118Te52+
energy: 6 MeV/u
cooler voltage:
purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)
Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +2.0 mm
run start: 11:36:01 27.05.2021
run stop: 12:39:18 27.05.2021
_______________________________________________________________________________
ON Rates
Si_ON_bdt: ~ 3 Hz
Xray_ON_bdt: 90 Hz
BaF_OR: ---
Si voltage: 100.0 V
Si leakage current: 380 nA
SIS particles before: 2.5e9 (one shoot)
ESR current at injection: 2.0e6 (stacked)
ESR particles after decel.: Unreadable
Target ON density: 1.0e14 |
494
|
Thu May 27 12:41:25 2021 |
Yuri, Jan | Runs | Run 0152 | 2021 |
No. of stacks: 10
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm
open file : e127b_run0152.lmd
beam: 118Te52+
energy: 6 MeV/u
cooler voltage:
purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)
Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +2.0 mm
run start: 12:39:21 27.05.2021
run stop: 13:23:57 27.05.2021
_______________________________________________________________________________
ON Rates
Si_ON_bdt: ~ 3 Hz
Xray_ON_bdt: 90 Hz
BaF_OR: ---
Si voltage: 100.0 V
Si leakage current: 380 nA
SIS particles before: 2.5e9 (one shoot)
ESR current at injection: 2.0e6 (stacked)
ESR particles after decel.: Unreadable
Target ON density: 1.0e14 |
495
|
Thu May 27 13:24:29 2021 |
Yuri, Jan | Runs | Run 0153 | 2021 |
All runs before (since 13:40 26/05/21) were taken without the degrader in FRS!
No. of stacks: 10
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm
open file : e127b_run0153.lmd
beam: 118Te52+
energy: 6 MeV/u
cooler voltage:
purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)
Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +2.0 mm
run start: 13:23:57 27.05.2021
run stop: 13:31:44 27.05.2021
_______________________________________________________________________________
ON Rates
Si_ON_bdt: ~ 3 Hz
Xray_ON_bdt: 90 Hz
BaF_OR: ---
Si voltage: 100.0 V
Si leakage current: 380 nA
SIS particles before: 2.5e9 (one shoot)
ESR current at injection: 2.0e6 (stacked)
ESR particles after decel.: Unreadable
Target ON density: 1.0e14 |
499
|
Thu May 27 14:38:14 2021 |
Yuri, Jan | Runs | Run 0155 | 2021 |
No. of stacks: 10
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm
open file : e127b_run0155.lmd
beam: 118Te52+
energy: 6 MeV/u
cooler voltage:
purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)
Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +2.0 mm
run start: 14:38:00 27.05.2021
run stop: 15:11:15 27.05.2021
_______________________________________________________________________________
ON Rates
Si_ON_bdt: ~ 3 Hz
Xray_ON_bdt: 90 Hz
BaF_OR: ---
Si voltage: 100.0 V
Si leakage current: 380 nA
SIS particles before: 2.5e9 (one shoot)
ESR current at injection: 2.0e6 (stacked)
ESR particles after decel.: Unreadable
Target ON density: 1.0e14 |
241
|
Mon Mar 23 12:30:39 2020 |
laszlo | Analysis | with and without scraper | |
|
Attachment 1: 124Xe_with_scraper_3D.png
|
|
Attachment 2: 124Xe_with_scraper_2D.png
|
|
Attachment 3: 124Xe_without_scraper_2D.png
|
|
Attachment 4: 124Xe_without_scraper_3D.png
|
|