ID |
Date |
Author |
Category |
Subject |
16
|
Tue Jun 17 16:33:16 2025 |
Carsten, Simon | DAQ | Gate Q-swtich |
The gate generator of the Q-switch signal for the Vuprom TDC was set to 10ms. |
74
|
Sun Jun 29 15:53:45 2025 |
Carsten, Simon | Accelerator | No scans possible |
15:45 Right now, we are waiting for stable conditions again. There are problems with both the ion source and the accumulation in SIS.
16:20 SIS and source look better now, but now ESR makes trouble. The stacking looks off and after deceleration and cooling barely any Th-ions are stored. The positions of the scrapers seem to be wrong. Carsten is trying to fix it. |
75
|
Sun Jun 29 18:04:11 2025 |
Carsten, Simon | Runs | Late shift 29.06. |
Time | Part. Det. Int | SIS-intensity | accumulations in ESR | cooler-HV PTB K250 | cooler-HV PTB red | cooler-HV orange | LMD-DAQ1 Files | LMD-DAQ2 Files | scan range(nm) | number of positions | dwell time (laser shots) | n loops | checked laser position | comments |
18:10 | 18000 | 2.8e9 | 30 | 104525 | - | - | 287 | 314 | 548.51 - 548.01 | 201 | 15 | 3 | yes | Part. Det Int. after 3 loops |
18:26 | 14000 | 2.8e9 | 30 | 104525 | - | - | 288 | 315 | 548.51 - 548.01 | 201 | 15 | 2 | yes | Part. Det Int. after 2 loops |
18:41 | 16000 | 2.8e9 | 35 | 104525 | - | - | 289 | 316 | 548.51 - 548.01 | 201 | 15 | 2 | yes | - |
18:57 | 21000 | 2.0e9 | 40 | 104525 | - | - | 290 | 317 | 548.51 - 548.01 | 201 | 15 | 3 | yes | - |
19:14 | 13000 | 2.0e9 | 40 | 104525 | - | - | 291 | 318 | 548.51 - 548.01 | 201 | 15 | 3 | yes | - |
19:53 | 15000 | 2.6e9 | 35 | 104525 | - | - | 292 | 319 | 548.51 - 548.01 | 201 | 15 | 2 | yes | - |
20:08 | 15000 | 2.0e9 | 40 | 104525 | - | - | 293 | 320 | 548.51 - 548.01 | 201 | 15 | 2 | yes | - |
20:25 | 19000 | 2.4e9 | 40 | 104525 | - | - | 294 | 321 | 548.51 - 548.01 | 201 | 15 | 3 | yes | - |
20:44 | 3000 | 2.4e9 | 45 | 104525 | - | - | 295 | 322 | 548.51 - 548.01 | 201 | 15 | 2 | yes | very low intensity, so we kept Go4 off and only ran mbs |
21:04 | 23000 | 2.8e9 | 30 | 104525 | - | - | 296 | 323 | 548.51 - 548.01 | 201 | 15 | 3 | yes | |
21:21 | 24000 | 2.8e9 | 30 | 104525 | - | - | 297 | 324 | 548.51 - 548.01 | 201 | 15 | 3 | yes | |
21:39 | 18000 | 2.8e9 | 30 | 104525 | - | - | 298 | 325 | 548.51 - 548.01 | 201 | 15 | 3 | yes | |
21:57 | 6400 | 2e9 | 40 | 104525 | - | - | 299 | 326 | 548.51 - 548.01 | 201 | 15 | 2 | yes | | low intensity, no Go4
|
All Particle detector values are after the respective number of scans.
AT 1700 Bernd came in since we couldn't properly accumulate. Bernd adjusted the deceleration HF and slightly changed the extraction kicker from 1.9mrad to 1.7 mrad. High intensity of 229/89.
At 19:35 again problems with transmission SIS -> ESR. Put the Kciker (GTE3Mu1/HIKICK) back to 1.9 mrad (orignal value). |
27
|
Wed Jun 18 20:24:01 2025 |
Carsten, Shahab, Ragandeep, Danyal et al. (the dayshift) and ESR crew | Accelerator | 229Th89+ identification |
We believe to have identified the 237U92+ and the 229Th89+ peaks.
