  113   Tue Jun 6 09:33:00 2023 RodolfoLaserFiber coupler
Cobra's new fiber-coupling unit for wavelength reading.
Attachment 1: Cobra_FiberCoupler.jpg
Cobra_FiberCoupler.jpg
  37   Wed May 11 03:17:14 2022 Patrick, Volker, Guy RunsStart laser scanning
Runs of the night shift. The wavelength range is 601.95 - 601.735 nm.
We started the wavemeter python script on atppc023 located in Desktop/wavemeter_Bi.
We changed the wavelength range in go4 to the actual scan range.

Questions:
There are many other masses/isotopes/etc. visible in the Shottky spectrum. Is this wanted?
Are the vacuum gauges turned off?
Laser timing correct? +- 3 bins?

TimeSIS-intensitycooler-HVLMD-DAQ2 file nr LMD-DAQ1 file nr scan range(nm)step width(nm)dwell time(s)comments
03:30 ?e7 223218 373 8073 601.95 - 601.735 0.005 5s 2 loops
03:38 9.0e7 223218 374-377 8074 601.95 - 601.735 0.005 5s 8 loops from now on
04:15 8.4e7 223218 378 - 381 8075 601.95 - 601.735 0.005 5s
04:49 7.4e7 223217 382 - 385 8076 601.95 - 601.735 0.005 5s
05:23 6.4e7 223216 386 - 388 8078 601.95 - 601.735 0.005 5s 8077 was skipped due to some "server socket locked"-error
05:59 8.9e7 223218 389 - 391 8079 601.95 - 601.735 0.005 5s
  39   Wed May 11 06:41:32 2022 Simon, Christopher, ShahabRunsshifts
new shift started.
  40   Wed May 11 06:47:58 2022 Simon, Christopher, ShahabRunsrestart laser scan

Time SIS-intensity ESR-intensity after orbit change cooler-HV LMD-DAQ2 file nr LMD-DAQ1 file nr scan range(nm) step width increment (nm) dwell time(s) comments
06:37 7.8e7 ? 223217 392-395 8080 601.95 - 601.735 0.005 5s 8 loops
07:10 4.8e7 ? 223217 396-399 8081 601.95 - 601.735 0.005 5s 8 loops
07:48 1.3E8 ? 223219 400-403 8082 601.95 - 601.735 0.005 5s 8 loops
08:25 3.9E7 ? 223219 404- 406 8083 601.95 - 601.735 0.005 5s 8 loops
08:59 5.8E7 ? 223219 407 8084 601.95 - 601.735 0.005 5s 8 loops
09:38 7.2E7 3.1E6 223219 408-411 8085 601.95 - 601.735 0.005 5s 8 loops
10:24 6.4E7 3.1E6 223219 412-415 8086 601.95 - 601.735 0.005 5s 8 loops
11:05 7E7 2.2E6 223219 416-418 8087 601.95 - 601.735 0.005 5s 8 loops
11:36 6E7 2.2E6 223219 419-420 8088 601.95 - 601.735 0.005 5s number of loops now 4
11:56 2E8 6.9E6 223219 421-423 8089 601.62 - 601.405 0.005 5s 4 loops, changed scan range
12:28 1.1E8 5E6 223219 424-426 8090 601.62 - 601.405 0.005 5s 4 loops
12:43 6.5E7 2.6E6 223219 427-429 8091 601.62 - 601.405 0.005 5s 4 loops
13:11 1.3E8 5.8E6 223219 430-432 8092 601.62 - 601.405 0.005 5s 4 loops
  42   Wed May 11 08:01:44 2022 ZoranRunsStart laser scanning

Quote:

Questions:
There are many other masses/isotopes/etc. visible in the Shottky spectrum. Is this wanted?
Are the vacuum gauges turned off?
Laser timing correct? +- 3 bins?


No, this is not wanted and yes, the machine is aware of the problem. The decision on Tuesday evening was to leave the ESR in that state in order to start the measurement and continue setting on Wednesday. Otherwise the night shift could not take place.

