Mon Mar 30 08:56:04 2020, Tino, Runs, Second measurement with 206Pb81+ beam at I_cooler = 10 mA
|
I_cooler = 10 mA
beam: 206Pb81+
energy: 400 MeV/u
purpose: data for the measurement of half life of 206Pb81+ due to its interaction with electron cooler and residual gas
Detector position (CsISiPHOS): 60 mm
Detector position (MWPC): -10 mm
run start at 08:52 , run file: run0271.lmd
_______________________________________________________________________________
time: 08:52
I_cooler = -01 mA
SIS particles: 3.4E8
ESR particles: 4.7E6
NTCAp file:
SC_2020-03-30_08-54-42
IQ_2020-03-30_08-54-42
(saved in drive: p)
copy and repeat the above (below the line) every 30 minutes |
Mon Mar 30 09:26:37 2020, Tino, Runs, Stopped Second measurement with 206Pb81+ beam at I_cooler = 10 mA
|
Stopped Data recording for Restart of the measurement
I_cooler = 10 mA
beam: 206Pb81+
energy: 400 MeV/u
purpose: data for the measurement of half life of 206Pb81+ due to its interaction with electron cooler and residual gas
Detector position (CsISiPHOS): 60 mm
Detector position (MWPC): -10 mm
run start at 08:52 , run file: run0271.lmd
_______________________________________________________________________________
time: 09:27
I_cooler = -01 mA
SIS particles: 3.4E8
ESR particles: 2.5E5
NTCAp file:
SC_2020-03-30_08-54-42
IQ_2020-03-30_08-54-42
(saved in drive: p)
copy and repeat the above (below the line) every 30 minutes |
Mon Mar 30 09:34:02 2020, Tino, Runs, Restart of second measurement with 206Pb81+ beam at I_cooler = 10 mA
|
Stopped Data recording for Restart of the measurement
I_cooler = 10 mA
beam: 206Pb81+
energy: 400 MeV/u
purpose: data for the measurement of half life of 206Pb81+ due to its interaction with electron cooler and residual gas
Detector position (CsISiPHOS): 60 mm
Detector position (MWPC): -10 mm
run start at 09:28 , run file: run0272.lmd
_______________________________________________________________________________
time: 09:28
I_cooler = -01 mA
SIS particles: 3.1E8
ESR particles: 5.7E6
NTCAp file:
SC_2020-03-30_09-28-14
IQ_2020-03-30_09-28-14
(saved in drive: p)
copy and repeat the above (below the line) every 30 minutes |
Mon Mar 30 10:01:31 2020, Tino, Runs, Second measurement with 206Pb81+ beam at I_cooler = 10 mA
|
I_cooler = 10 mA
beam: 206Pb81+
energy: 400 MeV/u
purpose: data for the measurement of half life of 206Pb81+ due to its interaction with electron cooler and residual gas
Detector position (CsISiPHOS): 60 mm
Detector position (MWPC): -10 mm
run start at 09:28 , run file: run0272.lmd
_______________________________________________________________________________
time: 10:00
I_cooler = -01 mA
ESR particles: 2.8E5
NTCAp file:
SC_2020-03-30_09-28-14
IQ_2020-03-30_09-28-14
(saved in drive: p)
copy and repeat the above (below the line) every 30 minutes |
Mon Mar 30 10:32:01 2020, Tino, Runs, Second measurement with 206Pb81+ beam at I_cooler = 10 mA
|
I_cooler = 10 mA
beam: 206Pb81+
energy: 400 MeV/u
purpose: data for the measurement of half life of 206Pb81+ due to its interaction with electron cooler and residual gas
Detector position (CsISiPHOS): 60 mm
Detector position (MWPC): -10 mm
run start at 09:28 , run file: run0272.lmd
_______________________________________________________________________________
time: 10:31
I_cooler = -01 mA
SIS particles: 3.1E8
ESR particles: 2.7E5
NTCAp file:
SC_2020-03-30_09-28-14
IQ_2020-03-30_09-28-14
(saved in drive: p)
copy and repeat the above (below the line) every 30 minutes |
Mon Mar 30 10:57:43 2020, Tino, Runs, Second measurement with 206Pb81+ beam at I_cooler = 10 mA
|
I_cooler = 10 mA
beam: 206Pb81+
energy: 400 MeV/u
purpose: data for the measurement of half life of 206Pb81+ due to its interaction with electron cooler and residual gas
Detector position (CsISiPHOS): 60 mm
Detector position (MWPC): -10 mm
run start at 09:28 , run file: run0272.lmd
_______________________________________________________________________________
time: 10:57
I_cooler = -01 mA
SIS particles: 3.1E8
ESR particles: 2.7E5
NTCAp file:
SC_2020-03-30_09-28-14
IQ_2020-03-30_09-28-14
(saved in drive: p)
copy and repeat the above (below the line) every 30 minutes |
Mon Mar 30 12:12:25 2020, ruijiu, General, Instruction of NTCAP for shift person
|
To start a new run with NTCap, we need to retart NTCAP.
