Sat May 29 13:43:39 2021, Tino, Rene, Yuri, Laszlo, Markus, Runs, change of settings, 2021
|
from now on, everal beam setting will be different. We could not get the beam back to were it was before. When the scraper was pulled back by 2 cm, the loe/energy beam was visible again via cap.
Schottky. However, no x-rays could be observerved during the target-on phase.
Hence the interpretation:
beam position has change at least a few mm, at most 2 cm.
Will now try to get a new overlap with target and need to adjust the detector position too.
All online - Si spectra are cleared. |
Sat May 29 15:26:40 2021, Sophia, Mario, Runs, Run 0195, 2021
|
No. of stacks: 10 00:52:43
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm
open file : e127b_run0195.lmd
beam: 118Te52+
energy: 6 MeV/u
cooler voltage:
purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)
Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +6.0 mm
run start: 15:25:44 29.05.2021
run stop: 16:27:45 29.05.2021
_______________________________________________________________________________
ON Rates
Si_ON_bdt: ~ 0 Hz
Xray_ON_bdt: ~ 0 Hz
BaF_OR: ---
Si voltage: 100.0 V
Si leakage current: 380 nA
SIS particles before: 2.7e9 (one shoot)
ESR current at injection: 3.0e6 (stacked)
ESR particles after decel.: not measurable
Target ON density: 8.0e13 |
Sat May 29 16:28:50 2021, Sophia, Mario, Runs, Run 0196, 2021
|
No. of stacks: 10 00:52:43
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm
open file : e127b_run0196.lmd
beam: 118Te52+
energy: 6 MeV/u
cooler voltage:
purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)
Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +6.0 mm
run start: 16:27:45 29.05.2021
run stop: 17:27:27 29.05.2021
_______________________________________________________________________________
ON Rates
Si_ON_bdt: ~ 0 Hz
Xray_ON_bdt: ~ 0 Hz
BaF_OR: ---
Si voltage: 100.0 V
Si leakage current: 380 nA
SIS particles before: 2.7e9 (one shoot)
ESR current at injection: 3.0e6 (stacked)
ESR particles after decel.: not measurable
Target ON density: 8.0e13 at 17:00 it increased during LN2 filling to 10E14 |
Sat May 29 17:28:26 2021, Sophia, Mario, Runs, Run 0197, 2021
|
No. of stacks: 10 00:52:43
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm
open file : e127b_run0197.lmd
beam: 118Te52+
energy: 6 MeV/u
cooler voltage:
purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)
Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +6.0 mm
run start: 17:27:32 29.05.2021
run stop: 18:28:10 29.05.2021
_______________________________________________________________________________
ON Rates
Si_ON_bdt: ~ 0 Hz
Xray_ON_bdt: ~ 0 Hz
BaF_OR: ---
Si voltage: 100.0 V
Si leakage current: 380 nA
SIS particles before: 2.7e9 (one shoot)
ESR current at injection: 3.0e6 (stacked)
ESR particles after decel.: not measurable
Target ON density: 10E14 |
Sat May 29 18:29:26 2021, Sophia, Mario, Runs, Run 0198, 2021
|
No. of stacks: 10 00:52:43
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm
open file : e127b_run0198.lmd
beam: 118Te52+
energy: 6 MeV/u
cooler voltage:
purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)
Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +6.0 mm
run start: 18:28:12 29.05.2021
run stop: 19:33:30 29.05.2021
_______________________________________________________________________________
ON Rates
Si_ON_bdt: ~ 0 Hz
Xray_ON_bdt: ~ 0 Hz
BaF_OR: ---
Si voltage: 100.0 V
Si leakage current: 380 nA
SIS particles before: 2.7e9 (one shoot)
ESR current at injection: 3.0e6 (stacked)
ESR particles after decel.: not measurable
Target ON density: 8.0E13 |
Sat May 29 19:34:41 2021, Sophia, Mario, Runs, Run 0199, 2021
|
No. of stacks: 10 00:52:43
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm
open file : e127b_run0199.lmd
beam: 118Te52+
energy: 6 MeV/u
cooler voltage:
purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)
Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +6.0 mm
run start: 19:33:33 29.05.2021
run stop: 20:33:33 29.05.2021
_______________________________________________________________________________
ON Rates
Si_ON_bdt: ~ 0 Hz
Xray_ON_bdt: ~ 0 Hz
BaF_OR: ---
Si voltage: 100.0 V
Si leakage current: 380 nA
SIS particles before: 2.7e9 (one shoot)