The additional isotopes are only a few percent and should contribute insignificantly to the background rate.
  53   Thu May 12 00:16:13 2022 Patrick, Volker, Guy, Rui-JiuRunsRuns night shift

Time SIS-intensity ESR-intensity after orbit change cooler-HV LMD-DAQ2 file nr LMD-DAQ1 file nr scan range(nm) step width increment (nm) dwell time(s) n loops comments
00:15 ~1.0e8 1.0e7 223219 433 - 435 8094 601.62 - 601.405 0.005 5s 4
00:47 3.7e8 2.0e7 223219 436 - 440 8095 - 8096 601.62 - 601.405 0.005 5s 8 increased to 15 injections, still not saturated
Tried 20 injections, got no injection into ESR, so moved back to 15. The scrapers were not moving for quiet some time, even after fixing them. The problem may have been that two ParamModi windows were opened. The injection into the SIS does not work when there are other users performing special operations with the machines.
02:28 1.0e8 1.0e7 223220 441 - 444 8097 601.62 - 601.405 0.005 5s 6
02:54 2.8e8 1.7e7 223219 445 - 449 8098 - 8099 601.62 - 601.405 0.005 5s 6
03:21 ~3e8 ~2e7 223219 DAQ did not write 8100 - 8101 601.62 - 601.405 0.005 5s 6
The vacuum gauge of the e-cooler failed. It was restarted. Apparently, upon changing/fixing anything related to our accelerator pattern, the scrapers forget to move. Then they have to be disabled and enabled again in ParamModi. Also, the injection does not work very well. Either the transition from SIS to ESR fails or we are just skipped completely. After fixing the injection, again the scrapers failed as expected. We keep the scrapers out now.
04:34 2.8e8 1.8e7 223219 DAQ failed, started late 454 - 458 8102 - 8103 601.62 - 601.405 0.005 5s 6 Set max.counts in PMT middle and north to 15 (In file 'SetConditions.C', applied when starting go4). No scrapers used.
05:02 2.5e8 1.7e7 223219 DAQ late, 461 - 464 8104 - 8105 601.62 - 601.405 0.005 5s 6
05:30 2.5e8 1.8e7 223219 465 - 469 8106 - 8107 601.62 - 601.405 0.005 5s 6 Set max.counts in PMT middle and north to 20.
05:57 2.5e8 1.7e7 223219 470 - 474 8108 - 8109 601.62 - 601.405 0.005 5s 6
06:25 2.6e8 1.7e7 223219 475 - 479 8110 - 8111 601.62 - 601.405 0.005 5s 6
06:50 2.4e8 1.6e7 223219 480 - 484 8112 - 8113 601.62 - 601.405 0.005 5s 6
07:24 2.2e8 1.5e7 223219 485 - 489 8114 - 8115 601.62 - 601.405 0.005 5s 6
07:49 - 6.5e6 223219 490 - 493 8116 601.62 - 601.405 0.005 5s 6 UNILAC tunnel will be open for ~1h -> no new beam from SIS available
08:15 - 2.6e6 223219 494 - 496 8117 601.62 - 601.405 0.005 5s 6 UNILAC tunnel will be open for ~1h -> no new beam from SIS available
  73   Mon May 16 18:01:23 2022 mbsMax, Danyal, Wilfried, Rodolfo, Patrick, Kristian, David, Ruijiu Runsruns afternoon/night shift 16.05
These are the runs after the septum was repaired on 16.05.

