85
Mon Sep 14 09:15:03 2020
Cristina
test without bakeout
BPM SIS100 2010-17
09.09.20
DONE
Accepted
09.09.2020
Mobile 3
10.09.2020
Cond. Accep.
17.09.2020
M. Müller
86
Mon Sep 14 09:16:54 2020
Cristina
test with bakeout
Mirror Laser cooling SIS100
14.09.2020
DONE
Accepted
6.10.20
Mobile 2
21.10.20
Cond. Accep.
KW47
D.Winters
87
Mon Sep 14 10:38:59 2020
Cristina
test without bakeout
Test on Viton gasket
14.09.2020
TO DO
Not Needed
88
Tue Sep 15 11:14:44 2020
Cristina
leak check
SIS100 Dipole chambers
on-going
July 2019
Customised
07-00006-001-0; 07-00006-002-7; 07-00006-003-4; 07-00006-004-1; 07-00006-005-8; 07-00006-006-5; 07-00006-007-2; 07-00006-009-6; 07-00006-010-2; 07-00006-011-9; 07-00006-012-6; 07-00006-013-3; 07-00006-014-0; 07-00006-016-4 (mit RGA); 07-00006-017-1; 07-00006-018-8; 07-00006-019-5; 07-00006-020-1 (mit RGA); 07-00006-022-5; 07-00006-023-2; 07-00006-024-9; 07-00006-026-3; 07-00006-028-7; 07-00006-033-1; 07-00006-048-5; 07-00006-036-2;07-00006-049-2; 07-00006-042-3; 07-00006-046-1;07-00006-040-9;07-00006-053-9;07-00006-039-3; 07-00006-050-8; 07-00006-047-8;07-00006-015-7; 07-00006-021-8;07-00006-025-6;07-00006-027-0;07-00006-034-8;07-00006-035-5;07-00006-049-2; 07-00006-046-1; 07-00006-047-8; 07-00006-045-4; 07-00006-044-7; 07-00006-055-3; 07-00006-054-6; 07-00006-051-5; 07-00006-059-1; 07-00006-041-6; 07-00006-065-2; 07-00006-038-6;07-00006-030-0; 07-00006-037-9; 07-00006-031-7;07-00006-032-4; 07-00006-056-0; 07-00006-057-7; 07-00006-052-2; 07-00006-069-0; 07-00006-060-7; 07-00006-072-0; 07-00006-058-4; 07-00006-067-6;07-00006-070-6; 07-00006-066-9;07-00006-062-1
90
Mon Sep 21 11:31:47 2020
Cristina
test without bakeout
HEBT BPM
KW39
On schedule
28.09.20
on-going
Accepted
05.10.20. Test started again: 08.12.20
Mobile 3
Refused
Since 19. Okt we wait for the decision of the diagnostic group on how(who has to clean the BPM. Since then the test is stopped. Meeting on the 2.12.20: The japanise company will clean a BPM and they will send us to be tested. Email from Dorn on the 07.12.20: we have to test the other 3 BPM, and later they will take care of the cleaning of the refused BPM. On the 7.12 we tested another BPM: dirty. We sent on the 14.12 an email to Dorn and Kowina, showing that the ceramic are dirty and that they make dirty also our reference chamber. Measurement stopped again.
91
Mon Sep 21 11:32:46 2020
Cristina
test with bakeout
Ferrite RI after firing
18.09.20
DONE
Accepted
22.09.2020
Setup 1
1.10.20
Accepted
---------
---------
92
Mon Sep 21 11:39:27 2020
Cristina
test without bakeout
Steel with epoxy glue
August 20
Delayed
21.09.20
DONE
22.09.2020
03.10.20
Refused
?
?
93
Mon Sep 21 14:26:53 2020
Cristina
leak check
Beam line CWT
21.08.20
21.09.20
Customised
23.09.2020
Accepted
24.09.20
S.Wilfert
95
Wed Sep 30 13:39:48 2020
Cristina
test with bakeout
Tube CCC Cryring
30.09.2020
DONE
Accepted
5.10.20
Setup 1
07.10.20
Accepted
8.10.20
T. Sieber
96
Tue Oct 13 07:54:31 2020
Phe
test without bakeout
Epoxy - baked in the oven @ 80 deg C
09.10.2020
On schedule
09.10.2020
DONE
Not Needed
22.10.20
Mobile 2
29.10.20
Refused
97
Wed Oct 21 10:51:47 2020
Cristina
test with bakeout
IPM SIS18 mit diagnostic parts
25.09.2020
DONE
Refused
20.10.20
29.10.20
Cond. Accep.
2.11.20
Grossmontage
98
Tue Nov 3 09:55:20 2020
Cristina
test with bakeout
MCPs for IPM SIS18
September
Delayed
2.11.20
DONE
Accepted
2.11.20
Setup 1
09.11.2020
Cond. Accep.
10.11.2020
T.Giacomini
99
Tue Nov 17 09:43:55 2020
Cristina
test without bakeout
BPM 2010-16
?
DONE
Accepted
16.11.20
Mobile 3
17.11.20
Cond. Accep.
01.12.20
Grossmontage
100
Thu Nov 19 10:15:23 2020
Cristina
Coating
Parts for Cryring
Beginning October 20
DONE
Accepted
16.11.20
Horizontal Coating Setup
17.11.20
Accepted
26.11.20
David (Cryring)
There are some parts which were not coated: 2 bellows (the Cryring people still did not receive them from the company) and a T-piece, which was brought to the cleaning on the 20 of October and it is not yet ready.
101
Thu Nov 19 10:17:06 2020
Cristina
Coating
Coating chamber Glorius
16.11.20
DONE
Accepted
27.11.20
Horizontal Coating Setup
3.12.20
Accepted
14.12.20
Glorius
102
Mon Nov 30 11:14:30 2020
Cristina
Coating
T-piece Cryring
Oct
KW48
DONE
Accepted
26.11.20
Horizontal Coating Setup
3.12.20
Accepted
10.12.20
Nikita
103
Mon Nov 30 11:15:48 2020
Cristina
test with bakeout
Antrieb for ESR-Beata
1 Dec.
Delayed
?
3.12.20
DONE
Not Needed
3.12.20
Setup 1
10.12.2020
Accepted
11.12.20
Müller
104
Mon Nov 30 11:17:49 2020
Cristina
leak check
5 x SIS100 Diagnostic Chambers
KW 49
DONE
Not Needed
01.12.20
16.12.20
Accepted
06-00401-099-1 (abgeholt on the 07.12 M.Müller); 06-00401-100-4 (abgeholt 09.12.20 Müller); 06-00401-101-1; 06-00401-103-5
105
Tue Dec 1 11:49:07 2020
Cristina
test with bakeout
Bellows IPM SIS18-Second measurement after cleaning
12.10.20
DONE
Not Needed
12.10.20
HEBT Setup
14.10.20
Accepted
15.10.20
Zentrale Dienst
106
Tue Jan 12 12:48:45 2021
Cristina
test without bakeout
HEBT BPM-Cleaned @Kyocera
Beginning 21
On schedule
12.01.21
DONE
Accepted
12.01.21
Mobile 3
14.01.21
Accepted
19.01.21
M.Müller