|  Thu Oct 14 10:21:46 2021, Rodolfo, Detectors, Dismounting Cu-Mirror   | 
| We have dismounted the parabolic Cu-Mirror at ESR on 13.10.2021 | 
|  Wed Sep  8 21:44:55 2021, Rodolfo, General, Preparation | 
| [TABLE border="1"] [B]Status[/B]|[B]Arbeitspaket[/B]|[COLOR=chocolate][B]Kommentare[/B][/COLOR]
 |-
 | 
|  Fri Aug  6 10:40:54 2021, Rodolfo, DAQ, mbs in atplaser | 
| The "MBS program" (data acquisition) is now located within atplaser@r4l-41, in the folder "th21": 
 
 | 
|  Fri Aug  6 10:23:00 2021, Rodolfo, DAQ, atplaser on r4l-41 | 
| The user [B]atplaser[/B] is now existing on the [B]R4L-XY[/B] nodes for the ESR group. 
 R4 = [B]R[/B]IO[B]4[/B] Computer
 | 
|  Mon Aug  2 10:54:44 2021, Rodolfo, Laser, Quanta Ray - Water leak in the power supply   | 
| After fixing the broken water connector in the laser head a new water leak appeared in the Quanta-Ray power supply. 
 We found the leak in the water pump flow line (arrow up). We took the water connectors appart, rebuilt the teflon seeling and put the water connectors
 | 
|  Wed Jul 28 13:31:22 2021, Rodolfo, General, Air filters   | 
| Filters (Hepa-) inside the flow boxes have been replaced. | 
|  Mon Jul 26 17:58:52 2021, Rodolfo, Laser, Water Filters   | 
| The water filters (deioninzing and particle) have been replaced. | 
|  Mon Jul 26 17:15:47 2021, Rodolfo, Laser, Quanta Ray - Water leak   | 
| On Thursday afternoon (22.07.2021) a water connector has broken and water has been spilled in the laser head. |