Using the Schottky resonator at about 244 MHz and taking the 123rd harmonic:
The Schottky peak of 237U92+ should be at 242.7975 MHz and the one of 229Th89+ at 242.8494 MHz.
The peaks should be ~52 kHz apart.
We have used:
gamma_t = 2.355
alpha_p = 0.18 |
19
|
Wed Jun 18 00:56:28 2025 |
Carsten, Rodolfo, Anton, Alexis, Esther | | beam overlap with t-scrapers |
When testing the overlap between laser and ion beam we observed the following positions:
motor | ion beam position (mm) from outside | ion beam position (mm) from inside
|
GECEDS1VU | 0 | 6.5
|
GECEDS1HA | -22.5 | -13.5
|
GECEDS2VU | 1 | 8.5
|
GECEDS2HA | -20.5 | -10
|
Note: To move GECEDS1VU the corresponding horizontal drive (GECEDS1HA) has to be at 2 mm. |
58
|
Tue Jun 24 18:49:18 2025 |
Carsten, Rima, Nils, Simon | Runs | Second attempt (Late shift 24.06.) |
Scan parameters:
scan range | 551.91 - 549.41
|
positions | 501
|
step width | 0.005nm
|
advance after | 15 rising edges
|
iterations | 1
|
rev. frequency | 1.54315e6
|
Time | Part. Det. Int | SIS-intensity | accumulations in ESR | cooler-HV PTB K250 | cooler-HV PTB red | cooler-HV orange | LMD-DAQ1 Files | LMD-DAQ2 Files | scan range(nm) | number of positions | dwell time (laser shots) | n loops | checked laser position | comments
|
19:03 | 14800 | 2.7e9 | 60 | 104510 | 104514 | 104514 | 118 | 147 | 551.91 - 549.41 | 501 | 15 | 1 | yes | TE beamline slits -30 +30
|
19:22 | 14900 | 2.7e9 | 60 | 104510 | 104514 | 104514 | 119 | 148 | 551.91 - 549.41 | 501 | 15 | 1 | yes | -
|
19:40 | 14500 | 2.7e9 | 60 | 104510 | 104514 | 104514 | 120 | 149 | 551.91 - 549.41 | 501 | 15 | 1 | yes | TE beamline slits -20 +20
|
19:58 | 13800 | 2.0e9 | 60 | 104510 | 104514 | 104514 | 121 | 150 | 551.91 - 549.41 | 501 | 15 | 1 | yes | -
|
20:18 | 10800 | 2.1e9 | 60 | 104510 | 104514 | 104514 | 122 | 151 | 551.91 - 549.41 | 501 | 15 | 1 | yes | -
|
20:37 | 10100 | 2.1e9 | 60 | 104510 | 104514 | 104514 | 123 | 152 | 551.91 - 549.41 | 501 | 15 | 1 | yes | -
|
20:56 | 10500 | 2.1e9 | 60 | 104510 | 104514 | 104514 | 124 | 153 | 551.91 - 549.41 | 501 | 15 | 1 | yes | -
|
21:15 | 9000 | 2.2e9 | 60 | 104510 | 104514 | 104514 | 125 | 154 | 551.91 - 549.41 | 501 | 15 | 1 | yes | after this we change to 45 injections to give more cave access to HHT
|
21:37 | 11500 | 2.3e9 | 45 | 104510 | 104514 | 104514 | 126 | 155 | 551.91 - 549.41 | 501 | 15 | 1 | yes | small pause for HHT
|
22:13 | 13400 | 2.5e9 | 45 | 104510 | 104514 | 104514 | 127 | 156 | 551.91 - 549.41 | 501 | 15 | 1 | yes | small pause for HHT, cooling time in SC7 increased from 20s to 28 s
|
|
45
|
Sun Jun 22 18:47:23 2025 |
Carsten, Peter, Marek, Simon | DAQ | PMT and particle detector settings |
Final voltages and thresholds for the PMTs and particle detectors before the first scan:
Detector | voltage (kV) | threshold (mV) | mesytec
|
particle detector | 1.35 | 100 | -
|
PMT north | 2.7 | 10 | 16
|
PMT mid | 2.65 | 10 | 16
|
PMT south | 2.7 | 10 | 16
|
All other mesytec thresholds remain unchanged and can be found in entry 7.
