HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018 E0028 E0038
  proton-capture on 118Te, Page 14 of 29  Not logged in ELOG logo
New entries since:Thu Jan 1 01:00:00 1970
Entry  Thu May 6 11:03:46 2021, Jan Glorius, DAQ, Cabling Documentation, 2021 e127b_cabling_docu_06_05_2021_.pdfe127b_2021_cabling_documentation.xlsx
 
Entry  Thu May 6 09:40:37 2021, Jan Glorius, Calibration, run0020 - Xray35 calib Ba133_hi d=334mm, 2021 
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 133Ba_hi
Distance: 334mm
Start time: 9:42:24 - 06.05.2021
Stop time: 14:03:44 - 06.05.2021

file name: e127b_run0020.lmd
avrg. rate: 90 Hz
dead-time:  1%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
Entry  Wed May 5 21:22:27 2021, Jan Glorius, Calibration, run0019 - Xray35 calib Pb210 d=334mm, 2021 
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 210Pb
Distance: 334mm
Start time: 21:22:06 - 05.05.2021
Stop time:   8:55:21 - 06.05.2021

file name: e127b_run0019.lmd
avrg. rate: 5 Hz
dead-time:  0%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.

Many rates spikes, check carefully!
Entry  Wed May 5 14:29:02 2021, Jan Glorius, Calibration, run0017 crap run, 2021 
 
Entry  Wed May 5 14:24:57 2021, Jan Glorius, Calibration, run0018 - Xray35 calib Pb210 d=334mm, 2021 
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 210Pb
Distance: 334mm
Start time: 14:28:26 - 05.05.2021
Stop time:  21:05:00 - 05.05.2021

file name: e127b_run0018.lmd
avrg. rate: 5 Hz
dead-time:  0%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.

There are many rate spikes in the first hour of measurement.
Entry  Wed May 5 10:02:27 2021, Jan Glorius, Calibration, run0016 - Xray35 calib Ba133_hi d=334mm, 2021 
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 133Ba_hi
Distance: 334mm
Start time: 10:02:07 - 05.05.2021
Stop time:  14:13:23 - 05.05.2021

file name: e127b_run0016.lmd
avrg. rate: 90 Hz
dead-time:  1%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
Entry  Tue May 4 21:18:15 2021, Jan Glorius, Calibration, run0015 - Xray35 calib Am241_hi d=334mm, 2021 
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 241Am_hi
Distance: 334mm
Start time: 21:17:44 - 04.05.2021
Stop time:   8:42:09 - 05.05.2021

file name: e127b_run0015.lmd
avrg. rate: 25 Hz
dead-time:  0%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
Entry  Tue May 4 16:04:42 2021, Jan Glorius, Calibration, run0014 - Xray35 calib Am241_hi d=334mm, 2021 
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 241Am_hi
Distance: 334mm
Start time: 16:04:21 - 04.05.2021
Stop time:  20:41:45 - 04.05.2021

file name: e127b_run0014.lmd
avrg. rate: 25 Hz
dead-time:  0%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
Entry  Tue May 4 14:46:15 2021, Jan Glorius, Calibration, run0013 - Xray35 calib Am241_hi d=334mm, 2021 
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 241Am_hi
Distance: 334mm
Start time: 14:45:51 - 04.05.2021
Stop time:  15:47:48 - 04.05.2021

file name: e127b_run0013.lmd
avrg. rate: 25 Hz
dead-time:  0%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
Entry  Tue May 4 14:34:51 2021, Jan Glorius, Detectors, Xray weight control induces signals, 2021 
When the readout/control of the det. weight control via serial connection is active, there is a clear signal induced in the Xray detectors for each command sent.

The solution for now is, that we activate weight control NOT DURING MEASUREMENT!
Entry  Tue May 4 14:31:51 2021, Jan Glorius, Calibration, run0012 - Xray35 calib Am241_hi d=334mm, 2021 
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 241Am_hi
Distance: 334mm
Start time: 14:31:24 - 04.05.2021
Stop time:  14:42:13 - 04.05.2021

file name: e127b_run0012.lmd
avrg. rate: 25 Hz
dead-time:  0%


Threshold < 10keV

The scale serial communication is off.

Stopped due to noise rate...
Entry  Mon May 3 14:55:44 2021, Jan Glorius, Calibration, Xray energy calib. table, 2021 
Here is the table of energy calibration lines from Am241 for all Xray detectors for several measurements.
Also the linear fit parameters [ E(ch) = A*ch + B ] are given at the bottom of the table.