Time ESR-intensity ESR-intensity after orbit change cooler-HV LMD-DAQ2 file nr LMD-DAQ1 file nr scan range(nm) step width increment (nm) dwell time(s) n loops comments
ca. 18:00 ? ~3.0e7 223222 504-505 8121 601.62 - 601.405 0.005 5s 4 DAQ2 late start
18:11 ? ~3.3e7 223222 506 8122 601.62 - 601.405 0.005 5s 4 aborted, ions are rapidly lost after Bahnverlagerung
18:33 4.5e7 ~3.7e7 223222 507-509 8123 601.62 - 601.405 0.005 5s 4 removed all scrapers, beam life time is better, high intensity, slight instability in bunching, ion intensity at the end: 1.4e7, DC Trafo was in saturation so ampl. was turned down
19:32 ? 3.7e7 223222 512 8125 601.62 - 601.405 0.005 5s 4 rf bunching amplitude decreased from 350V -> 280V
20:00 ? 3.6e7 223223 514 8127 601.745 - 601.32 0.01 5s 4 43 steps of laser, therefore 1 step offset from loop to loop, laser already of ?
20:30 ? 2.9e7 223223 520-522 8129 601.760 - 601.320 0.01 5s 6 aborted after 3 loops, laserstabi was off, probably during hole measurement
21:01 ? 3.6e7 223223 524-528 8130 601.760 - 601.320 0.01 5s 6
21:30 ? 3.4e7 223223 529-532 8131 601.760 - 601.320 0.01 5s 6 coasting beam now !!!
22:10 ? 3.7e7 223223 534-537 8132 601.760 - 601.320 0.01 5s 6 Night shift took over.
22:40 4.4e7 3.6e7 223223 539-542 8133 601.760 - 601.320 0.01 5s 6
23:52 3.8e7 3.2e7 223223 545-548 8134 - 8135 601.760 - 601.320 0.01 5s 6 Switched off the nitrogen before this run, stopped acq. after 4 loops (laser gone), PMT north saw a lot of noise? (ID:78) -> larger files.
00:39 4.3e7 3.2e7 223222 552-555 8136 601.760 - 601.320 0.01 5s 6 The secondary DAC always requires 2-3 restarts until it is working.
01:10 3.8e7 3.0e7 223222 557-558 8137 601.760 - 601.320 0.01 5s 6 Laser moved away during the third loop, acq. stopped.
01:33 4.2e7 3.1e7 223222 561-564 8139 601.760 - 601.320 0.01 5s 6 The previous laser scan was still running, restarted the acq. once.
02:00 > 4.0e7 3.4e7 223222 566-569 8140 601.760 - 601.320 0.01 5s 6
02:30 > 4.0e7 3.2e7 223222 573-576 8141 601.760 - 601.320 0.01 5s 6
02:58 4.2e7 3.4e7 223222 579-582 8142 601.760 - 601.320 0.01 5s 6
03:45 4.2e7 3.3e7 223222 585-589 8143 601.760 - 601.320 0.01 5s 6 Back to bunched beam mode
04:15 4.7e7 3.4e7 223222 594-598 8144 601.760 - 601.320 0.01 5s 6
04:43 4.4e7 3.4e7 223222 600-604 8145 601.960 - 601.520 0.01 5s 6 We changed the scan range Big grin
05:10 4.3e7 3.3e7 223221 606-610 8146 601.960 - 601.520 0.01 5s 6 It looks like the "overWavel" plots actually work better now, because the laser steps somehow are counted wrongly, leading to shifts between loops.
05:37 4.2e7 3.7e7 223221 611-615 8147 601.960 - 601.520 0.01 5s 6
06:27 ? ? 223219 616-618 8148 601.960 - 601.520 0.01 5s 6 Since Go4 was still running during startup we were not able to look into the data during the scan. After reduction of particle number, the dc trafo is no longer capable to measure to number of ions.
06:56 ? ? 223219 619-620 8149 601.960 - 601.520 0.01 5s 6
07:25 ? ? 223219 621-622 8150 601.9 - 601.7 0.01 5s 6 Changed the scan range t0 601.9-601.7 and reduced step increment to 0,005nm
07:56 ? ? 223219 623-624 8151 601.9 - 601.7 0.01 5s 6
08:20 ? ? 223219 625-625 8152 601.9 - 601.7 0.01 5s 6 Restarted scan without new injection, DAQ2 was not running during this scan
08:44 ? ? 223220 626-627 8153 601.9 - 601.7 0.01 5s 6 Reduced Loop number to 4 (aborted after 4 completed scans)
09:03 ? ? 223220 628 8154 601.9 - 601.7 0.01 5s 6 Reduced Loop number to 3 (before starting a new scan)
09:17 ? ? 223221 629 8155 601.9 - 601.7 0.01 5s 6
Since
  86   Mon May 30 02:12:20 2022 Max, Patrick, ~Konstantin, Ragan, KenRunsruns night shift 30.05.
These are the first runs of the substitute beamtime.