Important note: Ch0 of the PMT power supply broke, so we moved PMT north to CH3.
Current setup:
CH1: PMT mid
CH2: PMT south
CH3: PMT north |
41
|
Sat Jun 21 18:00:49 2025 |
Carsten, Peter, Danyal, Kristian, Marek | Detectors | PMT signals |
Carsten and Kristian changed (18:00) the thresholds of PMT north and PMT middle in the ESR cave.
See picture below. The effect is negligible.
Carsten will now (18:05) change the channel of the pre-amp of PMT north. |
40
|
Sat Jun 21 17:00:10 2025 |
Carsten, Peter, Danyal, Kristian, Marek | Detectors | PMT north exchanged |
We noted that the dark countrate of PMT north is about 20x higher than that of PMT middle.
We went into the ESR cave to look for stray light.
We decided to put a black cloth over all the PMTs.
Also we made the tubes of the N2 airflow to the PMTs black (tape around).
This did not make a significant change.
We exchanged PMT north (same type 9423B) to see if that made a difference.
It did not really change either.
PMT north = 2650 V
PMT middle = 2700 V
The first picture below shows the dark rates for these two voltages.
At about 17:00 we changed the voltage of PMT north to 2700 V.
This did not change anything.
Carsten swapped the cables of PMT north and PMt middle (inside ESR, before the amps) at 17:14.
The second picture below shows the result of that.
Now black and red are exchanged.
Carsten swapped the cables back again (17:20).
Carsten and Kristian will look at the PMT signals in the ESR cave. |
62
|
Thu Jun 26 19:01:46 2025 |
Carsten, Jan, Simon | Runs | Late shift 26.06. |
Scan parameters:
scan range | 551.91 - 549.41
|
positions | 501
|
step width | 0.005nm
|
advance after | 15 rising edges
|
iterations | 1
|
rev. frequency | 1.54315e6
|
Time | Part. Det. Int | SIS-intensity | accumulations in ESR | cooler-HV PTB K250 | cooler-HV PTB red | cooler-HV orange | LMD-DAQ1 Files | LMD-DAQ2 Files | scan range(nm) | number of positions | dwell time (laser shots) | n loops | checked laser position | mesytec rate ok | comments
|
19:15 | 23000 | 1.5-3.5 e9 | 33 | 104522 | 104526 | 104526 | 142 | 170 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -10 and +10, SIS intensity unstable
|
19:41 | 19000 | 3.3e9 | 30 | 104522 | 104526 | 104526 | 143 | 171 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -7 and +7
|
20:08 | 17500 | 3.3e9 | 25 | 104522 | 104526 | 104526 | 144 | 172 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | -
|
20:24 | 7000 | 3.3e9 | 25 | 104522 | 104526 | 104526 | 145 | 173 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | low Part. Det. Int maybe due to wrond position of part. det.
|
21:19 | 7000 | 3.5e9 | 25 | 104522 | 104526 | 104526 | 146 | 174 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | TE slits at -8 and +8
|
21:40 | 14000 | 3.3e9 | 25 | 104522 | 104526 | 104526 | 147 | 175 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes | adjusted particle detector position
|
22:04 | 19500 | 3.3e9 | 40 | 104522 | 104526 | 104526 | 148 | 176 | 551.91 - 549.41 | 501 | 15 | 1 | yes | yes |
|
After the scan at 22:04 teh FCU new was new calibrated. For this MEdusa was stopped and reopened in the laser lab. This stops the Labview server, and as a consequence the the connection to the MBS was lsot. Then, MBS needs to be restarted and the online data is cleared. Such an intervention should avoided as good as possible.
A pratical solution for beamtimes with such long data taking is still pending.
