HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018 E0028 E0038 E0075
  laser spectroscopy on Be-like Kr, Page 2 of 4  Not logged in ELOG logo
ID Date Author Category Subject Text Attachments
  51   Fri Feb 28 15:30:41 2025 Tamina, RodolfoDetectorsFast PD Nord WestWe had an issue with the PD NW. The signal
after the Rate Divider was not OK.
  
  50   Fri Feb 28 13:53:06 2025 Tamina, RodolfoRunsIonenbunchSignal of PMT-Mitte. Recorded with an unstable
ion bunch. With approx. 3E8 Ions
 Screenshot_Kr_Ionenbunch_2025-02-28_1350.pngScreenshot_Kr_Ionenbunch_2025-02-28_1445.png 
  Draft   Fri Feb 28 08:57:30 2025  General 28.2.25

Morning shift Aufgaben(nach Sebas:
  
  48   Fri Feb 28 05:53:59 2025 Denise, Sebastian, BenediktCalibrationLaserstrahllage mit ScraperLaserstrahl mit Hilfe der Blenden eingestellt,
dann mit Scrapern vermessen
  
  47   Fri Feb 28 04:47:39 2025 Denise, Sebastian, BenediktCalibrationGentec ESR KalibrationUnsicherheiten Gentecs im allgemeinen bei
etwa 6.5%
  
  46   Fri Feb 28 00:44:31 2025 Denise, SebastianDAQShutter auf - Shutter zu Go4 SignaleShutter auf (TDC Kanal 12 / Go4 Kanal TrS11_Shutter
auf)
Shutter zu (TDC Kanal 13 / Go4 Kanal TrS12_Shutter
 Screenshot_2025-02-28-Shutter_auf-Shutter_zu.png 
  45   Fri Feb 28 00:35:46 2025 BenediktGeneralAnleitung WellenlängenscanEs folgt eine Anleitung, wie die Wellenlänge
des Lasersystems mit der PulsedLaserGUI gesteuert
werden kann.
  
  44   Fri Feb 28 00:19:04 2025 Konstantin, SebastianDetectorsPMT power supplyWe checked the PMT signal inside the cave
by using the local HV power supply. If one
needs to repeat the measurement at another
 Screenshot_2025-02-28_00-33.png 
  43   Fri Feb 28 00:18:16 2025 TaminaGeneralKalibration UV Monitor im LaserlaborDer UV Monitor steht in der Reflexion des
PBSC auf dem Lasertisch im Laserlabor. 
  
  42   Thu Feb 27 23:01:26 2025 Danyal, Konstantin, Sebastian CalibrationDelaytime measurementYesterday, we forgot to take the transit time
of the PMT's into account when we set up
the delay times of the detectors PMT middle,
  
  41   Thu Feb 27 21:23:02 2025 JensDAQPython Data ChannelsPython variables and names in DAQ
Because the Name of variables in DAQ doesnt
match the the real variables, here is List
  
  40   Thu Feb 27 20:21:31 2025 SebastianGeneral10 MHz Ref Clock (Butis)10 MHz Referenz Clock von BUTIS mit AFG als
externl clock verbunden.
Wird benötigt das die Bunch Kavität (DDS
  
  39   Thu Feb 27 16:30:57 2025 Konstantin und RodolfoDAQHF SignalWe have exchanged the HF-Signal in the DAQ
Electronic: from Dummy-HF to the real HF
from accelerator. 
  
  38   Thu Feb 27 11:19:55 2025 DanyalAcceleratorpictures of the ion beam from 26.02.2025 Hier sind noch ein paar Bilder vom Ionenstrahl
, wie er gestern Abend war.
 Schottky_pick-up_Bild.jpgSchottky_resonator_Bild.jpggebunchter_Strahl_im_ESR.jpg 
  37   Thu Feb 27 10:52:32 2025 RodolfoDetectorsTaking PMTs into operation[B]PMT Nord[/B] the power supply (CH0) is
giving an alarm already by an applied voltage
of -1707 V (operation voltage is -2450 V).
 20250227_105411.jpg20250227_143518.jpg 
  36   Thu Feb 27 09:59:08 2025 KonstantinCalibrationHV-DividerThe readout of the HV divider runs on a RaspberryPi
inside the HV-cave of the electron cooler.
Starting from 2025-02-27 09:42, the divider
  
  35   Thu Feb 27 09:38:04 2025 Konstantin & DanyalDetectorsLaufzeit MessungenGestern Abend haben wir die Signal Laufzeiten
im gemessen:
  
  34   Wed Feb 26 18:53:52 2025 DanyalAcceleratorESR settingsDate: 26.02.2025 @  18:35

E-cooler:  HV = 157366 +/- 1 V  (HV divider
  
  33   Tue Feb 25 14:51:52 2025 DanyalGeneralExperiment AcceptanceToday, 25 FEB 2025 at 9:00 AM, our experiment
was reviewed carefully by the safety committee.
  
  32   Tue Feb 25 13:12:01 2025 SebastianDAQTeilungsverhältnis Ohmlabs HVS250Hochspannung Maßstabsfaktoren

-140 kV >>> 248514
 Bericht_PTB_verbesserter_Teiler_2018.pdfMaßstabsfaktoren.JPG 
ELOG V3.1.5-fc6679b