ID |
Date |
Author |
Category |
Subject |
Text |
|
28
|
Wed Mar 23 11:20:55 2022 |
Simon | Detectors | PMT operating voltages | During the testing of the PMTs the following
operating voltages were applied:
|
|
77
|
Sun Apr 3 11:40:48 2022 |
Shahab, Yuri | Detectors | 245 MHz Resonator tuned to new frequency | resonance frequency changed from 243.2 MHz
to 244.692 MHz by reducing the value of both
of the plungers from 395 to 325. Now after |
|
44
|
Thu Mar 31 20:07:47 2022 |
Shahab | DAQ | Schottky Display in AP Messhütte (or anywhere else) | [B]UPDATE: 16.05.2022[/B]
[I]Please use ATPPC029 instead of ATPPC019 |
|
137
|
Thu Apr 7 14:01:22 2022 |
Sebastian, Max | Detectors | Mesytec reset | The Mesytec crate was turned off and on again.
It is again possible to communicate with |
|
228
|
Thu Apr 14 12:27:18 2022 |
Sebastian, Max | Detectors | Connection to XUV / Cu reestablished! | The connection to the XUV and Cu were lost
during the nightshift.
|
|
95
|
Tue Apr 5 11:14:52 2022 |
Sebastian, Kristian, Pierre-Michel | DAQ | PMT over wavelength histogram v2 | Today we finalized the PMT over wavelength
histogram in:
|
|
100
|
Tue Apr 5 15:25:47 2022 |
Sebastian, Kristian, Pierre-Michel | DAQ | Go4 Conditions for "TDC_stacked_Sig/Ref" histogram | First version of TDC_stacked histograms implemented.
Under folder "Conditions" you can now integrate |
|
208
|
Mon Apr 11 11:05:35 2022 |
Sebastian, Kristian | Laser | Check laser power | we checked the UV laser power at 10:15 am.
It was only 8.5 - 9 mJ. Pumping power was
460 mJ, which is the maximum output of the |
|
87
|
Mon Apr 4 15:27:31 2022 |
Sebastian | DAQ | PMT over wavelength histogram | Kristian and I tried to create a "PMT over
wavelength" histogram under step2Anl/LabView.
Unfortunately, the code ist still in progress |
|
133
|
Thu Apr 7 09:23:18 2022 |
Sebastian | DAQ | DAQ1 & DAQ2 - 1kHz and delayed cs removed | Questionable 1kHz and delayed common stop
signals has been removed.
|
|
152
|
Fri Apr 8 08:17:52 2022 |
Sebastian | Laser | Laser Ext Trig is working | We (Max, Kristian, Danyal, me) fixed the "external
trigger" problem.
|
|
153
|
Fri Apr 8 08:40:06 2022 |
Sebastian | DAQ | DAQ improvement | Concerning ID 146 a new histogram for accumulating
data (no cleaning histograms) can be found
under
|
|
229
|
Thu Apr 14 12:38:38 2022 |
Sebastian | Detectors | XUV / Cu checkups | Check regularly if the connection to the XUV
and Cu detectors are still active and the
detectors move in (after injection) and out |
|
231
|
Thu Apr 14 19:46:46 2022 |
Sarper & Konstantin & Rodolfo | Laser | Laser power | We checked the laser power before starting
the new scan range.
|
|
78
|
Sun Apr 3 12:31:10 2022 |
Rui-Jiu Chen, Yuri, Carsten, Ragan | Runs | Measurement of beam with different circles | Parameters of NTCAP:
1. IQ rate 20 MSamples/s
|
6x |
64
|
Sat Apr 2 10:22:21 2022 |
Rui-Jiu Chen, Ragan | General | Beam is in the ESR | Primary beam is in the ESR now. Markus and
the team are tuning the ESR. |
|
51
|
Fri Apr 1 11:00:41 2022 |
Rui-Jiu Chen | DAQ | Start NTCAP | Start taking data with NTCAP.
Carrier Frequency: 245MHz
IQ rate:5 MSamples/s
|
|
52
|
Fri Apr 1 11:19:57 2022 |
Rui-Jiu Chen | DAQ | S.M.A.R.T check on NTCAP | S.M.A.R.T. check has been done on NTCAP on
March 28th. All 12 hard disk drivers are
normal. |
|
211
|
Mon Apr 11 18:46:31 2022 |
Rodolfo, Sarper and Konstantin | Laser | Alignment-check | We cross-checked the laser-ion alignment.
The horizontal position was optimized. However,
|
|
224
|
Wed Apr 13 15:41:29 2022 |
Rodolfo & Sarper | Laser | Laser status | We have checked the power of the laser and
optimized it "a bit".
|
|