386
|
Sat May 22 14:53:34 2021 |
Beatriz et al. | General | DSSD Movement | 2021 |
DSSSD moved to the most in position in order to try to scrape the beam with the scraper. |
387
|
Sat May 22 16:28:10 2021 |
Laszlo | General | DSSSD movement | 2021 |
The DSSSD is moved back to its measurement position(-25mm). |
388
|
Sat May 22 16:29:07 2021 |
Laszlo et al. | General | coolers are back | 2021 |
we can measure again.
(please write a detailed description if needed) |
389
|
Sat May 22 17:01:33 2021 |
Pierre-Michel & Enis | Runs | Run 0072 | 2021 |
open file : e127b_run0072.lmd
beam: 124Xe54+
energy: 7 MeV/u
cooler voltage:
purpose:
Reference measurement with scraper
Detector position (Si): -19 mm
GE01DD1IG position (Scraper): +12.5 mm
run start: 1:: 22-05-2021
run stop: 1:: 22-05-2021
_______________________________________________________________________________
ON Rates
Si_ON_bdt: 15 Hz
Xray_ON_bdt: 600 Hz
BaF_OR: ---
Si voltage: 100.0 V
Si leakage current: 340 uA
SIS particles before: 9e7
ESR current at injection: 3e7
ESR particles after decel.: 3e6
Target ON density: 1.0e14 |
390
|
Sat May 22 17:13:48 2021 |
Laszlo | Detectors | Si position changed | 2021 |
Si was moved from -25 mm to -19 mm before run 0072. |
391
|
Sat May 22 17:15:26 2021 |
Laszlo | General | Scraper position | 2021 |
We found the beam axis with the scraper to be -17.5 mm.
The Scraper position set to +12.5 mm, 30 mm relative from the beam. |
392
|
Sat May 22 18:05:22 2021 |
Pierre-Michel & Enis | Runs | Run 0073 | 2021 |
open file : e127b_run0073.lmd
beam: 124Xe54+
energy: 7 MeV/u
cooler voltage:
purpose:
Reference measurement with scraper
Detector position (Si): -19 mm
GE01DD1IG position (Scraper): +12.5 mm
run start: 1:: 22-05-2021
run stop: 1:: 22-05-2021
_______________________________________________________________________________
ON Rates
Si_ON_bdt: 10 Hz
Xray_ON_bdt: 600 Hz
BaF_OR: ---
Si voltage: 100.0 V
Si leakage current: 340 uA
SIS particles before: 9e7
ESR current at injection: 2e7
ESR particles after decel.: 2e6
Target ON density: 1.0e14 |
393
|
Sat May 22 19:05:13 2021 |
Pierre-Michel & Enis | Runs | Run 0074 | 2021 |
Note: SIS lost HF, no beam since 19:21. Stopped run early
open file : e127b_run0074.lmd
beam: 124Xe54+
energy: 7 MeV/u
cooler voltage:
purpose:
Reference measurement with scraper
Detector position (Si): -19 mm
GE01DD1IG position (Scraper): +12.5 mm
run start: 1:: 22-05-2021
run stop: 1:: 22-05-2021
_______________________________________________________________________________
ON Rates
Si_ON_bdt: 15 Hz
Xray_ON_bdt: 700 Hz
BaF_OR: ---
Si voltage: 100.0 V
Si leakage current: 340 uA
SIS particles before: 8e7
ESR current at injection: 3e7
ESR particles after decel.: 2e6
Target ON density: 1.0e14 |
394
|
Sat May 22 19:32:28 2021 |
Pierre-Michel & Enis | Runs | Run 0075 | 2021 |
open file : e127b_run0075.lmd
beam: 124Xe54+
energy: 7 MeV/u
cooler voltage:
purpose:
Reference measurement with scraper
Detector position (Si): -19 mm
GE01DD1IG position (Scraper): +12.5 mm
run start: 1:: 22-05-2021
run stop: 1:: 22-05-2021
_______________________________________________________________________________
ON Rates
Si_ON_bdt: 15 Hz
Xray_ON_bdt: 700 Hz
BaF_OR: ---
Si voltage: 100.0 V
Si leakage current: 340 uA
SIS particles before: 1.1e8
ESR current at injection: 2.5e7
ESR particles after decel.: 2.5e6
Target ON density: 1.0e14 |
395
|
Sat May 22 20:46:19 2021 |
Laszlo, Pierre.Michel, Enis | General | new scraper position | 2021 |
The scraper was moved into 2.5mm in absolute which corresponds to 20mm away from the beam axis.
