Entry  Sat Mar 1 07:14:48 2025, Night shift, Runs, Tabelle erster Scanbereich 257.175 nm - 257.300 nm 
[TABLE border="1"]
Zeit | Scanbereich (nm) | Anzahl Laserschritte | Start LMD | Stop LMD | Messdauer (s) | Elektronenkuehlung (s) | Gentec Files | Kommentar
|-
Entry  Fri Feb 28 00:19:04 2025, Konstantin, Sebastian, Detectors, PMT power supply Screenshot_2025-02-28_00-33.png
We checked the PMT signal inside the cave by using the local HV power supply. If one needs to repeat the measurement at another time, be careful to use
the correct HV power supply, since the second one, which is mounted in the crate has the wrong polarity. Both PMT's must be supplied with a negative voltage.
Entry  Thu Feb 27 16:30:57 2025, Konstantin und Rodolfo, DAQ, HF Signal 
We have exchanged the HF-Signal in the DAQ Electronic: from Dummy-HF to the real HF from accelerator. 
Entry  Thu Feb 27 09:38:04 2025, Konstantin & Danyal, Detectors, Laufzeit Messungen 
Gestern Abend haben wir die Signal Laufzeiten im gemessen:

PMT Mitte: 1213 ns (bis zur Messhütte ohne DAQ sind es 1025 ns) 
Entry  Sat Mar 1 13:45:42 2025, Konstantin , DAQ, Gentec readout 
Gentec readout is now included in the general slow data acquisition. The issue was that the buffer of the power meter was not emptied fast enough. Now,
the data is read out permanently in a separate thread. 
Entry  Thu Feb 27 09:59:08 2025, Konstantin, Calibration, HV-Divider 
The readout of the HV divider runs on a RaspberryPi inside the HV-cave of the electron cooler.
Starting from 2025-02-27 09:42, the divider ratio was changed from 248517 to 248515. Usually, the
readout should run continuously. If not, the following procedure must be applied:
Entry  Fri Feb 28 15:53:44 2025, Konstantin, Accelerator, HF Signal for Bunching 
The HF Bunching Frequency is: 1.7838817 MHz

Given (read) by the ESR operators in HKR.
Entry  Fri Feb 28 17:36:15 2025, Konstantin, DAQ, Trigger events from accelarator 2025-02-28_17-34-15-990_1.png
We will use GENESIS to create trigger events (injection, ...) for the DAQ
Entry  Fri Feb 28 20:44:04 2025, Konstantin, DAQ, Updated Channel list 
Final (hopefully) channel list:

[TABLE border="1"]
Entry  Sat Mar 1 16:33:22 2025, Konstantin, DAQ, Restart of Go4  Screenshot_2025-03-01_16-34-01.pngScreenshot_2025-03-01_16-45-30.png
Restart of go4 before new scan:

1. Stop Acquisition (Stop button right next to the green play buttons)
Entry  Sat Mar 1 19:19:25 2025, Konstantin, DAQ, Offline Analysis Unbenanntes_Bild.pngg042.png
A copy of the current go4 analysis is placed at atplaser@lxgs1055:/kr_analysis. Use the usual Go4 startup routine to start an offline analysis of data.
They should be copied into the folder lxgs1055:/data.local2/E0075. To combine different 
lmd-files, simply create a file like "FirstScan.lml" containing the paths of the individual files. After this you can read the lml file with go4.
Entry  Mon Mar 3 16:19:18 2025, Konstantin, DAQ, changed shutter timing - effective from 03.03.2025 @ 16:24 ! changed_shutter_timing.jpgchanged_shutter_timing_in_DAQ.jpg
The power and the fluctuations of the laser should be recorded in the XUV Data, so that the rate can be corrected. 
Therefore the shutter timing was changed. 
Entry  Sun Mar 2 16:51:53 2025, Jens, Rodolfo, General, Spaetschicht 
Die Spaetschicht macht weiter, bzw. mit dem naeschten Scan (um 16:41 Uhr) angefangen. Siehe bitte Eintrag 75 (noch als Preview gezeigt). 
Entry  Sun Mar 2 18:55:30 2025, Jens, Reinhold, Danyal, DAQ, Main Scan range (4 blocks) shifted by -0.0012 nm --> one more scan 
We decided to scan the main range, shifted by -0.0012nm, one more time in order to have the non-shifted AND the shifted range both 3x  ! 

[TABLE border="1"]
Entry  Mon Mar 3 20:14:10 2025, Jens, Konstantin, Reinhold, Danyal, DAQ, -0.0012 nm shifted whole range scan (@2nd ion energy) 

[TABLE border="1"]
Zeit | Scanbereich (nm) | Anzahl Laserschritte | Start LMD | Stop LMD | Messdauer (s) | Elektronenkuehlung (s) | Kommentar
Entry  Thu Feb 27 21:23:02 2025, Jens, DAQ, Python Data Channels 
Python variables and names in DAQ
Because the Name of variables in DAQ doesnt match the the real variables, here is List of the channels and what value is in this channel
Entry  Sat Mar 1 18:32:05 2025, Jens, Runs, Laser während Scan 
18:30Uhr:
Timestamp im Programm hat gehangen. Kurz im Anzeigefeld markieren und es ging wieder.
Laser Step Counter hat gehangen und lag mit 4Go 20 Steps auseinander. Da es früh genug auffiel wurde der Number of Steps von 52 auf 72 erhöht. Ist also
Entry  Mon Mar 3 16:40:42 2025, Denise, Tamina, Bene, Sebastian, Jens, Calibration, TODO Lasercrew vor Ende der Messzeit 
Prio 1: Gentec Suedwest kalibrieren (Wichtig fuer Auswertung)
Prio 2: Messung Strahlwanderung mit Wincam im Widescan und ohne (wichtig für zukünftige Strahlzeiten!!!!!)
Prio 3: Signal XUV Detektor in Abhängigkeit der Laserleistung
Entry  Thu Feb 20 13:51:04 2025, Denise, Sebatian, DAQ, DAQ Channelliste 
[TABLE border="1"]
TDC Channels |
|- 
Entry  Fri Feb 28 04:47:39 2025, Denise, Sebastian, Benedikt, Calibration, Gentec ESR Kalibration 
Unsicherheiten Gentecs im allgemeinen bei etwa 6.5%

Gentec NW (PH100 S:UV) OD.3 