HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018 E0028 E0038 E0075 E0203_E0018 E0052_2025
  Nuclear hyperfine mixing in 229Th89+ 2025, Page 3 of 4  Not logged in ELOG logo
New entries since:Thu Jan 1 01:00:00 1970
ID Date Author Category Subject Text Attachments
  26   Wed Jun 18 19:05:28 2025 Carlo, Zac, Shahab, RaganAnalysisPID with rionidFor PID with rionid we used: 
- LISE file from last year (attached.
- .npz file from 410 RSA (attached).
 Schermata_2025-06-18_alle_19.03.04.pngE142_TEline-ESR_229Th.lppRSA01-2025.06.18.18.06.02.162.tiq_spectrum.npzSchermata_2025-06-18_alle_19.32.53.png 
  25   Wed Jun 18 14:01:40 2025 Julien, Konstantin, Stefan, Rima Ion bunch on PMTsApparentlty did not work yesterday evening
(no signal), but looks fine now (see attachment)
 Bildschirmfoto_bunch_on_PMT_2025-06-18_14-11-44.png 
  24   Wed Jun 18 11:44:15 2025 Julien, Konstantin, Stefan, Rima Jitter CFDPulses on photodiodes in entry 20 

1. seem wide, so we also looked at qswitch
  
  23   Wed Jun 18 08:07:17 2025 Julien, KonstantinLasernew FCU CalibrationUV power was not stable but dropping from
~18 mW to 13 mW after opening COBRA shutters.
New calibration of frequency conversion unit:
 2025_06_18_FCU_Cal_Table.JPG2025_06_18_FCU_Cal_Plot.JPG 
  22   Wed Jun 18 07:59:00 2025 Julien, KonstantinLaserPhase-offsetWe recognized that the output power of the
pump laser was instable when externally triggered
with 200kHz (stand-bye frequency of ESR).
 IMG_3889.jpgIMG_3888.jpgIMG_3891.jpgIMG_3892.jpg 
  21   Wed Jun 18 03:12:25 2025 RodolfoLaserBeam StabilizationThe laser beam pointing has changed, even
thought the MRC settings are the same as
given in Entry 5 (this elog). This could
 MRC-Settings_2025_06-18.JPG 
  20   Wed Jun 18 01:58:02 2025 Esther, Alexis, Anton, Rodolfo, CarstenAcceleratorNo Bunched BeamThe Bunching is not set properly. We do not
see the ion bunch with the Photomultiplier.
 Bildschirmfoto_PD_PMTSouth_2025-06-18_1.png 
  19   Wed Jun 18 00:56:28 2025 Carsten, Rodolfo, Anton, Alexis, Esther beam overlap with t-scrapersWhen testing the overlap between laser and
ion beam we observed the following positions:
  
  18   Wed Jun 18 00:33:56 2025 Esther, Alexis, Anton, Rodolfo, CarstenDAQR4l-41We have some issues with the RIO computer
R4l-41 (DAQ1). We had to switch off completely
twice. Only then it reacted properly. Maybe
  
  17   Tue Jun 17 19:25:30 2025 RodolfoAcceleratorT-Scrapers in E-CoolerHere are the instructions to drive the T-Scrapers
inside the ESR Electron Cooler
 Scrapereinstellung_am_E-Kühler.pdf 
  16   Tue Jun 17 16:33:16 2025 Carsten, SimonDAQGate Q-swtichThe gate generator of the Q-switch signal
for the Vuprom TDC was set to 10ms.
  
  15   Tue Jun 17 09:32:00 2025 KonstantinDAQMeda data acquisitionThe metadata provided by Medusa (Labview channels
in former times) are now labelled correctly
in Go4. 
 Bildschirmfoto_2025-06-17_11-23-40.png 
  14   Mon Jun 16 22:27:01 2025 Carsten, SimonDAQThresholds We also adjusted the thresholds for the mesytic.
Doing that, we noticed that one of the trafos
is broken (5th row, middle).
  
  13   Mon Jun 16 21:47:23 2025 Carsten, Wilfried, Simon DetectorsNew threshold values for PMTsWe adjusted the thresholds for the PMTs and
increased the voltage of PMT middle to 2.7kV
  
  12   Mon Jun 16 07:26:01 2025 KonstantinDAQStartup of Medusa and DAQ1The laser control software (Medusa) is hosted
on the Windows PC ATPPC023. First connect
via the command
 MedusaStartup.JPGMedusa.png 
  11   Fri Jun 13 18:14:02 2025 Konstantin, CarstenDAQDAQ FolderDryrun of DAQs. Folders of the data are:

  
  10   Thu Jun 12 16:27:45 2025 CarstenDAQNTCAP directoryOn the NTCAP the directory is changed to 
Q:/E0052
  
  9   Thu Jun 12 15:53:13 2025 CarstenDAQdata directory and backup disk (WDBook2 RAID)The data of DAQ1 are stored in the following
directory and with the following filenames
  
  8   Wed Jun 11 17:33:04 2025 Carsten, JanDAQDAQ2 - new computer , new directories (lxg0155 doesn't work reliably)LXG0155 (used in form er beamtimes) crashes
and hangs and hangs eventually. 
  
  7   Wed Jun 11 16:44:39 2025 Carsten, JanDetectorsDetector voltages and thresholds for DAQ2 (Mesycontrol)Voltages (2650v) for the small band 9423 is
ok. Signals are nicely separated from background,
thresholds can be easily set.
 Mesytec_CFD_Settings_2025-06-11_17-17-33.png 
ELOG V3.1.5-fc6679b