Pressluft Antriebe
Position 1
GE01DD1IG (SC12) 2.5 mm
Position 2
GE01DD1IG (SC9) 2.5 mm |
396
|
Sat May 22 20:47:39 2021 |
Pierre-Michel & Enis | Runs | Run 0076 | 2021 |
open file : e127b_run0076.lmd
beam: 124Xe54+
energy: 7 MeV/u
cooler voltage:
purpose:
Reference measurement with scraper
Detector position (Si): -19 mm
GE01DD1IG position (Scraper): +2.5 mm (20mm away from beam)
run start: 1:: 22-05-2021
run stop: 21:21:38 22-05-2021
_______________________________________________________________________________
ON Rates
Si_ON_bdt: 15 Hz
Xray_ON_bdt: 700 Hz
BaF_OR: ---
Si voltage: 100.0 V
Si leakage current: 340 uA
SIS particles before: 1.3e8
ESR current at injection: 4.5e7
ESR particles after decel.: 2.5e6
Target ON density: 1.0e14 |
397
|
Sat May 22 21:33:22 2021 |
Laszlo | Detectors | Xray detectors are filled at 90 and 35 | 2021 |
|
398
|
Sat May 22 21:40:09 2021 |
Pierre-Michel & Enis | Runs | Run 0077 | 2021 |
open file : e127b_run0077.lmd
beam: 124Xe54+
energy: 7 MeV/u
cooler voltage:
purpose:
Reference measurement with scraper
Detector position (Si): -19 mm
GE01DD1IG position (Scraper): +2.5 mm (20mm away from beam)
run start: 21:39:33 22-05-2021
run stop: 22:20:22 22-05-2021
_______________________________________________________________________________
ON Rates
Si_ON_bdt: 15 Hz
Xray_ON_bdt: 700 Hz
BaF_OR: ---
Si voltage: 100.0 V
Si leakage current: 340 uA
SIS particles before: 1.2e8
ESR current at injection: 4e7
ESR particles after decel.: 2.5e6
Target ON density: 1.0e14 |
399
|
Sat May 22 22:25:42 2021 |
Pierre-Michel & Enis | Runs | Run 0078 | 2021 |
open file : e127b_run0078.lmd
beam: 124Xe54+
energy: 7 MeV/u
cooler voltage:
purpose:
Reference measurement with scraper
Detector position (Si): -19 mm
GE01DD1IG position (Scraper): +7.5 mm (25mm away from beam)
run start: 22:25:23 22-05-2021
run stop: 23:27:16 22-05-2021
_______________________________________________________________________________
ON Rates
Si_ON_bdt: 15 Hz
Xray_ON_bdt: 700 Hz
BaF_OR: ---
Si voltage: 100.0 V
Si leakage current: 340 uA
SIS particles before: 9e7
ESR current at injection: 2.2e7
ESR particles after decel.: 2.5e6
Target ON density: 1.0e14
Info: Dipole MU2 did no longer work, reset during the run, tried to reset at 23:12 Uhr. Problem not solved at the end of the run. Problematic injection structure |
400
|
Sat May 22 22:29:48 2021 |
Laszlo, Pierre.Michel, Enis | General | new scraper position | 2021 |
The scraper was moved into 7.5 mm in absolute which corresponds to 25 mm away from the beam axis.
