  539   Sat May 29 08:01:25 2021 Rene, TinoRunsRun 01912021
No. of stacks: 10 00:52:43
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm

open file : e127b_run0191.lmd

beam: 118Te52+
energy: 6 MeV/u
cooler voltage:

purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)

Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +2.0 mm

run start: 08:00:39 29.05.2021
run stop: 08:58:46 29.05.2021
_______________________________________________________________________________

ON Rates
Si_ON_bdt: ~ 3 Hz
Xray_ON_bdt: ~ 120 Hz
BaF_OR: ---

Si voltage: 100.0 V
Si leakage current: 380 nA


SIS particles before: 2.4e9 (one shoot)
ESR current at injection: 3.3e6 (stacked)
ESR particles after decel.: 1.5e6
Target ON density: 1.0e14
  540   Sat May 29 08:59:52 2021 Rene, TinoRunsRun 01922021
No. of stacks: 10 00:52:43
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm

open file : e127b_run0192.lmd

beam: 118Te52+
energy: 6 MeV/u
cooler voltage:

purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)

Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +2.0 mm

run start: 09:07:18 29.05.2021
run stop: 10:07:34 29.05.2021
_______________________________________________________________________________

ON Rates
Si_ON_bdt: ~ 2.5 Hz
Xray_ON_bdt: ~ 110 Hz
BaF_OR: ---

Si voltage: 100.0 V
Si leakage current: 380 nA


SIS particles before: 2.7e9 (one shoot)
ESR current at injection: 3.0e6 (stacked)
ESR particles after decel.: 2.5e6
Target ON density: 1.0e14
  541   Sat May 29 10:09:01 2021 Rene, TinoRunsRun 01932021
No. of stacks: 10 00:52:43
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm

open file : e127b_run0193.lmd

beam: 118Te52+
energy: 6 MeV/u
cooler voltage:

purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)

Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +2.0 mm

run start: 10:07:35 29.05.2021
run stop: 10:59:10 29.05.2021
_______________________________________________________________________________

ON Rates
Si_ON_bdt: ~ 3 Hz
Xray_ON_bdt: ~ 110 Hz
BaF_OR: ---

Si voltage: 100.0 V
Si leakage current: 380 nA


SIS particles before: 2.7e9 (one shoot)
ESR current at injection: 3.0e6 (stacked)
ESR particles after decel.: 2.5e6
Target ON density: 1.0e14
  542   Sat May 29 11:01:04 2021 Rene, TinoRunsRun 0194 crap run2021
No. of stacks: 10 00:52:43
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm

open file : e127b_run0194.lmd

beam: 118Te52+
energy: 6 MeV/u
cooler voltage:

purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)

Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +2.0 mm

run start: 12:22:55 29.05.2021
run stop: 12:49:05 29.05.2021
_______________________________________________________________________________

ON Rates
Si_ON_bdt: ~ 0 Hz
Xray_ON_bdt: ~ 0 Hz
BaF_OR: ---

Si voltage: 100.0 V
Si leakage current: 380 nA


SIS particles before: 2.7e9 (one shoot)
ESR current at injection: 3.0e6 (stacked)
ESR particles after decel.: not measurable
Target ON density: 1.0e14
  470   Wed May 26 14:48:10 2021 Ragan, ThanassisRunsRun 01342021
No. of stacks: 20
length SC10: 3x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -11.5 mm

open file : e127b_run0134.lmd

beam: 118Te52+
energy: 7 MeV/u
cooler voltage:

purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)

Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +3.5 mm (20mm away from beam)

run start: 14:47:22 26.05.2021
run stop: 15:28:45 26.05.2021
_______________________________________________________________________________

ON Rates
Si_ON_bdt: ~ 8 Hz
Xray_ON_bdt: 210 Hz
BaF_OR: ---

Si voltage: 100.0 V
Si leakage current: 380 nA


SIS particles before: 2.1e9 (one shoot)
ESR current at injection: 5.4e6 (stacked)
ESR particles after decel.: 0.9e6 (more or less)
Target ON density: 1.0e14
  471   Wed May 26 15:30:03 2021 Ragan, ThanassisGeneralChanging the beam energy 2021
The ESR team is changing the beam energy to 6 MeV/u.
  473   Wed May 26 18:36:08 2021 Ragan, ThanassisRunsRun 01352021
No. of stacks: 20
length SC10: 3x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -11.5 mm


open file : e127b_run0135.lmd

beam: 118Te52+
energy: 6 MeV/u
cooler voltage:

purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)

Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +1.0 mm (20mm away from beam)

run start: 18:18:10 26.05.2021
run stop: 19:07:37 26.05.2021
_______________________________________________________________________________

