Entry  Wed Aug 21 15:53:36 2019, Shahab, Status Update, Blade installed IMG_20190712_113334.jpg
[B]Installation of the blades[/B]

The ferritic blade was installed successfully. VNA Tests were successful afterwards. Data was taken for offline assessment.
Entry  Tue Nov 5 16:40:12 2019, Shahab, Status Update, Partly torn drive belt on resonator Motor IMG_20191105_155947.jpg
After an inspection, the drive belt of the 3rd motor of the 410MHz Schottky resonator was found to be partly torn. A new belt will be ordered and inserted
into the cavity asap.
Entry  Mon Nov 11 13:51:19 2019, Shahab, General, Fuse inside ESR IMG_20191111_095410.jpg
A fuse has been apparently blown last week inside ESR due to a clean power supply device. This fuse was unfortunately responsible for the Schottky Rack
inside ESR as well. GSI colleagues from electricity workshop fixed the fuse which was inside the ESR, on the right side of the high voltage cave. For information
the picture of the socket is attached.
Entry  Mon Nov 11 14:14:37 2019, Shahab, General, Resonator drive belt exchanged IMG_20191111_092731.jpg
The drive belt for the 400 MHz Schottky resonator has been replaced  by a new one. Tests have been performed, all works fine now. Fine electronics (LNA)
also turned on ready for beam time.
Entry  Mon Nov 11 14:41:30 2019, Shahab, General, Documentation of the problems found in Schototky 410 MHz Motor Bild1.jpgBild4.jpgBild2.jpgBild3.jpg
Hallo Herr Weber,

ich habe jetzt den Antrieb erfolgreich repariert. Nach weitere Untersuchung haben wir festgestellt was das Problem war, was die problematische Funktion