Pressluft Antriebe
Position 1
GE01DD1IG (SC12) 7.5 mm
Position 2
GE01DD1IG (SC9) 7.5 mm |
401
|
Sat May 22 23:28:08 2021 |
Philipp, Benny | Runs | Run 0079 | 2021 |
open file : e127b_run0079.lmd
beam: 124Xe54+
energy: 7 MeV/u
cooler voltage:
purpose:
Reference measurement with scraper
Detector position (Si): -19 mm
GE01DD1IG position (Scraper): +7.5 mm (25mm away from beam)
run start: 23:35:43 22.05.2021
run stop: 00:34:46 23.05.2021
_______________________________________________________________________________
ON Rates
Si_ON_bdt: 12 Hz
Xray_ON_bdt: 650 Hz
BaF_OR: ---
Si voltage: 100.0 V
Si leakage current: 340 uA
SIS particles before: 1e8
ESR current at injection: 3.5e7
ESR particles after decel.: 3e6
Target ON density: 1.0e14 |
402
|
Sun May 23 00:35:33 2021 |
Philipp, Benny | Runs | Run 0080 | 2021 |
open file : e127b_run0080.lmd
beam: 124Xe54+
energy: 7 MeV/u
cooler voltage:
purpose:
Reference measurement with scraper
Detector position (Si): -19 mm
GE01DD1IG position (Scraper): +7.5 mm (25mm away from beam)
run start: 00:35:16 23.05.2021
run stop: 01:35:21 23.05.2021
_______________________________________________________________________________
ON Rates
Si_ON_bdt: 12 Hz
Xray_ON_bdt: 650 Hz
BaF_OR: ---
Si voltage: 100.0 V
Si leakage current: 340 uA
SIS particles before: xxxxx scaling problem, value negative, seem to be a visual problem only
ESR current at injection: 1.6e7
ESR particles after decel.: 2e6
Target ON density: 1.0e14 |
403
|
Sun May 23 01:36:04 2021 |
Philipp, Benny | Runs | Run 0081 | 2021 |
open file : e127b_run0081.lmd
beam: 124Xe54+
energy: 7 MeV/u
cooler voltage:
purpose:
Reference measurement with scraper
Detector position (Si): -19 mm
GE01DD1IG position (Scraper): +7.5 mm (25mm away from beam)
run start: 01:36:06 23.05.2021
run stop: 02:35:37 23.05.2021
_______________________________________________________________________________
ON Rates
Si_ON_bdt: 12 Hz
Xray_ON_bdt: 650 Hz
BaF_OR: ---
Si voltage: 100.0 V
Si leakage current: 340 uA
SIS particles before: 5.5e7
ESR current at injection: 1.2e7
ESR particles after decel.: 2.2e6
Target ON density: 1.0e14 |
404
|
Sun May 23 02:16:47 2021 |
Laszlo | General | quick comparison between different scraping positions | 2021 |
So far we did 3 scraping positions:
- 30mm away from beam axis (lmd72-75)
- 25mm away from beam axis (lmd78-80)
- 20mm away from beam axis (lmd76-77)
below one can see the comparison of the scrapings for the DSSSD spectrum. Please mind the different amount of data collected |
405
|
Sun May 23 02:36:07 2021 |
Philipp, Benny | Runs | Run 0082 | 2021 |
open file : e127b_run0082.lmd
beam: 124Xe54+
energy: 7 MeV/u
cooler voltage:
purpose:
Reference measurement with scraper
Detector position (Si): -19 mm
GE01DD1IG position (Scraper): +7.5 mm (25mm away from beam)
run start: 02:36:02 23.05.2021
run stop: 03:34:08 23.05.2021
_______________________________________________________________________________
ON Rates
Si_ON_bdt: 12 Hz
Xray_ON_bdt: 600 Hz
BaF_OR: ---
Si voltage: 100.0 V
Si leakage current: 340 uA
SIS particles before: 4.5e7
ESR current at injection: 8e6
ESR particles after decel.: 2.5e6
Target ON density: 1.0e14 |