| ID | Date | Author | Category | Subject | Year | 
| 398 | Sat May 22 21:40:09 2021 | Pierre-Michel & Enis | Runs | Run 0077 | 2021 | 
| open file : e127b_run0077.lmd 
 beam: 124Xe54+
 energy: 7 MeV/u
 cooler voltage:
 
 purpose:
 Reference  measurement with scraper
 
 Detector position (Si): -19 mm
 GE01DD1IG position (Scraper): +2.5 mm (20mm away from beam)
 
 run start: 21:39:33 22-05-2021
 run stop: 22:20:22 22-05-2021
 _______________________________________________________________________________
 
 ON Rates
 Si_ON_bdt:    15 Hz
 Xray_ON_bdt:  700 Hz
 BaF_OR:     ---
 
 Si voltage:          100.0 V
 Si leakage current:  340 uA
 
 
 SIS particles before:      1.2e8
 ESR current at injection:  4e7
 ESR particles after decel.: 2.5e6
 Target ON density:          1.0e14
 | 
| 397 | Sat May 22 21:33:22 2021 | Laszlo | Detectors | Xray detectors are filled at 90 and 35 | 2021 | 
|  | 
| 396 | Sat May 22 20:47:39 2021 | Pierre-Michel & Enis | Runs | Run 0076 | 2021 | 
| open file : e127b_run0076.lmd 
 beam: 124Xe54+
 energy: 7 MeV/u
 cooler voltage:
 
 purpose:
 Reference  measurement with scraper
 
 Detector position (Si): -19 mm
 GE01DD1IG position (Scraper): +2.5 mm (20mm away from beam)
 
 run start: 1:: 22-05-2021
 run stop: 21:21:38 22-05-2021
 _______________________________________________________________________________
 
 ON Rates
 Si_ON_bdt:    15 Hz
 Xray_ON_bdt:  700 Hz
 BaF_OR:     ---
 
 Si voltage:          100.0 V
 Si leakage current:  340 uA
 
 
 SIS particles before:      1.3e8
 ESR current at injection:  4.5e7
 ESR particles after decel.: 2.5e6
 Target ON density:          1.0e14
 | 
| 395 | Sat May 22 20:46:19 2021 | Laszlo, Pierre.Michel, Enis | General | new scraper position | 2021 | 
| The scraper was moved into 2.5mm in absolute which corresponds to 20mm away from the beam axis.
Pressluft Antriebe
Position 1
GE01DD1IG (SC12) 2.5 mm
Position 2
GE01DD1IG (SC9) 2.5 mm | 
| 394 | Sat May 22 19:32:28 2021 | Pierre-Michel & Enis | Runs | Run 0075 | 2021 | 
| open file : e127b_run0075.lmd 
 beam: 124Xe54+
 energy: 7 MeV/u
 cooler voltage:
 
 purpose:
 Reference  measurement with scraper
 
 Detector position (Si): -19 mm
 GE01DD1IG position (Scraper): +12.5 mm
 
 run start: 1:: 22-05-2021
 run stop: 1:: 22-05-2021
 _______________________________________________________________________________
 
 ON Rates
 Si_ON_bdt:    15 Hz
 Xray_ON_bdt:  700 Hz
 BaF_OR:     ---
 
 Si voltage:          100.0 V
 Si leakage current:  340 uA
 
 
 SIS particles before:      1.1e8
 ESR current at injection:  2.5e7
 ESR particles after decel.: 2.5e6
 Target ON density:          1.0e14
 | 
| 393 | Sat May 22 19:05:13 2021 | Pierre-Michel & Enis | Runs | Run 0074 | 2021 | 
| Note: SIS lost HF, no beam since 19:21. Stopped run early 
 open file : e127b_run0074.lmd
 
 beam: 124Xe54+
 energy: 7 MeV/u
 cooler voltage:
 
 purpose:
 Reference  measurement with scraper
 
 Detector position (Si): -19 mm
 GE01DD1IG position (Scraper): +12.5 mm
 
 run start: 1:: 22-05-2021
 run stop: 1:: 22-05-2021
 _______________________________________________________________________________
 
 ON Rates
 Si_ON_bdt:    15 Hz
 Xray_ON_bdt:  700 Hz
 BaF_OR:     ---
 
 Si voltage:          100.0 V
 Si leakage current:  340 uA
 
 
 SIS particles before:      8e7
 ESR current at injection:  3e7
 ESR particles after decel.: 2e6
 Target ON density:          1.0e14
 | 
| 392 | Sat May 22 18:05:22 2021 | Pierre-Michel & Enis | Runs | Run 0073 | 2021 | 
| open file : e127b_run0073.lmd 
 beam: 124Xe54+
 energy: 7 MeV/u
 cooler voltage:
 
 purpose:
 Reference  measurement with scraper
 
 Detector position (Si): -19 mm
 GE01DD1IG position (Scraper): +12.5 mm
 
 run start: 1:: 22-05-2021
 run stop: 1:: 22-05-2021
 _______________________________________________________________________________
 
