ID |
Date |
Author |
Category |
Subject |
Text |
 |
89
|
Mon Mar 3 12:47:08 2025 |
Tamina, Rudolfo, Volker, Danyal, Florian | Calibration | Änderung der Ionenenergie | Änderung der Ionenenergie auf 287,62 MeV/u.
Gamma = 1,30877
|
|
98
|
Tue Mar 4 01:05:33 2025 |
Sebastian, Benedikt, Denise | Runs | whole range scan @2nd ion energy) without shift |
[TABLE border="1"]
Zeit | Scanbereich (nm) | Anzahl Laserschritte |
|
2
|
Fri Dec 13 13:44:13 2024 |
Sebastian | Runs | test | test 123 |
|
17
|
Thu Feb 6 11:03:26 2025 |
Danyal | General | pressures in laser window sections NW and SW | The pressures, read by the UHV technician,
are :
|
|
38
|
Thu Feb 27 11:19:55 2025 |
Danyal | Accelerator | pictures of the ion beam from 26.02.2025 | Hier sind noch ein paar Bilder vom Ionenstrahl
, wie er gestern Abend war. |
  |
92
|
Mon Mar 3 15:40:39 2025 |
Tamina, Danyal | Calibration | necessary tests on 03.03.2025 before 08:00 AM | We must test the following, shortly before
beamtime end:
|
|
22
|
Tue Feb 11 13:08:32 2025 |
Danyal | Accelerator | limited access to ESR this week | Hitrap will kein ESR Zugang erlauben bevor
22 Uhr. Außerdem merken die, wenn ESR-SIS
nicht laufen, dass sie morgens immer 1 Stunde |
|
19
|
Tue Feb 11 10:30:21 2025 |
Danyal | General | laser valves are open | Die Laserventile im ESR sind nun offen. Der
Laserstrahl darf durchgeschossen werden.
|
|
91
|
Mon Mar 3 13:15:51 2025 |
Tamina, Florian, Volker, Danyal | Calibration | frequencies | f_bunch = 1.785 615 785 047 41 MHz
f_rev = f_bunch
|
|
83
|
Mon Mar 3 01:53:18 2025 |
Sebastian, Denise, Benedikt | Runs | extending the lower wavelength range (no shift) | We will extend the min normal scan range by
3 measurements
|
|
82
|
Sun Mar 2 23:05:16 2025 |
*Jens, Reinhold, Danyal | DAQ | extended normal scan range (no shift): 257.6750 - 257.8022 nm | We will extend the max normal scan range by
3 measurements
|
|
93
|
Mon Mar 3 16:19:18 2025 |
Konstantin | DAQ | changed shutter timing - effective from 03.03.2025 @ 16:24 ! | The power and the fluctuations of the laser
should be recorded in the XUV Data, so that
the rate can be corrected.
|
 |
94
|
Mon Mar 3 16:27:13 2025 |
Danyal | Accelerator | bunch length @ new ion energy | the bunch length was measured for the new
ion energy.
It is about 60 ns. This is shorter than |
  |
20
|
Tue Feb 11 10:32:01 2025 |
Danyal | General | access to ESR | Es wird noch zwei 2 stündige Unterbrechungen
am UNILAC geben. Wahrscheinlich diesen Freitag
und nächste Woche Mittwoch oder Donnerstag. |
|
21
|
Tue Feb 11 10:35:38 2025 |
Danyal | Detectors | XUV detector | Volker Hannen kommt am 19.2 (Mittags) und
dann machen wir Abends den XUV Detektor bereid.
|
|
100
|
Tue Mar 4 07:07:11 2025 |
Sebastian, Benedikt, Denise | Detectors | XUV Detektor Laseruntergrund | Laseruntergrund XUV Detektor krzpton0554.lmd
Schritte:
|
|
103
|
Tue Mar 4 08:46:43 2025 |
| General | Wir sagen Daaaankeeeschooooeeeeeeeen | 40 Jahre die Flippers |
|
67
|
Sat Mar 1 09:49:06 2025 |
Tamina, Florian, Konstantin, Volker | Calibration | Vorbereitung Messung Verschiebung Laserpulse zum Bunch | Nach dem Zyklus liegen die PD-Pulse nicht
mehr bei 55 bins (SW) und 85 bins (NW) wie
vorher, sondern bei 56 (SW) und 87 (NW)
|
|
27
|
Fri Feb 14 16:41:54 2025 |
Sebastian | General | Verkabelung PD ESR | Verkabelung von den Photodioden im ESR. Sind
auch gelabelt.
|
|
57
|
Fri Feb 28 20:44:04 2025 |
Konstantin | DAQ | Updated Channel list | Final (hopefully) channel list:
[TABLE border="1"]
|
|