E [keV] xray90 (GEM1800) xray35 (GLP) xray145 (GEM3000)
lab (28.04.21) lab (04.05.21) lab (11.05.21)
59.54 2402.45 ch 2399.07 ch 2423.24 ch
26.34 1107.38 ch 1116.65 ch 1127.94 ch
13.95 624.33 ch 638.38 ch 645.39 ch
slope A 0.0256387 keV/ch 0.0258923 keV/ch 0.0256408 keV/ch
offset B -2.05477 keV -2.57638 keV -2.59114 keV
Entry  Sat May 1 22:34:16 2021, Jan Glorius, Calibration, efficiency xray 90deg, 2021 90deg_eff.png90deg_eff.dat
the attached efficiency curve for the 90 deg. xray detector is made with the runs
e127b_run0008.lmd
e127b_run0009.lmd
e127b_run0010.lmd

All of them were taken in the lab (not in ESR).

This is a very preliminary analysis!
E.g. deadtime has been estimated and was not cleanly determined.
Entry  Fri Apr 30 14:48:01 2021, Jan Glorius, Calibration, run0011 - Xray90 calib Ba133_low d=167.5mm , 2021 
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
Source: 133Ba_low
Distance: 167.5mm
Start time: 14:47:40 - 30.04.2021
Stop time:  15:50:12 - 30.04.2021

file name: e127b_run0011.lmd
avrg. rate: 310 Hz
dead-time:  3%

Now with blocking gate to get rid of multiple triggers.
Dead time determination using: adt/bdt

Threshold at ~10keV

We have some external noise on the signal. 
Entry  Fri Apr 30 14:35:11 2021, Jan Glorius, Calibration, run0010 - Xray90 calib Ba133 d=167.5mm , 2021 
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
Source: 133Ba_high
Distance: 167.5mm
Start time: 14:34:51 - 30.04.2021
Stop time:  14:45:42 - 30.04.2021

file name: e127b_run0010.lmd
avrg. rate: 2600 Hz
dead-time:  16%

Now with blocking gate to get rid of multiple triggers.
Dead time determination using: adt/bdt

Threshold at ~10keV

We have some external noise on the signal. 
Entry  Fri Apr 30 13:56:59 2021, Jan Glorius, Calibration, run0009 - Xray90 calib Am241 d=167.5mm , 2021 
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
Source: 241Am_high
Distance: 167.5mm
Start time: 13:56:33 - 30.04.2021
Stop time:  14:28:53 - 30.04.2021

file name: e127b_run0009.lmd
avrg. rate: 600Hz
dead-time:  4%

Now with blocking gate to get rid of multiple triggers.
Dead time determination using: adt/bdt

Threshold at ~10keV

We have some external noise on the signal. 
Entry  Fri Apr 30 13:10:44 2021, Jan Glorius, Calibration, run0007 - crap, 2021 
crap run
Entry  Fri Apr 30 13:07:23 2021, Jan Glorius, Calibration, run0008 - Xray90 calib Pb210 d=167.5mm , 2021 
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
Source: 210Pb
Distance: 167.5mm
Start time: 13:08:35 - 30.04.2021
Stop time:  13:49:56 - 30.04.2021

file name: e127b_run0008.lmd
avrg. rate: 60Hz
dead-time:  1%

Now with blocking gate to get rid of multiple triggers.
Dead time determination using: adt/bdt

Threshold at ~10keV

We have some external noise on the signal. 
Entry  Fri Apr 30 11:04:31 2021, Jan Glorius, Calibration, run0006 - Xray90 calib Pb210 d=167.5mm , 2021 
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
Source: 210Pb
Distance: 167.5mm
Start time: 11:04:03 - 30.04.2021
Stop time:  11:53:59 - 30.04.2021

file name: e127b_run0006.lmd
avrg. rate: 50Hz
dead-time:  1%

Now with blocking gate to get rid of multiple triggers.
Dead time determination using: adt/bdt

Threshold at ~20keV

Last seconds of run may not be usable: DAQ node2 has been deactivated shortly by mistake.
Entry  Fri Apr 30 10:55:34 2021, Jan Glorius, DAQ, Xray multi-triggers - blocking gate, 2021 
From the CFD of Xray1 (90deg) we get multiple trigger signals every ~5sec for unknown reasons. 
No hint from the preamp signals.

The solution now is to have an extended blocking gate active in the trlo-config of node 2.
There is now a trigger_stretch of 2500ns for all Xray triggers:

trig_stretch(1)  = 2500ns;
trig_stretch(2)  = 2500ns;
trig_stretch(3)  = 2500ns;

This prevents the multiple triggers to reach the lmu (before_dt scalers)
However, this will also lead to a very small loss of good signal triggers that overlay within 2.5 us.

This has to be taken into account for deadtime determination.
ELOG V3.1.5-fc6679b