ID Date Author Category Subject Yeardown Text Attachments
  284   Fri Oct 23 12:58:12 2020 JanDetectorsTest of 2nd DSSSD (gen2)2021This is the documentation of the source tests
with the 2nd micron DSSSD of 2nd generation
(label 3288-17, thickness 529um)
 Si2_run001.jpgSi2_run002.jpg 
  285   Fri Nov 20 15:53:37 2020 JanAnalysisTest of 2nd DSSSD (gen2)2021Here is the analysis of the test runs with
the 2nd DSSD (gen2) using the quick_si_plots.c
script attached.
 quick_si_plots.ce127.hSi2_run001.ana.rootSi2_run002.ana.root 
  286   Tue Nov 24 15:35:49 2020 JanCalibrationDSSD X/Y channel mapping2021During the detector test [URL=https://elog.gsi.de/esr/E127/284]test
measurements[/URL] with alpha source the
allocation of the 16 X- and 16 Y-channels
  
  287   Tue Feb 2 13:59:00 2021 JanDetectorsDSSD installation and alignment2021The DSSD has been exchanged and aligned in

November 2020. After bakeout at 140°C 
  
  288   Wed Mar 31 13:52:04 2021 JanDetectorsBaF Na22 Testing2021On 30. and 31.03.2021 we did some testing
with the 6 BaF detectors.
  
  289   Wed Mar 31 13:57:27 2021 JanDetectorsBaF HV settings2021The setting below is a start value for final
gain matching with a full ADC range of 12MeV:
  
  290   Wed Apr 14 09:42:50 2021 JanDetectorsBaF HV gain matching2021The preliminary gain matching settings for
a full ADC range of ~12MeV:
 BaF_Na22_gain_match_v1.pngbaf_gain_match_v1.dat 
  291   Thu Apr 22 10:56:20 2021 JanDetectorstarget-chamber det. distances2021This is the lookup table for the distance
of each x-ray detector to the target.
See attached foto for explanation.
 E127_xray_distances.jpg 
  292   Wed Apr 28 09:12:47 2021 Jan, YuriDetectorsUI-diagram2021The UI-curve of the detector
U I
10 0.11
 UI_meas_curve.pdf 
  293   Thu Apr 29 20:32:52 2021 Jan GloriusCalibrationcalibration sources2021We use the following list of sources for calibration
of the Xray detectors:
  
  294   Thu Apr 29 20:34:09 2021 Jan GloriusCalibrationrun0003 - Xray90 calib Pb210 d=167.5mm2021Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
  
  295   Fri Apr 30 09:27:27 2021 Jan GloriusCalibrationrun0004 - Xray90 calib Pb210 d=167.5mm 2021Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
  
  296   Fri Apr 30 09:57:57 2021 Jan GloriusCalibrationrun0005 - Xray90 calib Pb210 d=167.5mm 2021Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
  
  297   Fri Apr 30 10:55:34 2021 Jan GloriusDAQXray multi-triggers - blocking gate2021From the CFD of Xray1 (90deg) we get multiple
trigger signals every ~5sec for unknown reasons.
  
  298   Fri Apr 30 11:04:31 2021 Jan GloriusCalibrationrun0006 - Xray90 calib Pb210 d=167.5mm 2021Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
  
  299   Fri Apr 30 13:07:23 2021 Jan GloriusCalibrationrun0008 - Xray90 calib Pb210 d=167.5mm 2021Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
  
  300   Fri Apr 30 13:10:44 2021 Jan GloriusCalibrationrun0007 - crap2021crap run   
  301   Fri Apr 30 13:56:59 2021 Jan GloriusCalibrationrun0009 - Xray90 calib Am241 d=167.5mm 2021Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
  
  302   Fri Apr 30 14:35:11 2021 Jan GloriusCalibrationrun0010 - Xray90 calib Ba133 d=167.5mm 2021Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
  
  303   Fri Apr 30 14:48:01 2021 Jan GloriusCalibrationrun0011 - Xray90 calib Ba133_low d=167.5mm 2021Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
  