HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018 E0028 E0038 E0075
  laser spectroscopy on Be-like Kr, Page 3 of 4  Not logged in ELOG logo
ID Date Author Category Subject
  34   Wed Feb 26 18:53:52 2025 DanyalAcceleratorESR settings
Date: 26.02.2025 @  18:35

E-cooler:  HV = 157366 +/- 1 V  (HV divider value)
E-cooler current = 200 mA (scale:  1V --> 10 mA) 

Schottky f_rev = 1.782012 +/- 1 MHz (measured value) 
f_b = 1.782013 MHz (set value) 

orbit length in ESR = 108.5 m (based on experience, injection is further out, not central) 

y = 1.30796 
ß = 0.64456 


We have 1 ion bunch in the ESR
  33   Tue Feb 25 14:51:52 2025 DanyalGeneralExperiment Acceptance
Today, 25 FEB 2025 at 9:00 AM, our experiment was reviewed carefully by the safety committee. 
The experiment was accepted without problems. 
All was fine, only minor comments were made and only a few short-notice actions were required. 

The procedure has become much more complicated and lengthy over time, though. 
(Something to keep in mind.)

Many thanks to the whole team !!
  32   Tue Feb 25 13:12:01 2025 SebastianDAQTeilungsverhältnis Ohmlabs HVS250
Hochspannung Maßstabsfaktoren

-140 kV >>> 248514
-160 kV >>> 248515
Attachment 1: Bericht_PTB_verbesserter_Teiler_2018.pdf
Bericht_PTB_verbesserter_Teiler_2018.pdf Bericht_PTB_verbesserter_Teiler_2018.pdf Bericht_PTB_verbesserter_Teiler_2018.pdf Bericht_PTB_verbesserter_Teiler_2018.pdf Bericht_PTB_verbesserter_Teiler_2018.pdf
Attachment 2: Maßstabsfaktoren.JPG
Maßstabsfaktoren.JPG
  31   Thu Feb 20 13:51:04 2025 Denise, SebatianDAQDAQ Channelliste
TDC Channels
Channel # Kanalbelegung
0 Laser busy Kr
1 Next Laserstep Kr
2
3
4 PMT Mitte
5 PMT Nord
6 XUV Anode
7
8
9
10 Bunch HF (common stop)
11 Shutter auf
12 Shutter zu
13 PD NW / 89
14 PD SW / 89
15 AFG rep / 89



Scaler Channels
Channel # Kanalbelegung
16 DC Trafo
17 Injection
18 U-ecool
19 I-ecool
20 P-gun
21 P-coll
22 common stop (Bunch HF)
23 1MHz clock
24 Extraktion
25 Shutter (10kHz)
26
27
28
29
30
31
  30   Thu Feb 20 13:37:30 2025 Denise, SebastianDAQCFD Schwellen DAQ
Schwellen fuer den CFD N842 (lascool)
Lasereinstellungen dabei waren: 0.5 Skalenteile -> 115ps Pulse
3.56654MHz Reprate
Stab (3.FV) wurde auf 1.2A betrieben
mit dem PBSc wurdee eine Laserleistung von 50mW in den ESR geschickt

CFD Channel Kanalbelegung Schwelle
0PD NW 14
1 PD SW 14
2 Shutter auf 20
3 Shutter zu 20
4 AFG rep 5
  29   Thu Feb 20 00:45:00 2025 DanyalGeneralDoppler width of M1 transition
According to Paul Indelicato

transition 3P1 -> 3P0  in 78Kr32+ : 

A	2.48E+08	1/s   "transition rate" 
t	4.03E-09	s     "lifetime 3P1" 
dv	3.95E+07	Hz    "natural line width" 
dE	1.63E-07	eV    "width in eV" 

Now, if we assume a final dp/p (after electron cooling) of the ion beam of ~1E-5, 
this corresponds to a longitudinal temperature of ~16000 K (~1.4 eV). 
This implies a Doppler width of ~26 GHz (ION frame). 

This is broader than the width of the laser pulse in the ION frame (between 4.4 and 23.4 GHz). 

This, in turn, implies that we should "see" the transition if we make 0.0024 nm steps in the UV (LAB frame). 
The transition would then be something like 2-3 steps wide. 

If we use this stepsize, and scan from 257.175 nm to 257.675 nm (LAB frame), we need 207 steps. 

