Entry  Fri Oct 23 12:58:12 2020, Jan, Detectors, Test of 2nd DSSSD (gen2), 2021 Si2_run001.jpgSi2_run002.jpg
This is the documentation of the source tests with the 2nd micron DSSSD of 2nd generation (label 3288-17, thickness 529um)

The detector is put into vacuum (~5e-6 mbar) in our test chamber. The source is positioned a few cm above (see fotos).
    Reply  Fri Nov 20 15:53:37 2020, Jan, Analysis, Test of 2nd DSSSD (gen2), 2021 quick_si_plots.ce127.hSi2_run001.ana.rootSi2_run002.ana.root
Here is the analysis of the test runs with the 2nd DSSD (gen2) using the quick_si_plots.c script attached.

All 32 Si channels are working with acceptable performance. 
Entry  Tue Nov 24 15:35:49 2020, Jan, Calibration, DSSD X/Y channel mapping, 2021 
During the detector test [URL=https://elog.gsi.de/esr/E127/284]test measurements[/URL] with alpha source the allocation of the 16 X- and 16 Y-channels has
been checked.
Entry  Tue Feb 2 13:59:00 2021, Jan, Detectors, DSSD installation and alignment, 2021 
The DSSD has been exchanged and aligned in 
November 2020. After bakeout at 140°C 
(externally, 120°C at internal temp. Sensor) 
Entry  Wed Mar 31 13:52:04 2021, Jan, Detectors, BaF Na22 Testing, 2021 
On 30. and 31.03.2021 we did some testing with the 6 BaF detectors.

Trigger & threshold:
Entry  Wed Mar 31 13:57:27 2021, Jan, Detectors, BaF HV settings, 2021 
The setting below is a start value for final gain matching with a full ADC range of 12MeV:

[TABLE border="1"]
Entry  Wed Apr 14 09:42:50 2021, Jan, Detectors, BaF HV gain matching, 2021 BaF_Na22_gain_match_v1.pngbaf_gain_match_v1.dat
The preliminary gain matching settings for a full ADC range of ~12MeV:

[TABLE border="1"]
Entry  Thu Apr 22 10:56:20 2021, Jan, Detectors, target-chamber det. distances, 2021 E127_xray_distances.jpg
This is the lookup table for the distance of each x-ray detector to the target.
See attached foto for explanation.
Entry  Wed Apr 28 09:12:47 2021, Jan, Yuri, Detectors, UI-diagram, 2021 UI_meas_curve.pdf
The UI-curve of the detector
U I
10 0.11
Entry  Thu Apr 29 20:32:52 2021, Jan Glorius, Calibration, calibration sources, 2021 
We use the following list of sources for calibration of the Xray detectors:

- 210Pb [40.1 kBq (4%), 01.10.2020 12:00 UTC] [URL=https://elog.gsi.de/esr/X-Ray+Lab/201006_131944/20-006_210Pb_40kBq.pdf]SpecSheet[/URL]
Entry  Thu Apr 29 20:34:09 2021, Jan Glorius, Calibration, run0003 - Xray90 calib Pb210 d=167.5mm, 2021 
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
Entry  Fri Apr 30 09:27:27 2021, Jan Glorius, Calibration, run0004 - Xray90 calib Pb210 d=167.5mm , 2021 
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
Entry  Fri Apr 30 09:57:57 2021, Jan Glorius, Calibration, run0005 - Xray90 calib Pb210 d=167.5mm , 2021 
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
Entry  Fri Apr 30 10:55:34 2021, Jan Glorius, DAQ, Xray multi-triggers - blocking gate, 2021 
From the CFD of Xray1 (90deg) we get multiple trigger signals every ~5sec for unknown reasons. 
No hint from the preamp signals.
Entry  Fri Apr 30 11:04:31 2021, Jan Glorius, Calibration, run0006 - Xray90 calib Pb210 d=167.5mm , 2021 
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
Entry  Fri Apr 30 13:07:23 2021, Jan Glorius, Calibration, run0008 - Xray90 calib Pb210 d=167.5mm , 2021 
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
Entry  Fri Apr 30 13:10:44 2021, Jan Glorius, Calibration, run0007 - crap, 2021 
crap run
Entry  Fri Apr 30 13:56:59 2021, Jan Glorius, Calibration, run0009 - Xray90 calib Am241 d=167.5mm , 2021 
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
Entry  Fri Apr 30 14:35:11 2021, Jan Glorius, Calibration, run0010 - Xray90 calib Ba133 d=167.5mm , 2021 
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
Entry  Fri Apr 30 14:48:01 2021, Jan Glorius, Calibration, run0011 - Xray90 calib Ba133_low d=167.5mm , 2021 
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg