HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018 E0028 E0038
  proton-capture on 118Te, Page 25 of 29  Not logged in ELOG logo
ID Date Author Categorydown Subject Year
  296   Fri Apr 30 09:57:57 2021 Jan GloriusCalibrationrun0005 - Xray90 calib Pb210 d=167.5mm 2021
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
Source: 210Pb
Distance: 167.5mm
Start time: 9:58:00 - 30.04.2021
Stop time:  10:01:09 - 30.04.2021

file name: e127b_run0005.lmd
avrg. rate: 50Hz
dead-time:  1%

Still rate peaks, probably the ESR kicker or some other external influence. 
So we have double triggers occasionally. This might slightly affect deadtime determination.
  298   Fri Apr 30 11:04:31 2021 Jan GloriusCalibrationrun0006 - Xray90 calib Pb210 d=167.5mm 2021
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
Source: 210Pb
Distance: 167.5mm
Start time: 11:04:03 - 30.04.2021
Stop time:  11:53:59 - 30.04.2021

file name: e127b_run0006.lmd
avrg. rate: 50Hz
dead-time:  1%

Now with blocking gate to get rid of multiple triggers.
Dead time determination using: adt/bdt

Threshold at ~20keV

Last seconds of run may not be usable: DAQ node2 has been deactivated shortly by mistake.
  299   Fri Apr 30 13:07:23 2021 Jan GloriusCalibrationrun0008 - Xray90 calib Pb210 d=167.5mm 2021
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
Source: 210Pb
Distance: 167.5mm
Start time: 13:08:35 - 30.04.2021
Stop time:  13:49:56 - 30.04.2021

file name: e127b_run0008.lmd
avrg. rate: 60Hz
dead-time:  1%

Now with blocking gate to get rid of multiple triggers.
Dead time determination using: adt/bdt

Threshold at ~10keV

We have some external noise on the signal. 
  300   Fri Apr 30 13:10:44 2021 Jan GloriusCalibrationrun0007 - crap2021
crap run
  301   Fri Apr 30 13:56:59 2021 Jan GloriusCalibrationrun0009 - Xray90 calib Am241 d=167.5mm 2021
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
Source: 241Am_high
Distance: 167.5mm
Start time: 13:56:33 - 30.04.2021
Stop time:  14:28:53 - 30.04.2021

file name: e127b_run0009.lmd
avrg. rate: 600Hz
dead-time:  4%

Now with blocking gate to get rid of multiple triggers.
Dead time determination using: adt/bdt

Threshold at ~10keV

We have some external noise on the signal. 
  302   Fri Apr 30 14:35:11 2021 Jan GloriusCalibrationrun0010 - Xray90 calib Ba133 d=167.5mm 2021
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
Source: 133Ba_high
Distance: 167.5mm
Start time: 14:34:51 - 30.04.2021
Stop time:  14:45:42 - 30.04.2021

file name: e127b_run0010.lmd
avrg. rate: 2600 Hz
dead-time:  16%

Now with blocking gate to get rid of multiple triggers.
Dead time determination using: adt/bdt

Threshold at ~10keV

We have some external noise on the signal. 
  303   Fri Apr 30 14:48:01 2021 Jan GloriusCalibrationrun0011 - Xray90 calib Ba133_low d=167.5mm 2021
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
Source: 133Ba_low
Distance: 167.5mm
Start time: 14:47:40 - 30.04.2021
Stop time:  15:50:12 - 30.04.2021

file name: e127b_run0011.lmd
avrg. rate: 310 Hz
dead-time:  3%

Now with blocking gate to get rid of multiple triggers.
Dead time determination using: adt/bdt

Threshold at ~10keV

We have some external noise on the signal. 
  304   Sat May 1 22:34:16 2021 Jan GloriusCalibrationefficiency xray 90deg2021
the attached efficiency curve for the 90 deg. xray detector is made with the runs
e127b_run0008.lmd
e127b_run0009.lmd
e127b_run0010.lmd

All of them were taken in the lab (not in ESR).

This is a very preliminary analysis!
E.g. deadtime has been estimated and was not cleanly determined.
Attachment 1: 90deg_eff.png
90deg_eff.png
Attachment 2: 90deg_eff.dat
59.54	0.00212	0.00007
26.34	0.00180	0.00011
53.16	0.00215	0.00007
79.61	0.00218	0.00008
81.00	0.00222	0.00007
34.95	0.00209	0.00007
35.87	0.00209	0.00009
30.80	0.00199	0.00006
46.54	0.00217	0.00009
  305   Mon May 3 14:55:44 2021 Jan GloriusCalibrationXray energy calib. table2021
Here is the table of energy calibration lines from Am241 for all Xray detectors for several measurements.
Also the linear fit parameters [ E(ch) = A*ch + B ] are given at the bottom of the table.

E [keV] xray90 (GEM1800) xray35 (GLP) xray145 (GEM3000)
lab (28.04.21) lab (04.05.21) lab (11.05.21)
59.54 2402.45 ch 2399.07 ch 2423.24 ch
26.34 1107.38 ch 1116.65 ch 1127.94 ch
13.95 624.33 ch 638.38 ch 645.39 ch
slope A 0.0256387 keV/ch 0.0258923 keV/ch 0.0256408 keV/ch
offset B -2.05477 keV -2.57638 keV -2.59114 keV
  306   Tue May 4 14:31:51 2021 Jan GloriusCalibrationrun0012 - Xray35 calib Am241_hi d=334mm2021
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 241Am_hi
Distance: 334mm
Start time: 14:31:24 - 04.05.2021
Stop time:  14:42:13 - 04.05.2021

file name: e127b_run0012.lmd
avrg. rate: 25 Hz
dead-time:  0%


Threshold < 10keV

The scale serial communication is off.

Stopped due to noise rate...
  308   Tue May 4 14:46:15 2021 Jan GloriusCalibrationrun0013 - Xray35 calib Am241_hi d=334mm2021
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 241Am_hi
Distance: 334mm
Start time: 14:45:51 - 04.05.2021
Stop time:  15:47:48 - 04.05.2021

file name: e127b_run0013.lmd
avrg. rate: 25 Hz
dead-time:  0%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
  309   Tue May 4 16:04:42 2021 Jan GloriusCalibrationrun0014 - Xray35 calib Am241_hi d=334mm2021
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 241Am_hi
Distance: 334mm
Start time: 16:04:21 - 04.05.2021
Stop time:  20:41:45 - 04.05.2021

file name: e127b_run0014.lmd
avrg. rate: 25 Hz
dead-time:  0%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
  310   Tue May 4 21:18:15 2021 Jan GloriusCalibrationrun0015 - Xray35 calib Am241_hi d=334mm2021
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 241Am_hi
Distance: 334mm
Start time: 21:17:44 - 04.05.2021
Stop time:   8:42:09 - 05.05.2021

file name: e127b_run0015.lmd
avrg. rate: 25 Hz
dead-time:  0%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
  311   Wed May 5 10:02:27 2021 Jan GloriusCalibrationrun0016 - Xray35 calib Ba133_hi d=334mm2021
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 133Ba_hi
Distance: 334mm
Start time: 10:02:07 - 05.05.2021
Stop time:  14:13:23 - 05.05.2021

file name: e127b_run0016.lmd
avrg. rate: 90 Hz
dead-time:  1%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
  312   Wed May 5 14:24:57 2021 Jan GloriusCalibrationrun0018 - Xray35 calib Pb210 d=334mm2021
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 210Pb
Distance: 334mm
Start time: 14:28:26 - 05.05.2021
Stop time:  21:05:00 - 05.05.2021

file name: e127b_run0018.lmd
avrg. rate: 5 Hz
dead-time:  0%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.

There are many rate spikes in the first hour of measurement.
  313   Wed May 5 14:29:02 2021 Jan GloriusCalibrationrun0017 crap run2021
  314   Wed May 5 21:22:27 2021 Jan GloriusCalibrationrun0019 - Xray35 calib Pb210 d=334mm2021
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 210Pb
Distance: 334mm
Start time: 21:22:06 - 05.05.2021
Stop time:   8:55:21 - 06.05.2021

file name: e127b_run0019.lmd
avrg. rate: 5 Hz
dead-time:  0%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.

Many rates spikes, check carefully!
  315   Thu May 6 09:40:37 2021 Jan GloriusCalibrationrun0020 - Xray35 calib Ba133_hi d=334mm2021
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 133Ba_hi
Distance: 334mm
Start time: 9:42:24 - 06.05.2021
Stop time: 14:03:44 - 06.05.2021

file name: e127b_run0020.lmd
avrg. rate: 90 Hz
dead-time:  1%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
  317   Thu May 6 13:39:02 2021 Jan GloriusCalibrationefficiency xray 35deg2021
the attached efficiency curve for the 35 deg. xray detector is made with the runs
e127b_run0015.lmd
e127b_run0016.lmd
e127b_run0017.lmd + e127b_run0019.lmd

All of them were taken in the lab (not in ESR).

This is a very preliminary analysis!
E.g. deadtime has been estimated and was not cleanly determined.
Attachment 1: 35deg_eff.png
35deg_eff.png
Attachment 2: 35deg_eff.dat
59.54	0.00007352	0.00000236
26.34	0.00006169	0.00000377
53.16	0.00007545	0.00000262
79.61	0.00007978	0.00000292
81.00	0.00007687	0.00000242
34.95	0.00007378	0.00000237
35.87	0.00007433	0.00000309
46.54	0.00007075	0.00000302
  318   Thu May 6 14:30:35 2021 Jan GloriusCalibrationrun0021 - Xray35 calib Pb210 d=334mm2021
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 210Pb
Distance: 334mm
Start time: 14:21:09 - 06.05.2021
Stop time:  21:06:10 - 06.05.2021

file name: e127b_run0021.lmd
avrg. rate: 5 Hz
dead-time:  0%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
ELOG V3.1.5-fc6679b