HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018 E0028 E0038
  proton-capture on 118Te, Page 14 of 29  Not logged in ELOG logo
New entries since:Thu Jan 1 01:00:00 1970
ID Date Author Category Subject Year
  316   Thu May 6 11:03:46 2021 Jan GloriusDAQCabling Documentation2021
  315   Thu May 6 09:40:37 2021 Jan GloriusCalibrationrun0020 - Xray35 calib Ba133_hi d=334mm2021
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 133Ba_hi
Distance: 334mm
Start time: 9:42:24 - 06.05.2021
Stop time: 14:03:44 - 06.05.2021

file name: e127b_run0020.lmd
avrg. rate: 90 Hz
dead-time:  1%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
  314   Wed May 5 21:22:27 2021 Jan GloriusCalibrationrun0019 - Xray35 calib Pb210 d=334mm2021
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 210Pb
Distance: 334mm
Start time: 21:22:06 - 05.05.2021
Stop time:   8:55:21 - 06.05.2021

file name: e127b_run0019.lmd
avrg. rate: 5 Hz
dead-time:  0%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.

Many rates spikes, check carefully!
  313   Wed May 5 14:29:02 2021 Jan GloriusCalibrationrun0017 crap run2021
  312   Wed May 5 14:24:57 2021 Jan GloriusCalibrationrun0018 - Xray35 calib Pb210 d=334mm2021
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 210Pb
Distance: 334mm
Start time: 14:28:26 - 05.05.2021
Stop time:  21:05:00 - 05.05.2021

file name: e127b_run0018.lmd
avrg. rate: 5 Hz
dead-time:  0%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.

There are many rate spikes in the first hour of measurement.
  311   Wed May 5 10:02:27 2021 Jan GloriusCalibrationrun0016 - Xray35 calib Ba133_hi d=334mm2021
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 133Ba_hi
Distance: 334mm
Start time: 10:02:07 - 05.05.2021
Stop time:  14:13:23 - 05.05.2021

file name: e127b_run0016.lmd
avrg. rate: 90 Hz
dead-time:  1%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
  310   Tue May 4 21:18:15 2021 Jan GloriusCalibrationrun0015 - Xray35 calib Am241_hi d=334mm2021
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 241Am_hi
Distance: 334mm
Start time: 21:17:44 - 04.05.2021
Stop time:   8:42:09 - 05.05.2021

file name: e127b_run0015.lmd
avrg. rate: 25 Hz
dead-time:  0%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
  309   Tue May 4 16:04:42 2021 Jan GloriusCalibrationrun0014 - Xray35 calib Am241_hi d=334mm2021
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 241Am_hi
Distance: 334mm
Start time: 16:04:21 - 04.05.2021
Stop time:  20:41:45 - 04.05.2021

file name: e127b_run0014.lmd
avrg. rate: 25 Hz
dead-time:  0%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
  308   Tue May 4 14:46:15 2021 Jan GloriusCalibrationrun0013 - Xray35 calib Am241_hi d=334mm2021
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 241Am_hi
Distance: 334mm
Start time: 14:45:51 - 04.05.2021
Stop time:  15:47:48 - 04.05.2021

file name: e127b_run0013.lmd
avrg. rate: 25 Hz
dead-time:  0%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
  307   Tue May 4 14:34:51 2021 Jan GloriusDetectorsXray weight control induces signals2021
When the readout/control of the det. weight control via serial connection is active, there is a clear signal induced in the Xray detectors for each command sent.

The solution for now is, that we activate weight control NOT DURING MEASUREMENT!
  306   Tue May 4 14:31:51 2021 Jan GloriusCalibrationrun0012 - Xray35 calib Am241_hi d=334mm2021
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 241Am_hi
Distance: 334mm
Start time: 14:31:24 - 04.05.2021
Stop time:  14:42:13 - 04.05.2021

file name: e127b_run0012.lmd
avrg. rate: 25 Hz
dead-time:  0%


Threshold < 10keV

The scale serial communication is off.

Stopped due to noise rate...
  305   Mon May 3 14:55:44 2021 Jan GloriusCalibrationXray energy calib. table2021
Here is the table of energy calibration lines from Am241 for all Xray detectors for several measurements.
Also the linear fit parameters [ E(ch) = A*ch + B ] are given at the bottom of the table.