When trying to restart MBS MBS completlz hung up also resl etc did not work. Crate powered down and restarted. |
63
|
Thu Jun 26 21:27:34 2025 |
Carsten, Jan, Simon | Calibration | Timing |
We checked the timing (before the start of the measurement cycle documented below).
Timing and measurement window are attached. |
7
|
Wed Jun 11 16:44:39 2025 |
Carsten, Jan | Detectors | Detector voltages and thresholds for DAQ2 (Mesycontrol) |
Voltages (2650v) for the small band 9423 is ok. Signals are nicely separated from background, thresholds can be easily set.
For the broadband 9422 the voltage is by far too low. Signals and BG are barely separated.
Increased voltage from 1500V to 1750V (according to datasheet Vmax = 2500V, 1500 V is the min voltage for gain 1e6...).
Now signals are better separated - threshold with mesycontrol can be set in such a way that the dark count rate is 300 - 400 Hz (was > 1kHz).
For detector south, the threshold in DAQ1 has to be readjusted.
Below are the settings for Mesytec CFD. Internally the 2nC chips (x 20 ??? ) are used for both groups. Please note that thre are also 20nC chips in "parking position" (lower amplfication).
The Mesytec PMT CFD Version uses "nC" chips while the other Mesytec Shapers "gain / amplification", i.e. chips labeled with 20 are x20 and have a large amplification while the "nC" chips give the full scale range , aka the 20nC is tghe small amplification.
Note as well: Externally, the channel numbering starts with "1" while the Mesycontrol starts with "0".
The sum of all Mesytec signals defines the gate for the ADC.
Group 4-7 are the x100 signals (similar to Vuprom TDC). These signals go to the 25ps TDC and are used for timing, and to the ADC.
Group 0-3 is x10; thresholds are artificially set slightly too high, signals are used in the ADC only. Since these are the same signals as in the x100 group, their (correct) gate signals can be used in the ADC. The x10 capture also the very high signals.
ch 0(1): PMT South x10 (A).
ch 1(2): PMT Mid x10 (B).
ch 2(3): PMT North x10 (C).
ch 4(5): PMT South x100 (B).
ch 5(6): PMT Mid x100 (D).
ch 6(7): PMT North x100 (F). |
8
|
Wed Jun 11 17:33:04 2025 |
Carsten, Jan | DAQ | DAQ2 - new computer , new directories (lxg0155 doesn't work reliably) |
LXG0155 (used in form er beamtimes) crashes and hangs and hangs eventually.
Lgx0155 is shut-down and is replaced by lxg3113.
For the experiment (DAQ2) still user stoe_exp can be used (Credits - see monitor).
MBS runs on r4l-68 with user atplaser: (/esr/usr/atplaser/) mbsrun/E0052/DAQ2
The data is stored on the ATP NAS (atprfio.gsi.de). (instead of local disk).
In the home directory there is link to access the data (u/stoe_exp/nas/229Th2025 ; ~/nas/229Th2025)
( sshfs -o directport=7777 atprfio.gsi.de:/mnt/raw.data /u/stoe_exp/nas )
For mbs the correct location on atprfio.gsi.de need to be used (cf. @openfile).i.e.
con rfio atprfio -DISK
and
open file /mnt/raw.data/229Th2025/Th25DAQ2_.lmd .
This is the mountpoint in mbs.
The commands @connectdisk and @openfile are adjusted accordingly. |
46
|
Sun Jun 22 21:36:48 2025 |
Carsten, Finn, Zac, Esther | Runs | First laser scans |
Scan parameters:
scan range | 551.91 - 549.41
|
positions | 501
|
step width | 0.005nm
|
advance after | 15 rising edges
|
iterations | 1
|
rev. frequency | 1.54315e6
|
Note: When doing the accumulation in the ESR the particle number (observed in Lassie Monitor) seems to saturate. This is not actually the case but just an effect of the real signal being buried in other species.
