HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018 E0028
  Bound beta decay experiment on Tl-205, Page 5 of 9  Not logged in ELOG logo
Entry  Mon Mar 30 08:56:04 2020, Tino, Runs, Second measurement with 206Pb81+ beam at I_cooler = 10 mA  
I_cooler = 10 mA

beam: 206Pb81+
energy: 400 MeV/u

purpose: data for the measurement of half life of 206Pb81+ due to its interaction with electron cooler and residual gas

Detector position (CsISiPHOS): 60 mm
Detector position (MWPC):     -10 mm

run start at 08:52 , run file: run0271.lmd 


_______________________________________________________________________________

time: 08:52

I_cooler = -01 mA

SIS particles: 3.4E8      
ESR particles:  4.7E6
        


NTCAp file:
SC_2020-03-30_08-54-42
IQ_2020-03-30_08-54-42
(saved in drive: p)

copy and repeat the above (below the line) every 30 minutes
Entry  Mon Mar 30 09:26:37 2020, Tino, Runs, Stopped Second measurement with 206Pb81+ beam at I_cooler = 10 mA  
Stopped Data recording for Restart of the measurement

I_cooler = 10 mA

beam: 206Pb81+
energy: 400 MeV/u

purpose: data for the measurement of half life of 206Pb81+ due to its interaction with electron cooler and residual gas

Detector position (CsISiPHOS): 60 mm
Detector position (MWPC):     -10 mm

run start at 08:52 , run file: run0271.lmd 


_______________________________________________________________________________

time: 09:27

I_cooler = -01 mA

SIS particles: 3.4E8      
ESR particles:  2.5E5
        


NTCAp file:
SC_2020-03-30_08-54-42
IQ_2020-03-30_08-54-42
(saved in drive: p)

copy and repeat the above (below the line) every 30 minutes
Entry  Mon Mar 30 09:34:02 2020, Tino, Runs, Restart of second measurement with 206Pb81+ beam at I_cooler = 10 mA  
Stopped Data recording for Restart of the measurement

I_cooler = 10 mA

beam: 206Pb81+
energy: 400 MeV/u

purpose: data for the measurement of half life of 206Pb81+ due to its interaction with electron cooler and residual gas

Detector position (CsISiPHOS): 60 mm
Detector position (MWPC):     -10 mm

run start at 09:28 , run file: run0272.lmd 


_______________________________________________________________________________

time: 09:28

I_cooler = -01 mA

SIS particles: 3.1E8      
ESR particles:  5.7E6
        


NTCAp file:
SC_2020-03-30_09-28-14
IQ_2020-03-30_09-28-14
(saved in drive: p)

copy and repeat the above (below the line) every 30 minutes
Entry  Mon Mar 30 10:01:31 2020, Tino, Runs, Second measurement with 206Pb81+ beam at I_cooler = 10 mA  

I_cooler = 10 mA

beam: 206Pb81+
energy: 400 MeV/u

purpose: data for the measurement of half life of 206Pb81+ due to its interaction with electron cooler and residual gas

Detector position (CsISiPHOS): 60 mm
Detector position (MWPC):     -10 mm

run start at 09:28 , run file: run0272.lmd 


_______________________________________________________________________________

time: 10:00

I_cooler = -01 mA
      
ESR particles:  2.8E5
        


NTCAp file:
SC_2020-03-30_09-28-14
IQ_2020-03-30_09-28-14
(saved in drive: p)

copy and repeat the above (below the line) every 30 minutes
Entry  Mon Mar 30 10:32:01 2020, Tino, Runs, Second measurement with 206Pb81+ beam at I_cooler = 10 mA  

I_cooler = 10 mA

beam: 206Pb81+
energy: 400 MeV/u

purpose: data for the measurement of half life of 206Pb81+ due to its interaction with electron cooler and residual gas

Detector position (CsISiPHOS): 60 mm
Detector position (MWPC):     -10 mm

run start at 09:28 , run file: run0272.lmd 


_______________________________________________________________________________

time: 10:31

I_cooler = -01 mA

SIS particles: 3.1E8      
ESR particles:  2.7E5
        


NTCAp file:
SC_2020-03-30_09-28-14
IQ_2020-03-30_09-28-14
(saved in drive: p)

copy and repeat the above (below the line) every 30 minutes
Entry  Mon Mar 30 10:57:43 2020, Tino, Runs, Second measurement with 206Pb81+ beam at I_cooler = 10 mA  

