HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018
  Laser Cooling Experiment 2021, Page 1 of 5  Not logged in ELOG logo
ID Date Authorup Category Subject Text Attachments
  17   Wed May 12 09:59:08 2021 BenediktGeneralRigol DSG821 signal generator settings   CF7701F6-86BE-406C-8FAD-F3161463181C.jpeg 
  40   Thu May 13 20:18:10 2021 BenediktGeneralNew f_revThe ions were changed to a different speed
today at 2 PM. The new revolution frequency
is now 1.291278 MHz.
  
  41   Thu May 13 21:00:48 2021 BenediktGeneralNew reprate of the UV laserThe lasersystem now operates at a reparte
of 9.038946 MHz. This is 7 times f_rev (see
previous post).
  
  56   Fri May 14 11:01:33 2021 BenediktCalibrationLaser Beam adjustedThe laser beam was adjusted in his position
to overlap better with the ions.
  
  57   Fri May 14 16:37:01 2021 BenediktGeneralStatus updateRight now, we have to adjust the ion beam
a bit to overlap better with the laser beam.
We keep the laser wavelength at resonance
  
  67   Sat May 15 11:06:31 2021 BenediktGeneralNew Zoom link for the momenthttps://tu-darmstadt.zoom.us/j/4948400422?pwd=SEJaUzBuYXM0TTd6UWxoZTFlWEZxUT09

Meeting ID: 494 840 0422
  
  68   Sat May 15 11:26:44 2021 BenediktGeneralTEM parametersWe changed the horizontal position of the
laser beam a bit, so that we maximized the
fluorescence signal on the XUV detector (coasting
  
  69   Sat May 15 12:11:54 2021 BenediktGeneralLaser-bunch-timingOptimizing the laser bunch timing.
h = 7 (9.0389145 MHz)
Reprate is already set to same frequency.
  
  80   Sun May 16 11:29:22 2021 BenediktGeneralAutomated laser scans workingThe automated laser scans are working now.
We can now scan with the following functions:
triangle, sawtooth, cosinus/sinus, arctan,
  
  1   Mon May 10 07:58:13 2021 DanyalGeneralTelephone numbers and Zoom linkTelephone numbers:

ESR laser lab: +49 (0)6159 71 - 2882
  
  2   Tue May 11 15:49:30 2021 DanyalGeneralsetting up ESRThe operators are still busy extracting the
beam from the SIS18 towards the ESR and injecting
it into the ESR. 
  
  5   Tue May 11 18:29:35 2021 DanyalGeneralfirst ions in the ESR   first_ions_are_injected.jpg 
  6   Tue May 11 20:14:43 2021 DanyalGeneralbad ESR vacuumI just heard that the ESR vacuum is quite
bad. 
There was a small vacuum problem a while
  
  8   Wed May 12 02:34:29 2021 DanyalGeneralinstabilities in ion frequencyThe ion beam moves around, pretty arbitrary.
See picture. 
ESR crew believes that it is not due to fluctuations
 fluctuations.jpg 
  20   Wed May 12 13:11:58 2021 DanyalAcceleratorshort break between 13:30 and 15:30There will be a short break (2 hours) this
afternoon. Between 13:30 and 15:30. 
The so-calles "super lense" in the UNILAC
  
  21   Wed May 12 15:36:04 2021 DanyalAcceleratorgeneral power failure in ESR! (everything is off)The power in the ESR is off. 
(also on the gas-target area on top of ESR)
Electrical engineers are trying to find and
  
  25   Wed May 12 22:35:33 2021 DanyalGeneralfrev and fbSchottky frequency (189th harmonic) = 244.033662
MHz
  
  30   Thu May 13 00:58:53 2021 DanyalDAQunbunched ion beam   not_bunched_-_signal.jpgnot_bunched_-_Schottky.jpg 
  31   Thu May 13 01:01:27 2021 DanyalDAQbunched ion beam   bunched_-_signal.jpgbunched_-_Schottky.jpgzoom_bunched_-_Schottky.jpg 
  32   Thu May 13 01:05:45 2021 DanyalDAQmore stability by bunchingBecause the C3+ part of the ion beam is properly
bunched, it is no longer so sensitive to
the drifts caused by the e-cooler. 
  
ELOG V3.1.5-fc6679b