E [keV] xray90 (GEM1800) xray35 (GLP) xray145 (GEM3000)
lab (28.04.21) lab (04.05.21) lab (11.05.21)
59.54 2402.45 ch 2399.07 ch 2423.24 ch
26.34 1107.38 ch 1116.65 ch 1127.94 ch
13.95 624.33 ch 638.38 ch 645.39 ch
slope A 0.0256387 keV/ch 0.0258923 keV/ch 0.0256408 keV/ch
offset B -2.05477 keV -2.57638 keV -2.59114 keV
  304   Sat May 1 22:34:16 2021 Jan GloriusCalibrationefficiency xray 90deg2021
the attached efficiency curve for the 90 deg. xray detector is made with the runs
e127b_run0008.lmd
e127b_run0009.lmd
e127b_run0010.lmd

All of them were taken in the lab (not in ESR).

This is a very preliminary analysis!
E.g. deadtime has been estimated and was not cleanly determined.
  303   Fri Apr 30 14:48:01 2021 Jan GloriusCalibrationrun0011 - Xray90 calib Ba133_low d=167.5mm 2021
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
Source: 133Ba_low
Distance: 167.5mm
Start time: 14:47:40 - 30.04.2021
Stop time:  15:50:12 - 30.04.2021

file name: e127b_run0011.lmd
avrg. rate: 310 Hz
dead-time:  3%

Now with blocking gate to get rid of multiple triggers.
Dead time determination using: adt/bdt

Threshold at ~10keV

We have some external noise on the signal. 
  302   Fri Apr 30 14:35:11 2021 Jan GloriusCalibrationrun0010 - Xray90 calib Ba133 d=167.5mm 2021
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
Source: 133Ba_high
Distance: 167.5mm
Start time: 14:34:51 - 30.04.2021
Stop time:  14:45:42 - 30.04.2021

file name: e127b_run0010.lmd
avrg. rate: 2600 Hz
dead-time:  16%

Now with blocking gate to get rid of multiple triggers.
Dead time determination using: adt/bdt

Threshold at ~10keV

We have some external noise on the signal. 
  301   Fri Apr 30 13:56:59 2021 Jan GloriusCalibrationrun0009 - Xray90 calib Am241 d=167.5mm 2021
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
Source: 241Am_high
Distance: 167.5mm
Start time: 13:56:33 - 30.04.2021
Stop time:  14:28:53 - 30.04.2021

file name: e127b_run0009.lmd
avrg. rate: 600Hz
dead-time:  4%

Now with blocking gate to get rid of multiple triggers.
Dead time determination using: adt/bdt

Threshold at ~10keV

We have some external noise on the signal. 
  300   Fri Apr 30 13:10:44 2021 Jan GloriusCalibrationrun0007 - crap2021
crap run
  299   Fri Apr 30 13:07:23 2021 Jan GloriusCalibrationrun0008 - Xray90 calib Pb210 d=167.5mm 2021
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
Source: 210Pb
Distance: 167.5mm
Start time: 13:08:35 - 30.04.2021
Stop time:  13:49:56 - 30.04.2021

file name: e127b_run0008.lmd
avrg. rate: 60Hz
dead-time:  1%

Now with blocking gate to get rid of multiple triggers.
Dead time determination using: adt/bdt

Threshold at ~10keV

We have some external noise on the signal. 
  298   Fri Apr 30 11:04:31 2021 Jan GloriusCalibrationrun0006 - Xray90 calib Pb210 d=167.5mm 2021
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
Source: 210Pb
Distance: 167.5mm
Start time: 11:04:03 - 30.04.2021
Stop time:  11:53:59 - 30.04.2021

file name: e127b_run0006.lmd
avrg. rate: 50Hz
dead-time:  1%

Now with blocking gate to get rid of multiple triggers.
Dead time determination using: adt/bdt

Threshold at ~20keV

Last seconds of run may not be usable: DAQ node2 has been deactivated shortly by mistake.
  297   Fri Apr 30 10:55:34 2021 Jan GloriusDAQXray multi-triggers - blocking gate2021
From the CFD of Xray1 (90deg) we get multiple trigger signals every ~5sec for unknown reasons. 
No hint from the preamp signals.

The solution now is to have an extended blocking gate active in the trlo-config of node 2.
There is now a trigger_stretch of 2500ns for all Xray triggers:

trig_stretch(1)  = 2500ns;
trig_stretch(2)  = 2500ns;
trig_stretch(3)  = 2500ns;

This prevents the multiple triggers to reach the lmu (before_dt scalers)
However, this will also lead to a very small loss of good signal triggers that overlay within 2.5 us.

This has to be taken into account for deadtime determination.
ELOG V3.1.5-fc6679b