1. Stop labview program.
2. Start labview program and start a new run. |
Mon Mar 30 12:16:04 2020, Tino, Runs, Second measurement with 206Pb81+ beam at I_cooler = 10 mA
|
I_cooler = 10 mA
beam: 206Pb81+
energy: 400 MeV/u
purpose: data for the measurement of half life of 206Pb81+ due to its interaction with electron cooler and residual gas
Detector position (CsISiPHOS): 60 mm
Detector position (MWPC): -10 mm
run start at 09:28 , run file: run0272.lmd
_______________________________________________________________________________
time: 12:00
I_cooler = -01 mA
SIS particles: 3.1E8
ESR particles: 2.3E5
NTCAp file:
SC_2020-03-30_09-28-14
IQ_2020-03-30_09-28-14
(saved in drive: p)
copy and repeat the above (below the line) every 30 minutes |
Mon Mar 30 12:33:04 2020, Tino, Runs, Second measurement with 206Pb81+ beam at I_cooler = 10 mA
|
I_cooler = 10 mA
beam: 206Pb81+
energy: 400 MeV/u
purpose: data for the measurement of half life of 206Pb81+ due to its interaction with electron cooler and residual gas
Detector position (CsISiPHOS): 60 mm
Detector position (MWPC): -10 mm
run start at 09:28 , run file: run0272.lmd
_______________________________________________________________________________
time: 12:30
I_cooler = -01 mA
SIS particles: 3.1E8
ESR particles: 2.8E5
NTCAp file:
SC_2020-03-30_09-28-14
IQ_2020-03-30_09-28-14
(saved in drive: p)
copy and repeat the above (below the line) every 30 minutes |
Mon Mar 30 13:02:49 2020, Tino, Runs, Second measurement with 206Pb81+ beam at I_cooler = 10 mA
|
I_cooler = 10 mA
beam: 206Pb81+
energy: 400 MeV/u
purpose: data for the measurement of half life of 206Pb81+ due to its interaction with electron cooler and residual gas
Detector position (CsISiPHOS): 60 mm
Detector position (MWPC): -10 mm
run start at 09:28 , run file: run0272.lmd
_______________________________________________________________________________
time: 13:00
I_cooler = -01 mA
SIS particles: 3.1E8
ESR particles: 4.5E5
(Intensity rised @ around 12:52, without a new shot from SIS)
NTCAp file:
SC_2020-03-30_09-28-14
IQ_2020-03-30_09-28-14
(saved in drive: p)
copy and repeat the above (below the line) every 30 minutes |
Mon Mar 30 13:31:58 2020, Tino, Runs, Second measurement with 206Pb81+ beam at I_cooler = 10 mA
|
I_cooler = 10 mA
beam: 206Pb81+
energy: 400 MeV/u
purpose: data for the measurement of half life of 206Pb81+ due to its interaction with electron cooler and residual gas
Detector position (CsISiPHOS): 60 mm
Detector position (MWPC): -10 mm
run start at 09:28 , run file: run0272.lmd
_______________________________________________________________________________
time: 13:30
I_cooler = -01 mA
SIS particles: 3.1E8
ESR particles: 4.0E5
NTCAp file:
SC_2020-03-30_09-28-14
IQ_2020-03-30_09-28-14
(saved in drive: p)
copy and repeat the above (below the line) every 30 minutes |