If one step takes 1 minute (using 1 ESR cycle and fixed laser wavelength per cycle), one scan takes ~3.5 hours ! 
I think that is too long. 

For this procedure, the ESR cycle (injection -> next injection) must be faster. 
Also the time for the laser to change the wavelength must be fast, but that should be OK. 
We should try to get to sth. like 1 hour per scan. 
  28   Mon Feb 17 13:49:20 2025 DeniseGeneralFrequenzbreiten Ionensystem
Berechnung der Frequenzbreiten des Lasers im Ionensystem 
Werte für die Frequezbreiten des Lasers aus Dissertation Benedikt Langfeld übernommen

v=0.644c 

Pulsdauer = 115ps
Frequenzbreite (FWHM) = 10.87GHz
Frequenzbreite Ionensystem = 23.36GHz

Pulsdauer = 240ps
Frequenzbreite (FWHM) = 4.19GHz
Frequenzbreite Ionensystem = 9.00GHz

Pulsdauer = 400ps
Frequenzbreite (FWHM) = 2.82GHz
Frequenzbreite Ionensystem = 6.06GHz

Pulsdauer = 735ps
Frequenzbreite (FWHM) = 2.05GHz
Frequenzbreite Ionensystem = 4.41GHz
  27   Fri Feb 14 16:41:54 2025 Sebastian GeneralVerkabelung PD ESR
Verkabelung von den Photodioden im ESR. Sind auch gelabelt.

PD gehen zuerst einmal in den FAMP (10x Verstärkung) 

PD NW -> ESR Panel 4 
PD SW -> ESR Panel 7 

XUV LED -> ESR Panel 8 
Attachment 1: IMG_8368.jpeg
IMG_8368.jpeg
  26   Fri Feb 14 16:24:22 2025 DeniseGeneralStrahlprofil UV im ESR
Erstes Bild ist das Strahlptofil in Höhe des Nord 
towers. 
Zweites Bild (mit Millimeterpapier) und Video zeigt 
das Strahlprofils etwa in Höhe des XUV Detektors im 
ESR. Das Zittern ist dabei durch anstoßen am Nord 
Tower entstanden, wichtig hier zu sehen ist das 
„Atmen“ des Strahlprofils, welches dabei allerdings 
örtlich fest bleibt. 
Attachment 1: IMG_8360.jpeg
IMG_8360.jpeg
Attachment 2: 796a74f2-ec76-4396-82e5-9d3209b36612.jpeg
796a74f2-ec76-4396-82e5-9d3209b36612.jpeg
Attachment 3: 0bbd839a-f6cf-4fa4-a5a4-71d0c84044fa.mov
  25   Thu Feb 13 12:56:01 2025 DanyalGeneralESR access - monday (whole day) 8-18 hr
  24   Thu Feb 13 12:55:13 2025 DanyalGeneralESR access - tomorrow (Fri) 10:00 - 12:00
  23   Wed Feb 12 12:48:48 2025 DanyalGeneralESR access
We had the chance to shortly enter the ESR, between 10:45 and 11:30. 

- We repositioned the PSDs of the TEM system. 
- We could see the laser beam at the ESR exit window. 
- We used a wedge at the exit window and directed the two laser beams onto a power meter and a fast photodiode. 
- We photographed the XUV detector setup. 
- We documented the cabling at the central cable panel. 

In the ESR laser lab we could see the laser beam power on the power meter and we could also see 
the laser pulses on the fast photodiode (scope). 

We still need a longer break (~2 hrs) to further optimize the setup and all the settings. 
  22   Tue Feb 11 13:08:32 2025 DanyalAcceleratorlimited access to ESR this week
Hitrap will kein ESR Zugang erlauben bevor 22 Uhr. Außerdem merken die, wenn ESR-SIS nicht laufen, dass sie morgens immer 1 Stunde Zeit verlieren weil nicht "angefordert" wurde. Deswegen wollen sie, dass ESR-SIS nachts auch weiter Strahl anfordern, etwa 1 mal pro 5 Minuten. Aber wir finden schon noch eine Lücke um reinzugehen.