Time ESR-intensity ESR-intensity after orbit change cooler-HV LMD-DAQ2 file nr LMD-DAQ1 file nr scan range(nm) step width increment (nm) dwell time(s) n loops comments
02:00 ? 8.0e6 223217 632 8158 601.95 - 601.734 0.005 5s 5 Resonance found in first scan. The resonance is much broader due to a broader ion beam.
02:28 ? 7.0e6 223217 633 8159 601.9 - 601.7 0.005 5s 5 Same injection.
02:52 ? 6.3e6 223217 634 8160 601.9 - 601.7 0.005 5s 5 Same injection. Adjusted gates in go4.
03:14 ? 5.6e6 223439 637 8163 601.9 - 601.7 0.005 5s 5 Same injection. Auto-impedance enabled -> correct voltages. NTCAP started shortly after scan start.
03:35 ? 3.0e5 ? (under range) 223439 639 8164 601.9 - 601.7 0.005 5s 5 Scraped the lower charge state beam by moving S1HA to pos. -30 (from the left).
04:50 ? 1.1e7 223439 640 8165 601.9 - 601.7 0.005 5s 5 Laser beam position slightly corrected for new ion beam position, see ID:84.
05:05 ? 1.0e7 223439 641 8166 601.9 - 601.7 0.005 5s 5 See picture (this + previous run)
05:21 ? 1.0e7 223439 642 8167 601.9 - 601.7 0.005 5s 5 Moved S1HA to -35 to get rid of the charge transferred beam.
05:33 ? 9.0e6 223439 644 8169 601.9 - 601.7 0.005 5s 5
05:52 ? 9.0e5 223439 645 8170 601.9 - 601.7 0.005 5s 5 Scraped main beam for lower beam intensities.
There were some issue with the acquisition, and thus there are some empty files
06:08 ? 4.e5 223438 650- 8172 601.9 - 601.7 0.005 3s 25
07:11 ? 1.15e7 223438 653 8173 601.9 - 601.7 0.005 3s 25 laser scan was not started
07:17 ? 1.03e7 223438 654-656 8174 601.9 - 601.7 0.005 3s 25
08:33 ? 6e6 223438 658-660 8176 601.9 - 601.7 0.005 3s 25 ESR scraper not in
09:27 ? 5.5e6 223438 661-667 8178 601.9 - 601.7 0.005 3s 25 XUV detector running again, signal of PMT north and middle in att. 2
10:15 ? <<1e6 223438 668 8179 601.9 - 601.7 0.005 3s 25 very low beam intensity


Note: PMT north does not see a resonance.

Note: PMT north does see a resonance after changing to the broadband PMT.
Attachment 1: 2022-05-30_resonance_10_loops.png
2022-05-30_resonance_10_loops.png
Attachment 2: 48_AM.png
48_AM.png
  90   Mon May 30 16:17:39 2022 Kristian, David, Wilfried, Carsten, Danyal, RodolfoRunsAfternoon shift 30.05
These are the first runs after changing to the thick stripper foil.

Time ESR-intensity ESR-intensity after orbit change cooler-HV LMD-DAQ2 file nr LMD-DAQ1 file nr scan range(nm) step width increment (nm) dwell time(s) n loops comments
16:15 ? 5.0e6 223439 669-670 8180 601.95 - 601.734 0.005 5s 5 Resonance found in first scan. electron cooling current 100mA
16:45 ? 5.0e6 223439 671-672 8181 601.95 - 601.734 0.005 5s 5 electron cooling current 300mA


Width approximately the same with the higher electron current.
next we go to 208Bi
  96   Mon May 30 23:18:06 2022 Carsten, Rodolfo, Konstantin, Patrick, Ragan, Ken, Kristian, Max, DanyalRunsRuns shifts 31.05.
These are the first runs with 208Bi.