Time | Part. Det. Int | SIS-intensity | accumulations in ESR | cooler-HV PTB | cooler-HV ? | LMD-DAQ1 Files | LMD-DAQ2 Files | scan range(nm) | number of positions | dwell time (laser shots) | n loops | checked laser position | comments
|
22:33 | 10800 | 1.25e9 | 60 | 104513 | 104509 | 106 | 0136 | 551.91 - 549.41 | 501 | 15 | 1 | yes | -
|
22:50 | 10800 | 1.3e9 | 60 | 104513 | 104509 | 107 | 0137 | 551.91 - 549.41 | 501 | 15 | 1 | yes | -
|
23:07 | 14000 | 1.3e9 | 60 | 104513 | 104509 | 108 | 0138 | 551.91 - 549.41 | 501 | 15 | 1 | yes | -
|
23:24 | 11500 | 1.3e9 | 60 | 104513 | 104509 | 109 | 0139 | 551.91 - 549.41 | 501 | 15 | 1 | yes | -
|
23:40 | 13100 | 1.3e9 | 60 | 104513 | 104509 | 110 | 0140 | 551.91 - 549.41 | 501 | 15 | 1 | yes | -
|
23:57 | 13100 | 1.3e9 | 60 | 104513 | 104509 | 111 | 0141 | 551.91 - 549.41 | 501 | 15 | 1 | yes | -
|
00:14 | 12700 | 1.4e9 | 60 | 104513 | 104509 | 112 | 0142 | 551.91 - 549.41 | 501 | 15 | 1 | yes | -
|
00:31 | 13500 | 1.4e9 | 60 | 104513 | 104509 | 113 | 0143 | 551.91 - 549.41 | 501 | 15 | 1 | yes | -
|
00:48 | 10700 | 1.25e9 | 60 | 104513 | 104509 | 114 | 0144 | 551.91 - 549.41 | 501 | 15 | 1 | yes | -
|
01:05 | - | 1.0e9 | 60 | 104513 | 104509 | 115 | 0145 | 551.91 - 549.41 | 501 | 15 | 1 | yes | BAD FILE(?): beam was killed as some point during the scan, we don't know why; no counts on the particle detector
|
|
59
|
Tue Jun 24 22:24:36 2025 |
Carsten, Finn, Hiroshi, Esther, Rodolfo | Runs | Data taking (night shift 24.06.) |
Scan parameters:
scan range | 551.91 - 549.41
|
positions | 501
|
step width | 0.005nm
|
advance after | 15 rising edges
|
iterations | 1
|
rev. frequency | 1.54315e6
|
Notes:
1. NTCAP was only started after the fourth scan. The file timestamp is '2025-06-25_00-02-08'.
2. Beam was lost at 03:45 and could not be recovered by the operators. On-call service was called but they have to replace a part, current estimate is 2-3 hours.
Time | Part. Det. Int | SIS-intensity | accumulations in ESR | cooler-HV PTB K250 | cooler-HV PTB red | cooler-HV orange | LMD-DAQ1 Files | LMD-DAQ2 Files | scan range(nm) | number of positions | dwell time (laser shots) | n loops | checked laser position | comments
|
22:39 | 13500 | 2.4e9 | 60 | 104510 | 104514 | 104514 | 128 | 157 | 551.91 - 549.41 | 501 | 15 | 1 | yes | 60 injections, cooling time (SC 7) 28s -> 32s
|
23:00 | 11500 | 2.4e9 | 45 | 104510 | 104514 | 104514 | 129 | 158 | 551.91 - 549.41 | 501 | 15 | 1 | yes | 45 injections
|
23:20 | 10500 | 2.4e9 | 45 | 104510 | 104514 | 104514 | 130 | 159 | 551.91 - 549.41 | 501 | 15 | 1 | yes |
|
23:44 | 9100 | 2.4e9 | 45 | 104510 | 104514 | 104514 | 131 | 160 | 551.91 - 549.41 | 501 | 15 | 1 | yes |
|
00:05 | 14000 | 2.4e9 | 60 | 104510 | 104514 | 104514 | 132 | 161 | 551.91 - 549.41 | 501 | 15 | 1 | yes | back to 60 injections
|
00:28 | 14300 | 2.4e9 | 75 | 104510 | 104514 | 104514 | 133 | 162 | 551.91 - 549.41 | 501 | 15 | 1 | yes | 75 injections, mesytec channel threshold 0-3 set 255