Hitrap does not want to allow ESR access before 10 p.m. In addition, when ESR-SIS is not running, they notice that they always lose 1 hour of time in the morning because nothing was "requested". That's why they want ESR-SIS to continue requesting beams at night, about once every 5 minutes. But we can still find a gap to go in.
  21   Tue Feb 11 10:35:38 2025 DanyalDetectorsXUV detector
Volker Hannen kommt am 19.2 (Mittags) und dann machen wir Abends den XUV Detektor bereid.


Volker Hannen is coming on February 19th (in the afternoon) and then we'll get the XUV detector ready in the evening.
  20   Tue Feb 11 10:32:01 2025 DanyalGeneralaccess to ESR
Es wird noch zwei 2 stündige Unterbrechungen am UNILAC geben. Wahrscheinlich diesen Freitag und nächste Woche Mittwoch oder Donnerstag. Am 25.2. wird dann die Gleitringdichtung der Pumpe am UNILAC gewechselt (8-18 Uhr) daher rechne ich mit Xe Strahl im ESR frühestes am 26.2.
Von HITRAP habe ich erfahren, dass wir Abends/Nachts was im Ring machen dürfen, ab 22 Uhr. z.B. am 17.2, 19.2 und 21.2.  Aber ich denke vorher (diese Woche) geht es bestimmt auch mal.

There will be two 2-hour breaks at the UNILAC. Probably this Friday and next week Wednesday or Thursday. On February 25th the mechanical seal of the pump on the UNILAC will be changed (8 a.m. to 6 p.m.) so I expect the Xe beam in the ESR on February 26th at the earliest.
I found out from HITRAP that we can do something in the ring in the evening/night, from 10 p.m. onwards. For example on February 17th, February 19th and February 21st.  But I think there will be some possibilities already this week. 
  19   Tue Feb 11 10:30:21 2025 DanyalGenerallaser valves are open
Die Laserventile im ESR sind nun offen. Der Laserstrahl darf durchgeschossen werden.

The "laser valves" (NW and SW) are now open.  The laser beam may pass through the ESR. 
  18   Thu Feb 6 12:54:29 2025 DanyalAcceleratorHV-divider for electron cooler connected - readout is running
The HV-divider has been connected to the E-Cooler. 

The readout (by Konstantin) is running too. 

Konstantin will prepare the inclusion into our DAQ.
Attachment 1: HV_divider_readout.jpg
HV_divider_readout.jpg
  17   Thu Feb 6 11:03:26 2025 DanyalGeneralpressures in laser window sections NW and SW
The pressures, read by the UHV technician, are : 

E01VVLF1T (E01VMLF1X):           5.84e-10mbar
E02VVLF2T (E02VMLF2X):           2.91e-10mbar

In 2019 the pressures were sth like 2e-9 mbar. They are much lower now. 

Note: The valves LF1 and LF2 may only be opened after ESR extraction to HITRAP has been set up! 
Most likely this is Monday 10 Feb. 


The UHV technician is: 
Mr. Semjon Strohmenger 
Phone: +49 6159 71 1765 
PSA: 7651 
  16   Wed Feb 5 14:05:53 2025 Rodolfo, SebastianAnalysisSchirm / Iris ESR Entrance Fenster
Wegen defekt von Kippschirm wurde das alte System zum Verstellen der Iris und des Schirms für das ESR Eintrittsfensters genutzt.

Fleischmann Trafo:

Slot 1 -> Schirm verfahren
Slot 3 -> Iris verfahren
  15   Wed Feb 5 13:50:28 2025 SebastianAcceleratorTIF Modul - Injektions-/ Extraktionssignale
TIF Moduleinstellung
Remark: Die EV H und EV L Einstellungen sind die Eventnummern in Hexadezimal Basis 16

Injektionssignal
Super Zyklus          3 (steht für ESR)
MA                    2 (Maschine 2)
EV H                  3
EV L                  1 


Extraktionssignal (Richtung Hitrap) 
Super Zyklus          3 (steht für ESR)
MA                    7 (steht für Subchain 7)
EV H                  4 (steht für Event 72 in hexadezimal = 48)
EV L                  8 (steht für Event 72 in hexadezimal = 48)

Das Extraktionssignal muss für unseres ESR Pattern noch angepasst werden.
Der "Code" 3748 wurde nur genutzt, um für die Laser - DAQ Einstellungen das Extraktionssignal zu erhalten.
Prinzip: Nach Extraktion wird der Laser auf einen Wellenlängenschritt weiter eingestellt.
ELOG V3.1.5-fc6679b