 ON Rates
 Si_ON_bdt:    10 Hz
 Xray_ON_bdt:  600 Hz
 BaF_OR:     ---
 
 Si voltage:          100.0 V
 Si leakage current:  340 uA
 
 
 SIS particles before:      9e7
 ESR current at injection:  2e7
 ESR particles after decel.: 2e6
 Target ON density:          1.0e14
 | 
| 391 | Sat May 22 17:15:26 2021 | Laszlo | General | Scraper position | 2021 | 
| We found the beam axis with the scraper to be -17.5 mm.
The Scraper position set to +12.5 mm, 30 mm relative from the beam. | 
| 390 | Sat May 22 17:13:48 2021 | Laszlo | Detectors | Si position changed | 2021 | 
| Si was moved from -25 mm to -19 mm before run 0072. | 
| 389 | Sat May 22 17:01:33 2021 | Pierre-Michel & Enis | Runs | Run 0072 | 2021 | 
| open file : e127b_run0072.lmd 
 beam: 124Xe54+
 energy: 7 MeV/u
 cooler voltage:
 
 purpose:
 Reference  measurement with scraper
 
 Detector position (Si): -19 mm
 GE01DD1IG position (Scraper): +12.5 mm
 
 run start: 1:: 22-05-2021
 run stop: 1:: 22-05-2021
 _______________________________________________________________________________
 
 ON Rates
 Si_ON_bdt:    15 Hz
 Xray_ON_bdt:  600 Hz
 BaF_OR:     ---
 
 Si voltage:          100.0 V
 Si leakage current:  340 uA
 
 
 SIS particles before:      9e7
 ESR current at injection:  3e7
 ESR particles after decel.: 3e6
 Target ON density:          1.0e14
 | 
| 388 | Sat May 22 16:29:07 2021 | Laszlo et al. | General | coolers are back | 2021 | 
| we can measure again.
(please write a detailed description if needed) | 
| 387 | Sat May 22 16:28:10 2021 | Laszlo | General | DSSSD movement | 2021 | 
| The DSSSD is moved back to its measurement position(-25mm). | 
| 386 | Sat May 22 14:53:34 2021 | Beatriz et al. | General | DSSD Movement | 2021 | 
| DSSSD moved to the most in position in order to try to scrape the beam with the scraper. | 
| 384 | Sat May 22 11:22:02 2021 | Beatriz, Yury, Jan | Runs | Run 0071 | 2021 | 
| open file : e127b_run0071.lmd 
 beam: 124Xe54+
 energy: 7 MeV/u
 cooler voltage:
 
 purpose:
 Reference measurement
 
 Detector position (Si): -25 mm
 GE01DD1IG position (Scraper): out/not used
 
 run start: 11:21:40 22-05-2021
 run stop: 11:56:31 22-05-2021
 _______________________________________________________________________________
 
 ON Rates
 Si_ON_bdt:    50 Hz
 Xray_ON_bdt:  500 Hz
 BaF_OR:     ---
 
 Si voltage:          100.0 V
 Si leakage current:  340 uA
 
 
 SIS particles before:      9e7
 ESR current at injection:  3e7
 ESR particles after decel.: 3e6
 Target ON density:          1.0e14
 
 
 We close file so that Jan and Jury can go into the ESR to try to solve the problem with the control system of the scraper.
 | 
| 383 | Sat May 22 09:42:20 2021 | Beatriz, Yury, Jan | Runs | Run 0070 | 2021 | 
| open file : e127b_run0070.lmd 
 beam: 124Xe54+
 energy: 7 MeV/u
 cooler voltage:
 
 purpose:
 Reference measurement
 
 Detector position (Si): -25 mm
 GE01DD1IG position (Scraper): out/not used
 
 run start: 09:41:54 22-05-2021
 run stop: 11:20:39 22-05-2021
 _______________________________________________________________________________
 
 ON Rates
 Si_ON_bdt:    50 Hz
 Xray_ON_bdt:  500 Hz
 BaF_OR:     ---
 
 Si voltage:          100.0 V
 Si leakage current:  340 uA
 
 
 SIS particles before:      6e7
 ESR current at injection:  3.1e7
 ESR particles after decel.: 4.1e6
 Target ON density:          1.0e14
 