|
01:26 | 13200 | 2.4e9 | 75 | 104510 | 104514 | 104514 | 134 | 163 | 551.91 - 549.41 | 501 | 15 | 1 | yes | there was an interruption at the source, took some time to get the beam back
|
01:48 | 10500 | 2.3e9 | 75 | 104510 | 104514 | 104514 | 135 | 164 | 551.91 - 549.41 | 501 | 15 | 1 | yes |
|
02:10 | 12400 | 2.3e9 | 75 | 104510 | 104514 | 104514 | 136 | 165 | 551.91 - 549.41 | 501 | 15 | 1 | yes |
|
02:33 | 9600 | 2.4e9 | 75 | 104510 | 104514 | 104514 | 137 | 166 | 551.91 - 549.41 | 501 | 15 | 1 | yes |
|
02:55 | 10000 | 2.4e9 | 75 | 104510 | 104514 | 104514 | 138 | 167 | 551.91 - 549.41 | 501 | 15 | 1 | yes |
|
03:18 | 9600 | 2.4e9 | 75 | 104510 | 104514 | 104514 | 139 | 168 | 551.91 - 549.41 | 501 | 15 | 1 | yes |
|
|
76
|
Sun Jun 29 22:06:19 2025 |
Carsten, Alex, Hendrik | Runs | Night shift 29.06. |
[TABLE border="1"]
Time | Part. Det. Int | SIS-intensity| accumulations in ESR | cooler-HV PTB K250 | cooler-HV PTB red | cooler-HV orange | LMD-DAQ1 Files | LMD-DAQ2 Files | scan range(nm) | number of positions | dwell time (laser shots) | n loops | checked laser position | comments |-
22:16 | 4800 | 2e9 | 30 | 104525 | - | - | 300 | 327 | 548.51 - 548.01 | 201 | 15 | 2 | yes | only few ions, did not start Go4 |-
22:28 | 9000 | 2e9 | 15 | 104525 | - | - | 301 | 328 | 548.51 - 548.01 | 201 | 15 | 2 | yes | After this run: Shaky SIS intensity, cathode is being changed |-
22:53 | 7000 | 2e9 | 35 | 104525 | - | - | 302 | 329 | 548.51 - 548.01 | 201 | 15 | 2 | yes |-|-
[/TABLE]
23:20 Failure of GE01KP04, GE01KP05, GE01KP22 (or GE01KP24) devices in the ESR.
Problems with gtemu3-hick (??) , take injections since
01:49 Bernd tried to make things work by switching off and on some of the dipole correction coils. He was even able to get a quite good intensity by leaving one of the coils switched off (although with a slight shift of frequencies visible in Schottky spectrum) but then the ESR electron cooler has failed probably due to problems with the cooling water. So, the decision was taken to stop the measurement at this point. The laser was switched off. |
33
|
Sat Jun 21 01:40:31 2025 |
Carsten, Alex, Anton, Rodolfo, Hiroshi | Accelerator | beam parameters |
The revolution frequency for the measurement condition is: 1.542728 MHz.
The electron cooler voltage is: 104411 V. |
2
|
Mon Jun 2 16:36:57 2025 |
Carsten | Detectors | Mesycontrol |
ssh -Y litv-exp@x86l-204
'mesycontrol' |
9
|
Thu Jun 12 15:53:13 2025 |
Carsten | DAQ | data directory and backup disk (WDBook2 RAID) |
The data of DAQ1 are stored in the following directory and with the following filenames
open file /data.local2/2025_229Th/2025_229ThDAQ1_ -auto -rfio
(from @openfile)
The usual @connectdisk, @openfile @closefile are adapted and can be used.
The WDBook RAID0 disk is attached for backup of the data
/media/atplaser/WDBook2/2025E0052daq1 |
10
|
Thu Jun 12 16:27:45 2025 |
Carsten | DAQ | NTCAP directory |
On the NTCAP the directory is changed to
Q:/E0052 |