  13   Thu Mar 3 17:34:42 2022 RodolfoLaserPump Laser We have issues with our Pump Laser (Spectra
Physics - Quanta Ray Pro). After the service
made on mid Dezember (2021) the laser does
 QRAY_Simmern.jpg 
  12   Thu Mar 3 17:21:26 2022 RodolfoDetectorsMgF2 WindowsNew reentrance viewports made of MgF2 have
been installed at the ESR "Optical Detection
Region". In the middle the old window (also
 ESR_Schaufenster_Spiegelkammer_10-02-2022.jpg 
  11   Thu Mar 3 13:50:45 2022 RodolfoLaserDye cell cracketThe "20 mm" dye cell used for the Oscillator
and Pramplifier got damage. I supposed it
was produced by the pump beam hitting on
 Cobra_Farbstoffzelle_08-02-2022.jpg 
  10   Fri Nov 19 15:50:44 2021 RodolfoAcceleratorLaser windows at E-Cooler sideNew laser viewports have been installed at
the E-Cooler side of the ESR.
 ESR_Laserfenster_NO_Nov2021.jpgESR_Laserfenster_SO_Nov2021.jpg 
  9   Fri Nov 5 17:23:23 2021 RodolfoDAQDye trigger out The dye laser offers different scan settings
(see pages 65 - 67 of the user manual). We
are using the following setting, see attached
 Cobra_ScanSettings_05-11-2021.PNG 
  8   Thu Oct 14 10:21:46 2021 RodolfoDetectorsDismounting Cu-MirrorWe have dismounted the parabolic Cu-Mirror
at ESR on 13.10.2021 
 AusbauCuSpiegel_2021.jpg 
  Draft   Wed Sep 8 21:44:55 2021 RodolfoGeneralPreparation[TABLE border="1"]
[B]Status[/B]|[B]Arbeitspaket[/B]|[COLOR=chocolate][B]Kommentare[/B][/COLOR]
|-
  
  6   Fri Aug 6 10:40:54 2021 RodolfoDAQmbs in atplaserThe "MBS program" (data acquisition) is now
located within atplaser@r4l-41, in the folder
"th21":
  
  5   Fri Aug 6 10:23:00 2021 RodolfoDAQatplaser on r4l-41The user [B]atplaser[/B] is now existing on
the [B]R4L-XY[/B] nodes for the ESR group.
  
  4   Mon Aug 2 10:54:44 2021 RodolfoLaserQuanta Ray - Water leak in the power supplyAfter fixing the broken water connector in
the laser head a new water leak appeared
in the Quanta-Ray power supply. 
 QRAY_WaterPump.jpg 
  3   Wed Jul 28 13:31:22 2021 RodolfoGeneralAir filtersFilters (Hepa-) inside the flow boxes have
been replaced.
 Flowbox_Filters.png 
  2   Mon Jul 26 17:58:52 2021 RodolfoLaserWater FiltersThe water filters (deioninzing and particle)
have been replaced. 
 QRAY_DeionizingFilter.jpg 
  1   Mon Jul 26 17:15:47 2021 RodolfoLaserQuanta Ray - Water leakOn Thursday afternoon (22.07.2021) a water
connector has broken and water has been spilled
in the laser head. 
 QRAY_AnschlussGeplatzt.jpg 