  proton-capture on 118Te, Page 14 of 29  Not logged in ELOG logo
Entry  Thu May 6 11:03:46 2021, Jan Glorius, DAQ, Cabling Documentation, 2021 e127b_cabling_docu_06_05_2021_.pdfe127b_2021_cabling_documentation.xlsx
 
Entry  Thu May 6 09:40:37 2021, Jan Glorius, Calibration, run0020 - Xray35 calib Ba133_hi d=334mm, 2021 
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Entry  Wed May 5 21:22:27 2021, Jan Glorius, Calibration, run0019 - Xray35 calib Pb210 d=334mm, 2021 
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Entry  Wed May 5 14:29:02 2021, Jan Glorius, Calibration, run0017 crap run, 2021 
 
Entry  Wed May 5 14:24:57 2021, Jan Glorius, Calibration, run0018 - Xray35 calib Pb210 d=334mm, 2021 
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Entry  Wed May 5 10:02:27 2021, Jan Glorius, Calibration, run0016 - Xray35 calib Ba133_hi d=334mm, 2021 
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Entry  Tue May 4 21:18:15 2021, Jan Glorius, Calibration, run0015 - Xray35 calib Am241_hi d=334mm, 2021 
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Entry  Tue May 4 16:04:42 2021, Jan Glorius, Calibration, run0014 - Xray35 calib Am241_hi d=334mm, 2021 
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Entry  Tue May 4 14:46:15 2021, Jan Glorius, Calibration, run0013 - Xray35 calib Am241_hi d=334mm, 2021 
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Entry  Tue May 4 14:34:51 2021, Jan Glorius, Detectors, Xray weight control induces signals, 2021 
When the readout/control of the det. weight control via serial connection is active, there is a clear signal induced in the Xray detectors for each command
sent.
Entry  Tue May 4 14:31:51 2021, Jan Glorius, Calibration, run0012 - Xray35 calib Am241_hi d=334mm, 2021 
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Entry  Mon May 3 14:55:44 2021, Jan Glorius, Calibration, Xray energy calib. table, 2021 
Here is the table of energy calibration lines from Am241 for all Xray detectors for several measurements.
Also the linear fit parameters [ E(ch) = A*ch + B ] are given at the bottom of the table.
Entry  Sat May 1 22:34:16 2021, Jan Glorius, Calibration, efficiency xray 90deg, 2021 90deg_eff.png90deg_eff.dat
the attached efficiency curve for the 90 deg. xray detector is made with the runs
e127b_run0008.lmd
e127b_run0009.lmd
Entry  Fri Apr 30 14:48:01 2021, Jan Glorius, Calibration, run0011 - Xray90 calib Ba133_low d=167.5mm , 2021 
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
Entry  Fri Apr 30 14:35:11 2021, Jan Glorius, Calibration, run0010 - Xray90 calib Ba133 d=167.5mm , 2021 
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
Entry  Fri Apr 30 13:56:59 2021, Jan Glorius, Calibration, run0009 - Xray90 calib Am241 d=167.5mm , 2021 
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
Entry  Fri Apr 30 13:10:44 2021, Jan Glorius, Calibration, run0007 - crap, 2021 
crap run
Entry  Fri Apr 30 13:07:23 2021, Jan Glorius, Calibration, run0008 - Xray90 calib Pb210 d=167.5mm , 2021 
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
Entry  Fri Apr 30 11:04:31 2021, Jan Glorius, Calibration, run0006 - Xray90 calib Pb210 d=167.5mm , 2021 
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
Entry  Fri Apr 30 10:55:34 2021, Jan Glorius, DAQ, Xray multi-triggers - blocking gate, 2021 
From the CFD of Xray1 (90deg) we get multiple trigger signals every ~5sec for unknown reasons. 
No hint from the preamp signals.