Time SIS-intensity ESR-intensity after orbit change cooler-HV LMD-DAQ2 file nr LMD-DAQ1 file nr scan range(nm) step width increment (nm) dwell time(s) n loops Particle Detector comments
23:11 2.6e8 - 225259 673-677 8182 548 - 549 0.005 2.97s 4 139e3 Resonance found in first scan.
00:09 ? - 225259 678-679 8183 548 - 549 0.005 2.97s 2 92e3
00:45 2.8e8 - 225260 680-681 8184 548 - 549 0.005 2.97s 2 36e3 In this run, the ions seemed to be not ejected with the new request, compare picture.
01:27 2.7e8 - 225260 684-685 8187 548 - 549 0.005 2.97s 2 88e3 Started delayed due to several issues with both DAQs.
01:53 2.7e8 - 225260 686-687 8188 548 - 549 0.005 2.97s 2 93e3
02:20 2.7e8 - 225260 688-689 8189 548 - 549 0.005 2.97s 2 87e3
02:55 2.1e8 - 225260 692-694 8190 548.335 - 548.535 0.0025 4.95s 4 88e3 Scan range changed
03:30 2.75e8 - 225260 696-697 8191 548.335 - 548.535 0.0025 4.95s 3 95e3 Number of loops reduced to 3
03:59 2.6e8 - 225260 698-699 8192 548.335 - 548.535 0.0025 4.95s 3 89e3
04:30 2.7e8 - 225260 700-702 8194 548.335 - 548.535 0.0025 4.95s 3 89e3 DAQ1 needed to be restarted
04:52 2.55e8 - 225260 703-704 8195 548.335 - 548.535 0.0025 4.95s 3 83e3
05:18 2.62e8 - 225260 706-707 8196 548.335 - 548.535 0.0025 4.95s 3 89e3 DAQ2 needed to be restarted
05:44 ? - 225260 708-710 8197 548.335 - 548.535 0.0025 4.95s 3 ?
06:18 2.1e8 - 225260 711-712 8200 548.36 - 548.51 0.0025 4.95s 3 67e3 we changed the timing by 20 ns. The laser pulse is later now. The file number of DAC1 is correct. I do not know why there is the gap to the previous scan
06:51 2.4e8 6e5 225260 713-714 8201 548.36 - 548.51 0.0025 4.95s 3 67e3 xuv-detector parameters were varied
07:09 2.4e8 6e5 - 715- 8202 548.36 - 548.51 0.0025 4.95s 3 68e3 mbs did not start
07:15 2.4e8 6e5 225260 716-717 8203 548.36 - 548.51 0.0025 4.95s 3 68e3 xuv-detector walklevel and treshhold were lowered
07:35 2.4e8 6e5 225259 718-719 8204 548.36 - 548.51 0.0025 4.95s 3 68e3 we changed the timing by additional 20 ns. The laser pulse is later now. - no fluo. visible
08:01 2.4e8 6e5 225259 720-721 no file opened 548.43 - 548.44 0.00001 ~20s 3 71e3 varying the delay with each laser step to find optimum - It was more a plateau between +-30ns (relative to the night shift's position). We now continue data taking at the same dealy as used by the night shift.
08:23 2.4e8 6e5 225258 722 8205 548.36 - 548.51 0.0025 4.95s 3 68e3 xuv-detector walklevel and treshhold were lowered
08:28 2.4e8 6e5 225258 724-725 8208 548.36 - 548.51 0.0025 4.95s 3 68e3
08:48 2.4e8 6e5 225258 727-728 8209 548.36 - 548.51 0.0025 4.95s 3 ? xuv-detector in Park-position just as a try
09:08 2.4e8 6e5 225259 729-730 8210 548.36 - 548.51 0.0025 4.95s 3 ?
09:28 2.5e8 6e5 225259 731-732 8211 548.36 - 548.51 0.0025 4.95s 3 73e3
09:54 2.5e8 6e5 225259 734-735 8212 548.36 - 548.51 0.0025 4.95s 3 175e3