Mon Mar 30 13:40:57 2020, ruijiu, Detectors, HV of Inner detector
|
|
Mon Mar 30 13:41:20 2020, ruijiu, Detectors, HV of Inner detector
|
|
Mon Mar 30 14:02:41 2020, Tino, Runs, Second measurement with 206Pb81+ beam at I_cooler = 10 mA
|
I_cooler = 10 mA
beam: 206Pb81+
energy: 400 MeV/u
purpose: data for the measurement of half life of 206Pb81+ due to its interaction with electron cooler and residual gas
Detector position (CsISiPHOS): 60 mm
Detector position (MWPC): -10 mm
run start at 09:28 , run file: run0272.lmd
run stop at 14Ö23 , run file: run0272.lmd
_______________________________________________________________________________
time: 14:00
I_cooler = -01 mA
SIS particles: 3.1E8
ESR particles: 4.2E5
NTCAp file:
SC_2020-03-30_09-28-14
IQ_2020-03-30_09-28-14
(saved in drive: p)
copy and repeat the above (below the line) every 30 minutes |
Mon Mar 30 14:24:46 2020, ruijiu, DAQ, stop taking data.
|
1. stop run0272.lmd.
2. stop NTCAP.
NTCAp file:
SC_2020-03-30_09-28-14
IQ_2020-03-30_09-28-14
(saved in drive: p) |
Mon Mar 30 14:48:02 2020, ruijiu and Tino, Detectors, The leakcurrent of SiPad is over the setting limit.
|
Because the
leakcurrent of
6 Sipads is
over the
setting limit,
the alarm of
HV module is
sounding.
Ruijiu
increase the
limit to 1500
uA. |
Mon Mar 30 15:03:58 2020, ruijiu Tino, Yuri, Lasoro, Nik and Uli, Detectors, the gas bottle is empty.
|
The gas bottle was empty. Tino replaced
the previous old bottle with 50 bar gas left.
Yuri is getting new gas bottles.
Yuri got a new gas bottle. We repalced
the the new gas bottle. There is 200
bar gas left now.
The old gas bottle with 50 bar gas
left was kept inside ESR.
The empty gas bottle was removed from
ESR.
The outlet volve was decreased by 30 degree.
Since the value is almost 0, we can not read the value of outlet volve. |
Mon Mar 30 16:25:05 2020, ruijiu, Detectors, turn down the HV
|
turn down the hv and pull out the detector.
suggest by
Hi everybody,
is there now much more rate on the detectors?
Perhaps it helps to turn down the bias for a while and to take the detectors out of the beam, by pulling the pocket out?
And then see if the leakage current comes down again.
good luck, Thomas
***********************************************
Dr. Thomas Faestermann | Tel: +49-89-2723868
Physik Department E12 |
Techn. Univ. Muenchen | Mob: +49-1626193388 |
Mon Mar 30 16:37:53 2020, ruijiu, Analysis,
|
|
Mon Mar 30 20:51:06 2020, Ragan, Detectors, Bias Voltage for Si-pads
|
The limit of the leakage current for all Si pads was increased from 1500 nA to 5000nA and
leakage current was recorded for different Bias Voltages (32V, 35V, 40V, 45V, 50V, 55V
and 60V).
Since, I_leakage remains < 3 uA for 60 V, it is decided to increase the Bias Voltage for
all the Si-pads from 32 V to 60 V.
For DSSD and CsI, bias voltages remains same (-50V and -160V respectively). |