  84   Mon Mar 30 08:56:04 2020 TinoRunsSecond measurement with 206Pb81+ beam at I_cooler = 10 mA
I_cooler = 10 mA

beam: 206Pb81+
energy: 400 MeV/u

purpose: data for the measurement of half life of 206Pb81+ due to its interaction with electron cooler and residual gas

Detector position (CsISiPHOS): 60 mm
Detector position (MWPC):     -10 mm

run start at 08:52 , run file: run0271.lmd 


_______________________________________________________________________________

time: 08:52

I_cooler = -01 mA

SIS particles: 3.4E8      
ESR particles:  4.7E6
        


NTCAp file:
SC_2020-03-30_08-54-42
IQ_2020-03-30_08-54-42
(saved in drive: p)

copy and repeat the above (below the line) every 30 minutes
  85   Mon Mar 30 09:26:37 2020 TinoRunsStopped Second measurement with 206Pb81+ beam at I_cooler = 10 mA
Stopped Data recording for Restart of the measurement

I_cooler = 10 mA

beam: 206Pb81+
energy: 400 MeV/u

purpose: data for the measurement of half life of 206Pb81+ due to its interaction with electron cooler and residual gas

Detector position (CsISiPHOS): 60 mm
Detector position (MWPC):     -10 mm

run start at 08:52 , run file: run0271.lmd 


_______________________________________________________________________________

time: 09:27

I_cooler = -01 mA

SIS particles: 3.4E8      
ESR particles:  2.5E5
        


NTCAp file:
SC_2020-03-30_08-54-42
IQ_2020-03-30_08-54-42
(saved in drive: p)

copy and repeat the above (below the line) every 30 minutes
  86   Mon Mar 30 09:34:02 2020 TinoRunsRestart of second measurement with 206Pb81+ beam at I_cooler = 10 mA
Stopped Data recording for Restart of the measurement

I_cooler = 10 mA

beam: 206Pb81+
energy: 400 MeV/u

purpose: data for the measurement of half life of 206Pb81+ due to its interaction with electron cooler and residual gas

Detector position (CsISiPHOS): 60 mm
Detector position (MWPC):     -10 mm

run start at 09:28 , run file: run0272.lmd 


_______________________________________________________________________________

time: 09:28

I_cooler = -01 mA

SIS particles: 3.1E8      
ESR particles:  5.7E6
        


NTCAp file:
SC_2020-03-30_09-28-14
IQ_2020-03-30_09-28-14
(saved in drive: p)

copy and repeat the above (below the line) every 30 minutes
  87   Mon Mar 30 10:01:31 2020 TinoRunsSecond measurement with 206Pb81+ beam at I_cooler = 10 mA

I_cooler = 10 mA

beam: 206Pb81+
energy: 400 MeV/u

purpose: data for the measurement of half life of 206Pb81+ due to its interaction with electron cooler and residual gas

Detector position (CsISiPHOS): 60 mm
Detector position (MWPC):     -10 mm

run start at 09:28 , run file: run0272.lmd 


_______________________________________________________________________________

time: 10:00

I_cooler = -01 mA
      
ESR particles:  2.8E5
        


NTCAp file:
SC_2020-03-30_09-28-14
IQ_2020-03-30_09-28-14
(saved in drive: p)

copy and repeat the above (below the line) every 30 minutes
  88   Mon Mar 30 10:32:01 2020 TinoRunsSecond measurement with 206Pb81+ beam at I_cooler = 10 mA

I_cooler = 10 mA

beam: 206Pb81+
energy: 400 MeV/u

purpose: data for the measurement of half life of 206Pb81+ due to its interaction with electron cooler and residual gas

Detector position (CsISiPHOS): 60 mm
Detector position (MWPC):     -10 mm

run start at 09:28 , run file: run0272.lmd 


_______________________________________________________________________________

time: 10:31

I_cooler = -01 mA

SIS particles: 3.1E8      
ESR particles:  2.7E5
        


NTCAp file:
SC_2020-03-30_09-28-14
IQ_2020-03-30_09-28-14
(saved in drive: p)

copy and repeat the above (below the line) every 30 minutes
  89   Mon Mar 30 10:57:43 2020 TinoRunsSecond measurement with 206Pb81+ beam at I_cooler = 10 mA