I_cooler = 10 mA

beam: 206Pb81+
energy: 400 MeV/u

purpose: data for the measurement of half life of 206Pb81+ due to its interaction with electron cooler and residual gas

Detector position (CsISiPHOS): 60 mm
Detector position (MWPC):     -10 mm

run start at 09:28 , run file: run0272.lmd 


_______________________________________________________________________________

time: 10:57

I_cooler = -01 mA

SIS particles: 3.1E8      
ESR particles:  2.7E5
        


NTCAp file:
SC_2020-03-30_09-28-14
IQ_2020-03-30_09-28-14
(saved in drive: p)

copy and repeat the above (below the line) every 30 minutes
Entry  Mon Mar 30 12:12:25 2020, ruijiu, General, Instruction of NTCAP for shift person Instruction_of_NTCAP_for_shift_person.png
To start a new run with NTCap, we need to retart NTCAP. 
1. Stop labview program. 
2. Start labview program and start a new run.
Entry  Mon Mar 30 12:16:04 2020, Tino, Runs, Second measurement with 206Pb81+ beam at I_cooler = 10 mA  

I_cooler = 10 mA

beam: 206Pb81+
energy: 400 MeV/u

purpose: data for the measurement of half life of 206Pb81+ due to its interaction with electron cooler and residual gas

Detector position (CsISiPHOS): 60 mm
Detector position (MWPC):     -10 mm

run start at 09:28 , run file: run0272.lmd 


_______________________________________________________________________________

time: 12:00

I_cooler = -01 mA

SIS particles: 3.1E8      
ESR particles:  2.3E5
        


NTCAp file:
SC_2020-03-30_09-28-14
IQ_2020-03-30_09-28-14
(saved in drive: p)

copy and repeat the above (below the line) every 30 minutes
Entry  Mon Mar 30 12:33:04 2020, Tino, Runs, Second measurement with 206Pb81+ beam at I_cooler = 10 mA  

I_cooler = 10 mA

beam: 206Pb81+
energy: 400 MeV/u

purpose: data for the measurement of half life of 206Pb81+ due to its interaction with electron cooler and residual gas

Detector position (CsISiPHOS): 60 mm
Detector position (MWPC):     -10 mm

run start at 09:28 , run file: run0272.lmd 


_______________________________________________________________________________

time: 12:30

I_cooler = -01 mA

SIS particles: 3.1E8      
ESR particles:  2.8E5
        


NTCAp file:
SC_2020-03-30_09-28-14
IQ_2020-03-30_09-28-14
(saved in drive: p)

copy and repeat the above (below the line) every 30 minutes
Entry  Mon Mar 30 13:02:49 2020, Tino, Runs, Second measurement with 206Pb81+ beam at I_cooler = 10 mA  

I_cooler = 10 mA

beam: 206Pb81+
energy: 400 MeV/u

purpose: data for the measurement of half life of 206Pb81+ due to its interaction with electron cooler and residual gas

Detector position (CsISiPHOS): 60 mm
Detector position (MWPC):     -10 mm

run start at 09:28 , run file: run0272.lmd 


_______________________________________________________________________________

time: 13:00

I_cooler = -01 mA

SIS particles: 3.1E8      
ESR particles:  4.5E5
(Intensity rised @ around 12:52, without a new shot from SIS)
        


NTCAp file:
SC_2020-03-30_09-28-14
IQ_2020-03-30_09-28-14
(saved in drive: p)

copy and repeat the above (below the line) every 30 minutes
Entry  Mon Mar 30 13:31:58 2020, Tino, Runs, Second measurement with 206Pb81+ beam at I_cooler = 10 mA  

I_cooler = 10 mA

beam: 206Pb81+
energy: 400 MeV/u

purpose: data for the measurement of half life of 206Pb81+ due to its interaction with electron cooler and residual gas