10:19 2.5e8 6e5 225258 737-738 8214 548.36 - 548.51 0.0025 4.95s 3 175e3
10:38 2.5e8 6e5 225258 741-742 8215 548.36 - 548.51 0.0025 4.95s 3 64e3 DAQ2 started late. Missed the first 25 scan numbers Coasting beam measurement from now on
10:57 2.5e8 6e5 225258 743-744 8216 548.36 - 548.51 0.0025 4.95s 3 87e3
11:17 2.5e8 6e5 225258 745 8217 548.36 - 548.51 0.0025 4.95s 3 e3 MBS crashed
11:18 2.5e8 6e5 225257 746-747 8218 548.36 - 548.51 0.0025 4.95s 3 87e3
11:37 2.5e8 6e5 225257 749-750 8219 548.36 - 548.51 0.0025 4.95s 3 87e3 DAQ2 started late. Missed the first 10 scan numbers
11:56 2.5e8 6e5 225257 751 8220 548.36 - 548.51 0.0025 4.95s 3 78e3
12:59 2.5e8 6e5 225257 752 8221 548.385 - 548.485 0.0025 4.95s 3 78e3 RF off from now on, laser scan settings not correct. stopped
13:06 2.5e8 6e5 225257 753 8223 548.385 - 548.485 0.0025 4.95s 3 87e3
13:26 2.5e8 6e5 225257 754 8224 548.385 - 548.485 0.0025 4.95s 3 - MBS crashed
13:29 2.5e8 6e5 225257 755-756 8226 548.385 - 548.485 0.0025 4.95s 3 93e3 RF now really off (wrong cavity was turned off)
13:43 2.5e8 6e5 225257 757-758 8227 548.385 - 548.485 0.0025 4.95s 3 67e3
14:08 2.5e8 6e5 225257 759 8228 548.385 - 548.485 0.0025 4.95s 3 70e3
14:22 2.2e8 6e5 225257 760 8229 548.385 - 548.485 0.0025 4.95s 3 56e3
14:35 2.2e8 6e5 225257 761 8230 548.385 - 548.485 0.0025 4.95s 3 64e3
14:50 2.2e8 6.5e5 225256 762 8231 548.385 - 548.485 0.0025 4.95s 3 59e3
15:04 2.0e8 7e5 225255 763 8232 548.385 - 548.485 0.0025 4.95s 3 58e3
15:28 1.9e8 ?e5 225272 764 8233 548.385 - 548.485 0.0025 4.95s 2 66e3 changed to 200mA cooler current, NTCAP stopped and started again folder: prob. ~31.5.22 13:??, loops decreased to 2
15:37 2e8 6e5 225272 765 8234 548.385 - 548.485 0.0025 4.95s 2 70e3
15:47 2.5e8 6e5 225272 766 8235 548.385 - 548.485 0.0025 4.95s 2 73e3 laser gone at loop 2 scan 35 (15:56), Ecooler went also off
16:17 ?e8 6e5 225272 767 8236 548.385 - 548.485 0.0025 4.95s 2 48e3 bad shot?, probably ecooler reset
16:27 2.2e8 6e5 225271 768 8237 548.385 - 548.485 0.0025 4.95s 2 77e3
16:36 2.0e8 7e5 225272 769 8238 548.385 - 548.485 0.0025 4.95s 2 65e3
16:52 ?e8 ?e5 225272 770 8239 548.385 - 548.485 0.0025 4.95s 2 61e3 we had to restart go4, it was very slow
17:02 2.0e8 7e5 225272 771 8240 548.385 - 548.485 0.0025 4.95s 2 63e3
17:16 2.3e8 7e5 225248 772-773 8241 548.385 - 548.485 0.0025 4.95s 4 45e3 changed to 50mA cooler current, NTCAP stopped and started again folder: 31.5.22 15:18, loops increased to 4
17:36 2.7e8 7.5e5 225248 774-775 8242 548.385 - 548.485 0.0025 4.95s 6 67e3 loops increased to 6
18:05 2.5e8 7.5e5 225248 776-777 8243 548.385 - 548.485 0.0025 4.95s 6 66e3
18:31 1.7e8 6e5 225247 778-779 8244 548.385 - 548.485 0.0025 4.95s 6 44.6e3 SIS pattern looked strange
Attachment 1: 2022-05-31_particles.png
2022-05-31_particles.png
  105   Wed Jun 1 00:03:08 2022 Konstantin, Simon, KristianRuns209-Bi runs