I_cooler = 10 mA

beam: 206Pb81+
energy: 400 MeV/u

purpose: data for the measurement of half life of 206Pb81+ due to its interaction with electron cooler and residual gas

Detector position (CsISiPHOS): 60 mm
Detector position (MWPC):     -10 mm

run start at 09:28 , run file: run0272.lmd 


_______________________________________________________________________________

time: 10:57

I_cooler = -01 mA

SIS particles: 3.1E8      
ESR particles:  2.7E5
        


NTCAp file:
SC_2020-03-30_09-28-14
IQ_2020-03-30_09-28-14
(saved in drive: p)

copy and repeat the above (below the line) every 30 minutes
  90   Mon Mar 30 12:12:25 2020 ruijiuGeneralInstruction of NTCAP for shift person
To start a new run with NTCap, we need to retart NTCAP. 
1. Stop labview program. 
2. Start labview program and start a new run.
  91   Mon Mar 30 12:16:04 2020 TinoRunsSecond measurement with 206Pb81+ beam at I_cooler = 10 mA

I_cooler = 10 mA

beam: 206Pb81+
energy: 400 MeV/u

purpose: data for the measurement of half life of 206Pb81+ due to its interaction with electron cooler and residual gas

Detector position (CsISiPHOS): 60 mm
Detector position (MWPC):     -10 mm

run start at 09:28 , run file: run0272.lmd 


_______________________________________________________________________________

time: 12:00

I_cooler = -01 mA

SIS particles: 3.1E8      
ESR particles:  2.3E5
        


NTCAp file:
SC_2020-03-30_09-28-14
IQ_2020-03-30_09-28-14
(saved in drive: p)

copy and repeat the above (below the line) every 30 minutes
  92   Mon Mar 30 12:33:04 2020 TinoRunsSecond measurement with 206Pb81+ beam at I_cooler = 10 mA

I_cooler = 10 mA

beam: 206Pb81+
energy: 400 MeV/u

purpose: data for the measurement of half life of 206Pb81+ due to its interaction with electron cooler and residual gas

Detector position (CsISiPHOS): 60 mm
Detector position (MWPC):     -10 mm

run start at 09:28 , run file: run0272.lmd 


_______________________________________________________________________________

time: 12:30

I_cooler = -01 mA

SIS particles: 3.1E8      
ESR particles:  2.8E5
        


NTCAp file:
SC_2020-03-30_09-28-14
IQ_2020-03-30_09-28-14
(saved in drive: p)

copy and repeat the above (below the line) every 30 minutes
  93   Mon Mar 30 13:02:49 2020 TinoRunsSecond measurement with 206Pb81+ beam at I_cooler = 10 mA

I_cooler = 10 mA

beam: 206Pb81+
energy: 400 MeV/u

purpose: data for the measurement of half life of 206Pb81+ due to its interaction with electron cooler and residual gas

Detector position (CsISiPHOS): 60 mm
Detector position (MWPC):     -10 mm

run start at 09:28 , run file: run0272.lmd 


_______________________________________________________________________________

time: 13:00

I_cooler = -01 mA

SIS particles: 3.1E8      
ESR particles:  4.5E5
(Intensity rised @ around 12:52, without a new shot from SIS)
        


NTCAp file:
SC_2020-03-30_09-28-14
IQ_2020-03-30_09-28-14
(saved in drive: p)

copy and repeat the above (below the line) every 30 minutes
  94   Mon Mar 30 13:31:58 2020 TinoRunsSecond measurement with 206Pb81+ beam at I_cooler = 10 mA

I_cooler = 10 mA

beam: 206Pb81+
energy: 400 MeV/u

purpose: data for the measurement of half life of 206Pb81+ due to its interaction with electron cooler and residual gas