Detector position (CsISiPHOS): 60 mm
Detector position (MWPC):     -10 mm

run start at 09:28 , run file: run0272.lmd 


_______________________________________________________________________________

time: 13:30

I_cooler = -01 mA

SIS particles: 3.1E8      
ESR particles:  4.0E5
        


NTCAp file:
SC_2020-03-30_09-28-14
IQ_2020-03-30_09-28-14
(saved in drive: p)

copy and repeat the above (below the line) every 30 minutes
Entry  Mon Mar 30 13:40:57 2020, ruijiu, Detectors, HV of Inner detector IMG_20200330_133909.jpg
 
Entry  Mon Mar 30 13:41:20 2020, ruijiu, Detectors, HV of Inner detector IMG_20200330_133909.jpg
 
Entry  Mon Mar 30 14:02:41 2020, Tino, Runs, Second measurement with 206Pb81+ beam at I_cooler = 10 mA  

I_cooler = 10 mA

beam: 206Pb81+
energy: 400 MeV/u

purpose: data for the measurement of half life of 206Pb81+ due to its interaction with electron cooler and residual gas

Detector position (CsISiPHOS): 60 mm
Detector position (MWPC):     -10 mm

run start at 09:28 , run file: run0272.lmd 
run stop at  14Ö23 , run file: run0272.lmd


_______________________________________________________________________________

time: 14:00

I_cooler = -01 mA

SIS particles: 3.1E8      
ESR particles:  4.2E5
        


NTCAp file:
SC_2020-03-30_09-28-14
IQ_2020-03-30_09-28-14
(saved in drive: p)

copy and repeat the above (below the line) every 30 minutes
Entry  Mon Mar 30 14:24:46 2020, ruijiu, DAQ, stop taking data. 
1. stop run0272.lmd.
2. stop NTCAP. 
  NTCAp file:
  SC_2020-03-30_09-28-14
  IQ_2020-03-30_09-28-14
  (saved in drive: p)
Entry  Mon Mar 30 14:48:02 2020, ruijiu and Tino, Detectors, The leakcurrent of SiPad is over the setting limit. Screenshot_20200330-145551.jpgScreenshot_20200330-145544.jpgIMG_20200330_144915.jpg
Because the 
leakcurrent of 
6 Sipads is 
over the 
setting limit, 
the alarm of 
HV module is 
sounding.
Ruijiu 
increase the 
limit to 1500 
uA.
Entry  Mon Mar 30 15:03:58 2020, ruijiu Tino, Yuri, Lasoro, Nik and Uli, Detectors, the gas bottle is empty. IMG_20200330_152735_mh1585575780057.jpg
The gas bottle was empty. Tino replaced 
the previous old bottle with 50 bar gas left. 

Yuri is getting new gas bottles.

Yuri got a new gas bottle. We repalced 
the the new gas bottle. There is 200 
bar gas left now. 
The old gas bottle with 50 bar gas 
left was kept inside ESR.
The empty gas bottle was removed from 
ESR.  

The outlet volve was decreased by 30 degree.
Since the value is almost 0, we can not read the value of outlet volve.
Entry  Mon Mar 30 16:25:05 2020, ruijiu, Detectors, turn down the HV 
turn down the hv and pull out the detector.  

suggest by 
Hi everybody,

is there now much more rate on  the detectors?

Perhaps it helps to turn down the bias for a while and to take the detectors out of the beam, by pulling the pocket out?

And then see if the leakage current comes down again.
good luck, Thomas

***********************************************
Dr. Thomas Faestermann  |  Tel:  +49-89-2723868
Physik Department E12   |
Techn. Univ. Muenchen   |  Mob:  +49-1626193388
Entry  Mon Mar 30 16:37:53 2020, ruijiu, Analysis,  
 
Entry  Mon Mar 30 20:51:06 2020, Ragan, Detectors, Bias Voltage for Si-pads Si_pad_bias-voltage.ods
The limit of the leakage current for all Si pads was increased from 1500 nA to 5000nA and 
leakage current was recorded for different Bias Voltages (32V, 35V, 40V, 45V, 50V, 55V 
and 60V).
Since, I_leakage remains < 3 uA for 60 V, it is decided to increase the Bias Voltage for 
all the Si-pads from 32 V to 60 V.
For DSSD and CsI, bias voltages remains same (-50V and -160V respectively).
ELOG V3.1.5-fc6679b