  58   Sun Mar 29 21:02:43 2020 ruijiuAnalysisdecay line of 205Tl beam
decay line of 205Tl beam with Ar gas jet target on and S6 slit at -8mm
Ar gas density~1e12 atoms/cm2
Attachment 1: main_s6_neg8_202003292059.png
main_s6_neg8_202003292059.png
Attachment 2: main_s6_neg8_202003292059.root
  60   Sun Mar 29 21:39:27 2020 ruijiuAnalysisIonization cross section and capture cross section for h-206Pb + Ar
Ionization cross section = 1320 barn.
Capture cross section = 557+137+44.5=735.5.
sigma_Ion/sigma_cap=1.79
Attachment 1: 202003292141.png
202003292141.png
  90   Mon Mar 30 12:12:25 2020 ruijiuGeneralInstruction of NTCAP for shift person
To start a new run with NTCap, we need to retart NTCAP. 
1. Stop labview program. 
2. Start labview program and start a new run.
Attachment 1: Instruction_of_NTCAP_for_shift_person.png
Instruction_of_NTCAP_for_shift_person.png
  95   Mon Mar 30 13:40:57 2020 ruijiuDetectorsHV of Inner detector
Attachment 1: IMG_20200330_133909.jpg
IMG_20200330_133909.jpg
  96   Mon Mar 30 13:41:20 2020 ruijiuDetectorsHV of Inner detector
Attachment 1: IMG_20200330_133909.jpg
IMG_20200330_133909.jpg
  98   Mon Mar 30 14:24:46 2020 ruijiuDAQstop taking data.
1. stop run0272.lmd.
2. stop NTCAP. 
  NTCAp file:
  SC_2020-03-30_09-28-14
  IQ_2020-03-30_09-28-14
  (saved in drive: p)
  101   Mon Mar 30 16:25:05 2020 ruijiuDetectorsturn down the HV
turn down the hv and pull out the detector.  

suggest by 
Hi everybody,

is there now much more rate on  the detectors?

Perhaps it helps to turn down the bias for a while and to take the detectors out of the beam, by pulling the pocket out?

And then see if the leakage current comes down again.
good luck, Thomas

***********************************************
Dr. Thomas Faestermann  |  Tel:  +49-89-2723868
Physik Department E12   |
Techn. Univ. Muenchen   |  Mob:  +49-1626193388
  102   Mon Mar 30 16:37:53 2020 ruijiuAnalysis 
  106   Tue Mar 31 15:30:43 2020 ruijiuAnalysisThe beam position on DSSD with different conditions on on the dEE spectrum
I analyzed the first measurement with 206Pb81+ beam, run 250-267. 
Attachment 1: 202003311453.png
202003311453.png
Attachment 2: 202003311508.png
202003311508.png
Attachment 3: 202003311509.png
202003311509.png
Attachment 4: 202003311510.png
202003311510.png
Attachment 5: 202003311511.png
202003311511.png
Attachment 6: 202003311512.png
202003311512.png
  111   Wed Apr 1 01:02:24 2020 ruijiuDetectorsthe detector cable is disconnected.
  112   Wed Apr 1 01:05:36 2020 ruijiuDetectorsdetector position fixed at 60 mm
The detector cable was disconnected . We entered and reconnected the cable.
Attachment 1: VID_20200401_010416.mp4
  120   Thu Apr 2 12:52:25 2020 ruijiuGeneralsetup FRS with 2 g/cm2 degrader.
Yuri sets up FRS with 2 g/cm2 degrader (previous one is 1 g/cm2). The ESR is not changed. The contamination should be less now.
  171   Thu Apr 9 11:47:28 2020 ruijiu backup data
2020.4.8  12:00Started copying 
data。
2020.4.9    11:52 Copied 6.3TB 
data. The copy speed is 65-67MB/s.
2020.4.15 Bring the two USB hard disk to my office and connect them  to lxg1298 computer.
Send email 
Attachment 1: IMG_20200409_113155.jpg
IMG_20200409_113155.jpg
  172   Fri Apr 17 18:27:34 2020 ruijiuAnalysisbackup data
1.connect USB harddisk  to computer lxg1254 and  copy data to sever.