Time SIS-intensity ESR-intensity after orbit change cooler-HV cooler current (mA) LMD-DAQ2 file nr LMD-DAQ1 file nr scan range(nm) step width increment (nm) dwell time(s) n loops comments
23:40 - - 223432 100 781 8246 601.735 - 601.95 0.005 4.95s 2 resonance found but problems with go4
00:02 - 5.5e6 223432 100 783 8248 601.735 - 601.95 0.005 4.95s 2 resonance not in the middle of the scan range
00:14 - 4.5e6 223432 100 784 8249 601.7 - 601.9 0.005 4.95s 2 change the scan range
00:14 - 1.1e7 223432 100 785 8250 601.7 - 601.9 0.005 4.95s 2
00:35 - 9e6 223432 100 786-787 8251 601.7 - 601.9 0.005 4.95s 4
00:50 - 6.5e6 223432 100 788-789 8252 601.7 - 601.9 0.005 4.95s 4
01:07 2.2e8 1.15e7 223432 100 790-791 8253 601.7 - 601.9 0.005 4.95s 4
01:26 2.3e8 1.15e7 223448 200 792 8254 601.7 - 601.9 0.005 4.95s 2 switched to coasting beam, changed ESR current
01:37 2.2e8 1.15e7 223448 200 793 8255 601.7 - 601.9 0.005 4.95s 2
01:48 2.2e8 1.15e7 223448 200 794 8256 601.7 - 601.9 0.005 4.95s 2
01:58 2.2e8 1.15e7 223448 200 795 8257 601.7 - 601.9 0.005 4.95s 2
02:08 2.2e8 1.15e7 223448 200 796 8258 601.7 - 601.9 0.005 4.95s 2
02:19 2.15e8 1.15e7 223432 100 797-799 8259 601.7 - 601.9 0.005 4.95s 6 changed ESR current again
02:45 2.17e8 1.15e7 223432 100 800-802 8260 601.7 - 601.9 0.005 4.95s 6
03:12 2.12e8 1.15e7 223423 50 803-805 8261 601.7 - 601.9 0.005 4.95s 6 changed cooler current again, restarted NTCAP folder 2022-05-31_19-33-57
03:38 2.2e8 1.15e7 223423 50 806-808 8262 601.7 - 601.9 0.005 4.95s 6
04:05 2.13e8 1.15e7 223427 75 809-811 8263 601.7 - 601.9 0.005 4.95s 6 changed cooler current again, restarted NTCAP folder 2022-06-01_03-12-03
04:30 2.03e8 1.15e7 223427 75 812-814 8264 601.7 - 601.9 0.005 4.95s 6
04:57 2.08e8 1.15e7 223440 150 815-817 8265 601.7 - 601.9 0.005 4.95s 6 changed cooler current again, restarted NTCAP folder 2022-06-01_04-04-32
05:21 2.1e8 1.15e7 223440 150 818-820 8266 601.7 - 601.9 0.005 4.95s 6
05:57 2.08e8 1.15e7 223440 150 821-823 8267 601.7 - 601.9 0.005 4.95s 6 laser power reduced from 46.2mJ to 25mJ
06:25 1.5e8 9e6 223440 150 824-826 8268 601.7 - 601.9 0.005 4.95s 6
06:56 2.1e8 1.15e7 223440 150 827-829 8269 601.7 - 601.9 0.005 4.95s 6 Laser power reset to 48mJ. Last ion injection!!!
07:18 - 4.7e6 223439 150 830-832 8270 601.7 - 601.9 0.005 4.95s 6 Last ion injection!!!
7:40 - 2.0e6 223439 150 833-834 8271 601.7 - 601.9 0.005 4.95s 6 old beam
8:02 - 1.0e6 223439 150 835-836 8272 601.7 - 601.9 0.005 4.95s 6 older beam
8:25 - 5e5 223439 150 837-8 8273 601.7 - 601.9 0.005 4.95s 6 very old beam
8:48 - 3e5 223439 150 839-840 8274 601.7 - 601.9 0.005 4.95s 6 very very old beam
9:10 - 2e5 223439 150 841-842 8275 601.7 - 601.9 0.005 4.95s 6 beam close to retirement - no signal anymore


9:33 beam retired. Turned laser off. Stopped NTcap

post beamtime change: replaced "ESR current" by "cooler current" in the comments column. (Wilfried)