ON Rates
Si_ON_bdt: ~ 6 Hz
Xray_ON_bdt: 80 Hz
BaF_OR: ---

Si voltage: 100.0 V
Si leakage current: 380 nA


SIS particles before: 2.1e9 (one shoot)
ESR current at injection: 5.4e6 (stacked)
ESR particles after decel.: 0.9e6 (more or less)
Target ON density: 1.0e14
  474   Wed May 26 18:37:34 2021 Ragan, ThanassisGeneralOptimization target bump/overlapp2021
| bump position | X ray bdt ON | Si rate ON |
---------------------------------------------
| -11.5         | 120          | 4-5        |
| -10.5         | 100          | 5          |
| -9.5          | 85           | 5          |
| -12.5         | 100          | 6          |

-11.5 bump position is taken for the data taking.
  475   Wed May 26 19:08:17 2021 Ragan, ThanassisGeneralCooler down2021
We stopped data taking as the electron cooler is down.
  476   Wed May 26 19:25:54 2021 Ragan, ThanassisRunsRun 01362021
No. of stacks: 20
length SC10: 3x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -11.5 mm

open file : e127b_run0136.lmd

beam: 118Te52+
energy: 6 MeV/u
cooler voltage:

purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)

Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +1.0 mm (20mm away from beam)

run start: 19:25:38 26.05.2021
run stop: 20:25:40 26.05.2021
_______________________________________________________________________________

ON Rates
Si_ON_bdt: ~ 6 Hz
Xray_ON_bdt: 80 Hz
BaF_OR: ---

Si voltage: 100.0 V
Si leakage current: 380 nA


SIS particles before: 2.1e9 (one shoot)
ESR current at injection: 5.4e6 (stacked)
ESR particles after decel.: 0.9e6 (more or less)
Target ON density: 1.0e14
  477   Wed May 26 20:26:15 2021 Ragan, ThanassisRunsRun 01372021
No. of stacks: 20
length SC10: 3x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -11.5 mm


open file : e127b_run0137.lmd

beam: 118Te52+
energy: 6 MeV/u
cooler voltage:

purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)

Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +1.0 mm (20mm away from beam)

run start: 20:25:44 26.05.2021
run stop: 20:33:11 26.05.2021
_______________________________________________________________________________

ON Rates
Si_ON_bdt: ~ 6 Hz
Xray_ON_bdt: 100 Hz
BaF_OR: ---

Si voltage: 100.0 V
Si leakage current: 380 nA


SIS particles before: 7.0e8 (one shoot)
ESR current at injection: 5.0e6 (stacked)
ESR particles after decel.: 2.3e5 (more or less)
Target ON density: 1.0e14
  478   Wed May 26 20:34:50 2021 Ragan, ThanassisDetectorsDetector filling2021
Jan is filling the detector after run0137.
  479   Wed May 26 20:42:31 2021 Ragan, ThanassisRunsRun 01382021
No. of stacks: 20
length SC10: 3x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -11.5 mm


open file : e127b_run0138.lmd

beam: 118Te52+
energy: 6 MeV/u
cooler voltage:

purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)

Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +1.0 mm

run start: 20:42:13 26.05.2021
run stop: 21:42:20 26.05.2021
_______________________________________________________________________________

ON Rates
Si_ON_bdt: ~ 5 Hz
Xray_ON_bdt: 100 Hz
BaF_OR: ---

Si voltage: 100.0 V
Si leakage current: 380 nA


SIS particles before: 7.0e8 (one shoot)
ESR current at injection: 5.0e6 (stacked)
ESR particles after decel.: 2.3e5 (more or less)
Target ON density: 1.0e14
  480   Wed May 26 21:42:46 2021 Ragan, ThanassisRunsRun 01392021
No. of stacks: 20
length SC10: 3x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -11.5 mm


open file : e127b_run0139.lmd

beam: 118Te52+
energy: 6 MeV/u
cooler voltage:

purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)

Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +1.0 mm

run start: 21:42:25 26.05.2021
run stop: 22:42:33 26.05.2021
_______________________________________________________________________________

ON Rates
Si_ON_bdt: ~ 5-6 Hz
Xray_ON_bdt: 120 Hz
BaF_OR: ---

Si voltage: 100.0 V
Si leakage current: 380 nA


SIS particles before: 7.0e8 (one shoot)
ESR current at injection: 2.0e6 (stacked)
ESR particles after decel.: 2.3e5 (more or less)
Target ON density: 1.0e14
  481   Wed May 26 22:42:56 2021 Ragan, ThanassisRunsRun 01402021
No. of stacks: 20
length SC10: 3x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -11.5 mm


open file : e127b_run0140.lmd

beam: 118Te52+
energy: 6 MeV/u
cooler voltage:

purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)

Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +1.0 mm

run start: 22:42:39 26.05.2021
run stop: 23:45:58 26.05.2021
_______________________________________________________________________________

ON Rates
Si_ON_bdt: ~ Hz
Xray_ON_bdt: Hz
BaF_OR: ---

Si voltage: 100.0 V
Si leakage current: 380 nA


SIS particles before: 7.0e8 (one shoot)
ESR current at injection: 3.0e6 (stacked)
ESR particles after decel.: 2.3e5 (more or less)
Target ON density: 1.0e14
  524   Fri May 28 16:41:07 2021 Ragan, ThanassisRunsRun 01762021
No. of stacks: 10
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm

open file : e127b_run0176.lmd

beam: 118Te52+
energy: 6 MeV/u
cooler voltage:

purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)

Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +2.0 mm

run start: 16:51:23 28.05.2021
run stop: 17:51:26 28.05.2021
_______________________________________________________________________________

ON Rates
Si_ON_bdt: ~4 Hz
Xray_ON_bdt: 100 Hz
BaF_OR: ---

Si voltage: 100.0 V
Si leakage current: 380 nA


SIS particles before: 2.2e8 (one shoot)
ESR current at injection: 2.4e6 (stacked)
ESR particles after decel.: 3.6e5
Target ON density: 1.0e14
  525   Fri May 28 17:51:54 2021 Ragan, ThanassisRunsRun 01772021
No. of stacks: 10
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm

open file : e127b_run0177.lmd

beam: 118Te52+
energy: 6 MeV/u
cooler voltage:

purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)

Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +2.0 mm

run start: 17:51:33 28.05.2021
run stop: 18:51:36 28.05.2021
_______________________________________________________________________________

ON Rates
Si_ON_bdt: ~3-4 Hz
Xray_ON_bdt: ~100-110 Hz
BaF_OR: ---

Si voltage: 100.0 V
Si leakage current: 380 nA


SIS particles before: 7.0e8 (one shoot)
ESR current at injection: 2.2e6 (stacked)
ESR particles after decel.: 6.9e5
Target ON density: 1.0e14
  526   Fri May 28 18:52:00 2021 Ragan, ThanassisRunsRun 01782021
No. of stacks: 10
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm

open file : e127b_run0178.lmd

beam: 118Te52+
energy: 6 MeV/u
cooler voltage:

purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)

Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +2.0 mm

run start: 18:51:44 28.05.2021
run stop: 19:51:50 28.05.2021
_______________________________________________________________________________

ON Rates
Si_ON_bdt: ~4-5 Hz
Xray_ON_bdt: ~110 Hz
BaF_OR: ---

Si voltage: 100.0 V
Si leakage current: 380 nA


SIS particles before: 2.0e9 (one shoot)
ESR current at injection: 2.3e6 (stacked)
ESR particles after decel.: 6.9e5
Target ON density: 1.0e14
  527   Fri May 28 19:52:09 2021 Ragan, ThanassisRunsRun 01792021
No. of stacks: 10
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm

open file : e127b_run0179.lmd

beam: 118Te52+
energy: 6 MeV/u
cooler voltage:

purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)

Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +2.0 mm

run start: 19:51:51 28.05.2021
run stop: 20:37:04 28.05.2021
_______________________________________________________________________________

ON Rates
Si_ON_bdt: ~4-5 Hz
Xray_ON_bdt: ~ 110 Hz
BaF_OR: ---

Si voltage: 100.0 V
Si leakage current: 380 nA


SIS particles before: 2.4e9 (one shoot)
ESR current at injection: 2.6e6 (stacked)
ESR particles after decel.: 6.9e5
Target ON density: 1.0e14
  528   Fri May 28 20:37:42 2021 Ragan, ThanassisRunsRun 01802021
No. of stacks: 10
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm

open file : e127b_run0180.lmd

beam: 118Te52+
energy: 6 MeV/u
cooler voltage:

purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)

Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +2.0 mm

run start: 20:54:46 28.05.2021
run stop: 21:53:27 28.05.2021
_______________________________________________________________________________

ON Rates
Si_ON_bdt: ~ 4-5 Hz
Xray_ON_bdt: ~ 120 Hz
BaF_OR: ---

Si voltage: 100.0 V
Si leakage current: 380 nA


SIS particles before: 2.4e9 (one shoot)
ESR current at injection: 2.6e6 (stacked)
ESR particles after decel.: 6.9e5
Target ON density: 1.0e14