  4   Tue May 4 17:54:08 2021 Helmut WeickSimulationsupdated LISE file
This file really has the optics like used in the last ESR experiments.
Now the transmission is also a little bit better.
  52   Fri May 7 17:12:52 2021 Kanika, Sergey, Yuri, Wolfram, Michael, ShahabRunstrigger time back to 50%
changed trigger time back to 50%

last file with 1% trigger: 

245 MHZ 2021.05.07.16.07.03.493
410 MHz 2021.05.07.17.07.21.104


p.s. remember the 245 Analyzer has a 1 hour offset in time stamp because of daylight saving time in win10 setting
  136   Mon Jun 28 18:32:49 2021 Ruijiu, ShahabDAQtrigger signal connected to NTCAP and analyzers
  143   Wed Jun 30 00:12:55 2021 Rui-Jiu ChenAnalysisthe ID of 72Ge setting
  168   Thu Jul 1 15:25:42 2021 Rui-Jiu ChenDAQtemplate of NTCAP shift
Please check following two parameters every two hours into make sure the NTCAP is running well: 
1. The latest recorded tdms files:
 p:\E143\sc\**\**
 p:\E143\iq\**\**
2. The rest space on the drive "p:\" : **GByte.
  114   Sun May 9 21:29:26 2021 Helmut, Wolfram, YuriAcceleratortarget TE3UF shift
We see a very strong line at the left end of spectrum, stronger than all other lines.
When we shift the target TE3UF_H from -35mm to -37mm it gete weaker, but -37mm is the end of drive.
In the other direction it becomes worse. At -25mm completely new peaks.
Also scan in Y direction TE3UF_V +25mm and +40mm give other fragments even very intense, while the old lines vanish.
So production happens in target frame. 
Use -37mm, +37mm as best target position.

We run with TCap only for some time hoping there is somewhere a peak of 70Se
running from about 9:15 to about 22:00
We set scrapers in the previous optimum position
  84   Sat May 8 22:06:36 2021 Michael, Ruijiu, Helmut, Yuri, Wolfram, Sergey, ShahabRunsstopped data taking with the high resolution
last file names with this setting:

245: 2021.05.08.20.42.03.743.tiq
410: 2021.05.08.21.42.17.969.tiq
  109   Sun May 9 18:41:33 2021 Rodostina, Oliver, Wolfram, Helmut, Yuri, ShahabRunsstopped data taking for energy change
last recorded filename:

E143-245MHz-2021.05.09.17.38.44.410.tiq

410 was not recording data in this period.
  218   Fri Jul 2 19:57:12 2021 Shahab, Yuri, Wolfram, LilianaRunsstop DAQ
At 19:56 the DAQ is stopped. 
The plan is to go back to the setting for Se. 
Last files for the Br setting:
- 410MHz-2021.07.02.19.57.36.721.tiq
-245MHz-2021.07.02.19.58.03.216.tiq
  60   Fri May 7 22:48:33 2021 ChrisGeneralstill running...
In anticipation of the imminent shift change, a snapshot of how things look.
  146   Wed Jun 30 02:12:46 2021 ShahabRunsstill no beam in ESR: quadrupole magnets are down
still no beam in ESR: quadrupole magnets are down
  26   Fri May 7 03:33:11 2021 Liliana, Akash, Wolfram, Helmut, ShahabGeneralstatus
At the moment we see a second peak to the left of each intense line. 
The distance between the main peak and the shadow is roughly the same for each peak.
This hints to a sharp fall in the isochronicity curve.
Helmut now tries to remove these shadows by using different scrappers. 
  216   Fri Jul 2 17:31:04 2021 LilianaGeneralstatus
at 17:20 intensity was suddenly reduced. 
Operators checked. Intensity was a bit low and fluctuating for a while (8e8 to 1e9)
Intensity (1.5e9) was recovered at 18:16
  81   Sat May 8 21:10:24 2021 Michael, Ruijiu, Helmut, Yuri, Wolfram, Sergey, ShahabRunsstarted data taking with high resolution
after optimizing the scrapers we have reached the highest possible resolution in the isochronous mode.

starting data taking with this new settings.

last files before this settings were:

245: 2021.05.08.14.14.55.114.tiq
410: 2021.05.08.15.15.10.101.tiq
  139   Tue Jun 29 20:40:55 2021 Rui-Jiu ChenDAQstart NTCAP
Carrier Frequency:245MHz.
IQ rate 20MSamples/s
Reference level:-90 dBm
SC_2021-06-29_20-38-27
IQ_2021-06-29_20-38-27
  184   Thu Jul 1 23:07:47 2021 IK, KKRunssource
The change of the source happens now
  183   Thu Jul 1 21:35:32 2021 ShahabRunssome pics after 15dB amplification on 410
  153   Wed Jun 30 13:57:14 2021 Yuri, SergeyAcceleratorsextupoles + scraper positions
After setting up on 70Se
Sextupoles were adjusted to optimize the resolution of the 70Se peak. Set values are in the ESR logbook.
The SIS energy was shifted up to increase the number of 70Se.
Esis original = 445.68 MeV/u
Esis now = 451.68 MeV/u (+6 MeV/u)

See attached pictures

scrapers: 
GE01DS_HI =  -58mm
GE01DS_HA = +130mm (fully open)
GEEXDD2_S = -0.5mm
GEEXDS2HA = -11.2mm
  173   Thu Jul 1 18:07:10 2021 ShahabRunssettings of the analyzers after changing the trigger position
  85   Sun May 9 01:19:47 2021 Helmut, SergeyAcceleratorscraper optmisation for 72Ge
To improve the isochronicity still in the 72Ge setting sextupoles and scrapers wer varies successfully.

1. E02DS_H -130 / 130mm  - wide rect. peak
            -30 / 130    - nothing 
            -35 / 130    - weak but broad
            -70 / 130    - wider rect peak
            -70 / 30     - two peaks
            -70 / 10     - peak + small peak
            -70 / 0      - peak  + tiny peak
   cannot move more than 0, back to 
            -70 / 50

2. E01DS_H: -130/130    -   wide rectangular distribution
             -40/130    -   wide double structure
             -35/130    -   very weak, still double structure
 some sectupole tuning
             -40/0      - nothing
             -50 / 0    - peak with right tail
  cannot close outside more than 0 so try next scraper

3. E01DS3HG
  -130  - peak with long right tail
   -30  - nothing
   -50  - peak with long right tail
   -40  - weak peak with long right tail
  ==> no good scraper do not use

4. EEXDS2HA
   -120   - peak with long right tail
    -20   - peak with long right tail
      0   - peak with less right tail
     10   - small peak, no right tail
      5   - smaller peak, no right tail
      3   - smaller peak, with small right tail
      8   - small peak no tail
  ==> good scraper use it (+8mm)

5. Try a bit more on E01DS_H
  -55 / 0  - small peak no tail
  -55 / 50 - same no loss in resolution
  -55 /130 - same no loss in resolution
  -65 /130 - same good peak, but higher intensity
  -75 /130 - same good peak, but left tail
  -65 /130 - same good peak, no tail

6. open E02DS_H to -130 / 130 no difference 

7. again E01DS_H
 -125 /130 - very wide left tail
  -68 /130 - one peak
  -72 /130 - one peak, samll left tail
  -66 /130 - one peak, no tail
 ==> good scraper use it (-66mm)

8. check all other scrapers they are all open
   also open TE4 slits to +/-20mm, no difference, back to +/-10 mm

9. bit more sextupole variation on the narrow line

===>  only two scrapers used 

   

picture shows result afterropimisation, the markers indicate the lines,
the left marker is for a +691 keV isomer.








 