Detector position (CsISiPHOS): 60 mm
Detector position (MWPC):     -10 mm

run start at 09:28 , run file: run0272.lmd 


_______________________________________________________________________________

time: 13:30

I_cooler = -01 mA

SIS particles: 3.1E8      
ESR particles:  4.0E5
        


NTCAp file:
SC_2020-03-30_09-28-14
IQ_2020-03-30_09-28-14
(saved in drive: p)

copy and repeat the above (below the line) every 30 minutes
  95   Mon Mar 30 13:40:57 2020 ruijiuDetectorsHV of Inner detector
  96   Mon Mar 30 13:41:20 2020 ruijiuDetectorsHV of Inner detector
  97   Mon Mar 30 14:02:41 2020 TinoRunsSecond measurement with 206Pb81+ beam at I_cooler = 10 mA

I_cooler = 10 mA

beam: 206Pb81+
energy: 400 MeV/u

purpose: data for the measurement of half life of 206Pb81+ due to its interaction with electron cooler and residual gas

Detector position (CsISiPHOS): 60 mm
Detector position (MWPC):     -10 mm

run start at 09:28 , run file: run0272.lmd 
run stop at  14Ö23 , run file: run0272.lmd


_______________________________________________________________________________

time: 14:00

I_cooler = -01 mA

SIS particles: 3.1E8      
ESR particles:  4.2E5
        


NTCAp file:
SC_2020-03-30_09-28-14
IQ_2020-03-30_09-28-14
(saved in drive: p)

copy and repeat the above (below the line) every 30 minutes
  98   Mon Mar 30 14:24:46 2020 ruijiuDAQstop taking data.
1. stop run0272.lmd.
2. stop NTCAP. 
  NTCAp file:
  SC_2020-03-30_09-28-14
  IQ_2020-03-30_09-28-14
  (saved in drive: p)
  99   Mon Mar 30 14:48:02 2020 ruijiu and TinoDetectorsThe leakcurrent of SiPad is over the setting limit.
Because the 
leakcurrent of 
6 Sipads is 
over the 
setting limit, 
the alarm of 
HV module is 
sounding.
Ruijiu 
increase the 
limit to 1500 
uA.
  100   Mon Mar 30 15:03:58 2020 ruijiu Tino, Yuri, Lasoro, Nik and UliDetectorsthe gas bottle is empty.
The gas bottle was empty. Tino replaced 
the previous old bottle with 50 bar gas left. 

Yuri is getting new gas bottles.

Yuri got a new gas bottle. We repalced 
the the new gas bottle. There is 200 
bar gas left now. 
The old gas bottle with 50 bar gas 
left was kept inside ESR.
The empty gas bottle was removed from 
ESR.  

The outlet volve was decreased by 30 degree.
Since the value is almost 0, we can not read the value of outlet volve.
  101   Mon Mar 30 16:25:05 2020 ruijiuDetectorsturn down the HV
turn down the hv and pull out the detector.  

suggest by 
Hi everybody,

is there now much more rate on  the detectors?

Perhaps it helps to turn down the bias for a while and to take the detectors out of the beam, by pulling the pocket out?

And then see if the leakage current comes down again.
good luck, Thomas

***********************************************
Dr. Thomas Faestermann  |  Tel:  +49-89-2723868
Physik Department E12   |
Techn. Univ. Muenchen   |  Mob:  +49-1626193388
  102   Mon Mar 30 16:37:53 2020 ruijiuAnalysis 
  103   Mon Mar 30 20:51:06 2020 RaganDetectorsBias Voltage for Si-pads
The limit of the leakage current for all Si pads was increased from 1500 nA to 5000nA and 
leakage current was recorded for different Bias Voltages (32V, 35V, 40V, 45V, 50V, 55V 
and 60V).
Since, I_leakage remains < 3 uA for 60 V, it is decided to increase the Bias Voltage for 
all the Si-pads from 32 V to 60 V.
For DSSD and CsI, bias voltages remains same (-50V and -160V respectively).