sync -rtmv --progress --checkSUD  /media/sdgl/data/ litv-exp@sbk0l9:/lustre/ap/litv-exp/2020-
04-14 e121_rchen/NTCAP/hd1

2020.05.02 08:06 copied 11.88 TB to server. 1.8TB rest on hard disk.
Attachment 1: IMG_20200417_162202.jpg
IMG_20200417_162202.jpg
  100   Mon Mar 30 15:03:58 2020 ruijiu Tino, Yuri, Lasoro, Nik and UliDetectorsthe gas bottle is empty.
The gas bottle was empty. Tino replaced 
the previous old bottle with 50 bar gas left. 

Yuri is getting new gas bottles.

Yuri got a new gas bottle. We repalced 
the the new gas bottle. There is 200 
bar gas left now. 
The old gas bottle with 50 bar gas 
left was kept inside ESR.
The empty gas bottle was removed from 
ESR.  

The outlet volve was decreased by 30 degree.
Since the value is almost 0, we can not read the value of outlet volve.
Attachment 1: IMG_20200330_152735_mh1585575780057.jpg
IMG_20200330_152735_mh1585575780057.jpg
  99   Mon Mar 30 14:48:02 2020 ruijiu and TinoDetectorsThe leakcurrent of SiPad is over the setting limit.
Because the 
leakcurrent of 
6 Sipads is 
over the 
setting limit, 
the alarm of 
HV module is 
sounding.
Ruijiu 
increase the 
limit to 1500 
uA.
Attachment 1: Screenshot_20200330-145551.jpg
Screenshot_20200330-145551.jpg
Attachment 2: Screenshot_20200330-145544.jpg
Screenshot_20200330-145544.jpg
Attachment 3: IMG_20200330_144915.jpg
IMG_20200330_144915.jpg
  15   Tue Feb 25 15:48:31 2020 ruijiu chenDAQImportant folders of e121 eperiments:
1. slow control:
1.1. program file:
litv-exp@ringpi:~/epics/mrcc_ioc/db $ 
1.2 user interface:
litv-exp@lxg1298:~/E121/scripts/GUIe121_new

2. mbs:
2.1 user interface
litv-exp@lxg1298:~/E121/scripts$ ./e121_gui.py
2.2 source 
R4L-47 litv-exp > pwd                                                                                                                                                                                                                        
/esr/usr/litv-exp/2020_e121/                                                                                                                                                                                             
3. go4:
litv-exp@lxg1298:~/2019Munich-Det/Go4-Muenchen-3V785AH-1V830-simple$ go4
4. experimenal data:
litv-exp@lxg1298:/data.local1/E121
  16   Tue Feb 25 15:49:58 2020 ruijiu chenDAQchanging scaling factor.
How to change scaling factor?
The parameters is in f_user.c file:
1. open the f_user.C with emacs :
R4L-47 litv-exp > emacs -nw f_user.c
2. change the parameters "MWPC_downscale". 
#define MWPC_downscale 0    // reduction = 2**MWPC_downscale                  
3.compile the program.
4. restart mbs.                                                                                                                                                               
  41   Sat Mar 28 16:12:25 2020 ruijiu, raganDAQenter ESR and turn on the HV crate.
  80   Mon Mar 30 07:02:15 2020 tinoRunsSecond measurement with 206Pb81+ beam at I_cooler = 20 mA
I_cooler = 20 mA

beam: 206Pb81+
energy: 400 MeV/u

purpose: data for the measurement of half life of 206Pb81+ due to its interaction with electron cooler and residual gas

Detector position (CsISiPHOS): 60 mm
Detector position (MWPC):     -10 mm

run start at 06:08 , run file: run0270.lmd  (0268.lmd and 0269.lmd are garbage files)
file running: 0270.lmd

_______________________________________________________________________________

time: 07:00

I_cooler = -01 mA

      
ESR particles:  2.2E5
        


NTCAp file:
SC_2020-03-30_06-11-08
IQ_2020-03-30_06-11-16
(saved in drive: p)

copy and repeat the above (below the line) every 30 minutes
ELOG V3.1.5-fc6679b