Entry  Sat Mar 1 18:32:05 2025, Jens, Runs, Laser während Scan 
18:30Uhr:
Timestamp im Programm hat gehangen. Kurz im Anzeigefeld markieren und es ging wieder.
Laser Step Counter hat gehangen und lag mit 4Go 20 Steps auseinander. Da es früh genug auffiel wurde der Number of Steps von 52 auf 72 erhöht. Ist also
Entry  Sat Mar 1 11:40:17 2025, Tamina, Florian, Konstantin, Calibration, Laser position Screenshot_2025-03-01_11-39-32.png
The laser position stayed perfect after the scans. See picture.
Entry  Fri Feb 28 19:33:03 2025, Volker, Detectors, Laser background 
We measured the laser background rate in the XUV detector.
The voltages applied were +300V at the MCP front, +2250V at the MCP anode.
Trigger threshold at ca. 40 mV.
Entry  Fri Feb 28 00:18:16 2025, Tamina, General, Kalibration UV Monitor im Laserlabor 
Der UV Monitor steht in der Reflexion des PBSC auf dem Lasertisch im Laserlabor. 

P_Transmission = (1-(0.9364 * cos^2(0.0349* (theta - 33.014)) + 0.06043)) *P_Reflex
Entry  Tue Jan 28 15:57:05 2025, Denise, General, Justage: ESR Labor Periskop Spiegel oben 
obere Schraube nach rechts drehen, dann bewegt sich Laserspot nach links 
selbe Schraube nach links dann Laser rechts
Entry  Mon Mar 3 08:02:00 2025, Danyal, Accelerator, Ionenenergie wird geändert 
Hallo zusammen,

wir haben den Scan-Bereich vom Laser - bei der Ionenenergie die wir jetzt haben - nun komplett ausgenützt. 
Entry  Fri Feb 28 13:53:06 2025, Tamina, Rodolfo, Runs, Ionenbunch Screenshot_Kr_Ionenbunch_2025-02-28_1350.pngScreenshot_Kr_Ionenbunch_2025-02-28_1445.png
Signal of PMT-Mitte. Recorded with an unstable ion bunch. With approx. 3E8 Ions

The second screenshot is showing a similar signal with 1.8E8 Ions
Entry  Thu Feb 6 12:54:29 2025, Danyal, Accelerator, HV-divider for electron cooler connected - readout is running HV_divider_readout.jpg
The HV-divider has been connected to the E-Cooler. 

The readout (by Konstantin) is running too. 
Entry  Thu Feb 27 09:59:08 2025, Konstantin, Calibration, HV-Divider 
The readout of the HV divider runs on a RaspberryPi inside the HV-cave of the electron cooler.
Starting from 2025-02-27 09:42, the divider ratio was changed from 248517 to 248515. Usually, the
readout should run continuously. If not, the following procedure must be applied:
Entry  Fri Feb 28 15:53:44 2025, Konstantin, Accelerator, HF Signal for Bunching 
The HF Bunching Frequency is: 1.7838817 MHz

Given (read) by the ESR operators in HKR.
Entry  Thu Feb 27 16:30:57 2025, Konstantin und Rodolfo, DAQ, HF Signal 
We have exchanged the HF-Signal in the DAQ Electronic: from Dummy-HF to the real HF from accelerator. 
Entry  Sat Mar 1 13:45:42 2025, Konstantin , DAQ, Gentec readout 
Gentec readout is now included in the general slow data acquisition. The issue was that the buffer of the power meter was not emptied fast enough. Now,
the data is read out permanently in a separate thread. 
Entry  Tue Mar 4 08:35:45 2025, Denise, Benedikt, Sebastian, Calibration, Gentec Kalibrationen im ESR 
Gentec NW (PH100 S:UV) OD.3 
[TABLE border="1"]
Leistung Gentec PH100 NW/ mW|0.475 | 0.961 | 1.2 | 1.36 | 1.66 | 1.8 | 2.18
Entry  Fri Feb 28 04:47:39 2025, Denise, Sebastian, Benedikt, Calibration, Gentec ESR Kalibration 
Unsicherheiten Gentecs im allgemeinen bei etwa 6.5%

Gentec NW (PH100 S:UV) OD.3 
Entry  Sun Mar 2 08:18:25 2025, Tamina, Volker, Florian, Rodolfo, General, Frühschicht  
Die Frühschicht scan weiter. Siehe Eintrag 75 (noch als Draft)
Entry  Mon Mar 3 08:33:00 2025, Tamina, Florian, Volker und Rodolfo , General, Fruehsicht 
Die Fruehsicht macht den letzten Scan, bevor die Ionenenergie geaendert wird.
Entry  Mon Feb 17 13:49:20 2025, Denise, General, Frequenzbreiten Ionensystem 
Berechnung der Frequenzbreiten des Lasers im Ionensystem 
Werte für die Frequezbreiten des Lasers aus Dissertation Benedikt Langfeld übernommen
Entry  Fri Feb 28 15:30:41 2025, Tamina, Rodolfo, Detectors, Fast PD Nord West 
We had an issue with the PD NW. The signal after the Rate Divider was not OK.

The problem was that the signal after the Delay Generator GG8000, the width, was too short. We have now changed the width and made similar to the PD SW.
Entry  Mon Mar 3 08:52:19 2025, Tamina, Rodolfo, Volker, Danyal, Florian, Runs, Extended Scan 257.0574 nm - 257.8022 nm bei hoeherer Ionenenergie 
Neue Umlauffrequenz: 1.78561578504741 MHz
rep freq laser :  3.57123157001
ecooler voltage: 157823 V
Entry  Mon Mar 3 13:00:51 2025, Danyal, General, Experiment will end tomorrow (04.03), at 08:00 AM precisely 
The experiment will end tomorrow (04.03) at 08:00 AM precisely. 
Everything (magnets, RF, etc) will be shut down.