 No beam or low intensity for quite some while...
 | 
| 382 | Sat May 22 09:07:55 2021 | Beatriz, Yury | Runs | Run 0069 | 2021 | 
| open file : e127b_run0069.lmd 
 beam: 124Xe54+
 energy: 7 MeV/u
 cooler voltage:
 
 purpose:
 Reference measurement
 
 Detector position (Si): -25 mm
 GE01DD1IG position (Scraper): out/not used
 
 run start: 09:07:16 22-05-2021
 run stop: 09:20:24 22-05-2021
 _______________________________________________________________________________
 
 ON Rates
 Si_ON_bdt:    30 Hz
 Xray_ON_bdt:  200 Hz
 BaF_OR:     ---
 
 Si voltage:          100.0 V
 Si leakage current:  330 uA
 
 
 SIS particles before:      2e7
 ESR current at injection:  6e6
 ESR particles after decel.: ~1e6
 Target ON density:          1.0e14
 
 | 
| 381 | Sat May 22 08:00:52 2021 | Beatriz, Yury | Runs | Run 0068 | 2021 | 
| open file : e127b_run0068.lmd 
 beam: 124Xe54+
 energy: 7 MeV/u
 cooler voltage:
 
 purpose:
 Reference measurement
 
 Detector position (Si): -25 mm
 GE01DD1IG position (Scraper): out/not used
 
 run start: 08:00:43 22-05-2021
 run stop: 09:06:54 22-05-2021
 _______________________________________________________________________________
 
 ON Rates
 Si_ON_bdt:    30 Hz
 Xray_ON_bdt:  200 Hz
 BaF_OR:     ---
 
 Si voltage:          100.0 V
 Si leakage current:  330 uA
 
 
 SIS particles before:      2e7
 ESR current at injection:  6e6
 ESR particles after decel.: ~1e6
 Target ON density:          1.0e14
 
 The intensity is lower because they are preparing a 4He beam for R3B.
 For a few minutes there were no particles in SIS.
 We have quite some fluctuations in intensity in this run.
 | 
| 380 | Sat May 22 06:53:08 2021 | Cobus, Michele, Beatriz, Yury | Runs | Run 0067 | 2021 | 
| open file : e127b_run0067.lmd 
 beam: 124Xe54+
 energy: 7 MeV/u
 cooler voltage:
 
 purpose:
 Reference measurement
 
 Detector position (Si): -25 mm
 GE01DD1IG position (Scraper): out/not used
 
 run start: 06:52:32 22-05-2021
 run stop: 08:00 22-05-2021
 _______________________________________________________________________________
 
 ON Rates
 Si_ON_bdt:    50 Hz
 Xray_ON_bdt:  450 Hz
 BaF_OR:     ---
 
 Si voltage:          100.0 V
 Si leakage current:  340 uA
 
 
 SIS particles before:      7.3e7
 ESR current at injection:  1.9e7
 ESR particles after decel.: 2.0e6
 Target ON density:          1.0e14
 
 For some minutes there was no beam in the ESR because Yuri was trying to move the scraper.
 | 
| 379 | Sat May 22 05:49:31 2021 | Cobus, Michele | Runs | Run 0066 | 2021 | 
| open file : e127b_run0066.lmd 
 beam: 124Xe54+
 energy: 7 MeV/u
 cooler voltage:
 
 purpose:
 Reference measurement
 
 Detector position (Si): -25 mm
 GE01DD1IG position (Scraper): out/not used
 
 run start: 05:48:52 22-05-2021
 run stop: 06:52:28 22-05-2021
 _______________________________________________________________________________
 
 ON Rates
 Si_ON_bdt:    70 Hz
 Xray_ON_bdt:  700 Hz
 BaF_OR:     ---
 
 Si voltage:          100.0 V
 Si leakage current:  340 uA
 
 
 SIS particles before:      1.25e8
 ESR current at injection:  3.75e7
 ESR particles after decel.: 3.5e6
 Target ON density:          1.0e14
 
 | 
| 378 | Sat May 22 04:50:21 2021 | Cobus, Michele | Runs | Run 0065 | 2021 | 
| open file : e127b_run0065.lmd 
 beam: 124Xe54+
 energy: 7 MeV/u
 cooler voltage:
 
 purpose:
 Reference measurement
 
 Detector position (Si): -25 mm
 GE01DD1IG position (Scraper): out/not used
 
 run start: 04:49:47 22-05-2021
 run stop: 04:48:41 22-05-2021
 _______________________________________________________________________________
 
 ON Rates
 Si_ON_bdt:    70 Hz
 Xray_ON_bdt:  650 Hz
 BaF_OR:     ---
 
 Si voltage:          100.0 V
 Si leakage current:  340 uA
 
 
 SIS particles before:      1.1e8
 ESR current at injection:  3.5e7
 ESR particles after decel.: 2.5e6
 Target ON density:          1.0e14
 
 |