ESR current at injection: 3.0e6 (stacked)
ESR particles after decel.: not measurable
Target ON density: 8.0E13 |
Sat May 29 20:35:22 2021, Sophia, Mario, Runs, Run 0200, 2021
|
No. of stacks: 10 00:52:43
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm
open file : e127b_run0200.lmd
beam: 118Te52+
energy: 6 MeV/u
cooler voltage:
purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)
Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +6.0 mm
run start: 20:33:36 29.05.2021
run stop: 20:45:12 29.05.2021
_______________________________________________________________________________
ON Rates
Si_ON_bdt: ~ 0 Hz
Xray_ON_bdt: ~ 0 Hz
BaF_OR: ---
Si voltage: 100.0 V
Si leakage current: 380 nA
SIS particles before: 2.7e9 (one shoot)
ESR current at injection: 3.0e6 (stacked)
ESR particles after decel.: not measurable
Target ON density: 8.0E13 |
Sat May 29 21:37:54 2021, Sophia, Mario, Runs, Run 0201, 2021
|
No. of stacks: 10 00:52:43
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm
open file : e127b_run0201.lmd
beam: 118Te52+
energy: 6 MeV/u
cooler voltage:
purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)
Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +6.0 mm
run start: 21:01:17 29.05.2021
run stop: 22:02:02 29.05.2021
_______________________________________________________________________________
ON Rates
Si_ON_bdt: ~ 0 Hz
Xray_ON_bdt: ~ 0 Hz
BaF_OR: ---
Si voltage: 100.0 V
Si leakage current: 380 nA
SIS particles before: 2.7e9 (one shoot)
ESR current at injection: 3.0e6 (stacked)
ESR particles after decel.: not measurable
Target ON density: 8.0E13 |
Sat May 29 22:03:28 2021, Sophia, Mario, Runs, Run 0202, 2021
|
No. of stacks: 10 00:52:43
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm
open file : e127b_run0202.lmd
beam: 118Te52+
energy: 6 MeV/u
cooler voltage:
purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)
Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +6.0 mm
run start: 22:02:07 29.05.2021
run stop: 22:24:21 29.05.2021
_______________________________________________________________________________
ON Rates
Si_ON_bdt: ~ 0 Hz
Xray_ON_bdt: ~ 0 Hz
BaF_OR: ---
Si voltage: 100.0 V
Si leakage current: 380 nA
SIS particles before: 2.7e9 (one shoot)
ESR current at injection: 3.0e6 (stacked)
ESR particles after decel.: not measurable
Target ON density: 8.0E13 |
Sat May 29 23:18:43 2021, Oliver, Diego, Runs, Run 0203, 2021
|
No. of stacks: 10 00:52:43
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm
open file : e127b_run0203.lmd
beam: 118Te52+
energy: 6 MeV/u
cooler voltage:
purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)
Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +6.0 mm
run start: 23:18:19 29.05.2021
run stop: 00:18:53 30.05.2021
_______________________________________________________________________________
ON Rates
Si_ON_bdt: ~ 0 Hz
Xray_ON_bdt: ~ 0 Hz
BaF_OR: ---
Si voltage: 100.0 V
Si leakage current: 410 nA
SIS particles before: 2.0-2.9e9 (one shoot)
ESR current at injection: 1.8e6 (stacked)
ESR particles after decel.: not measurable
Target ON density: 8.0E13 |
Sun May 30 00:19:33 2021, Oliver, Diego, Runs, Run 0204, 2021
|
No. of stacks: 10 00:52:43
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm
open file : e127b_run0204.lmd
beam: 118Te52+
energy: 6 MeV/u
cooler voltage:
purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)
Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +6.0 mm
run start: 00:18:57 30.05.2021
run stop: 01:21:30 30.05.2021
_______________________________________________________________________________
ON Rates
Si_ON_bdt: ~ 0 Hz
Xray_ON_bdt: ~ 100 Hz
BaF_OR: ---
Si voltage: 100.0 V
Si leakage current: 400 nA
SIS particles before: 2.5e9 (one shoot)
ESR current at injection: 2.5e6 (stacked)
ESR particles after decel.: not measurable
Target ON density: 8.0E13 |
Sun May 30 01:22:03 2021, Oliver, Diego, Runs, Run 0205, 2021
|
No. of stacks: 10 00:52:43
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm
open file : e127b_run0205.lmd
beam: 118Te52+
energy: 6 MeV/u
cooler voltage:
purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)
Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +6.0 mm
run start: 01:21:34 30.05.2021
run stop: 02:25:32 30.05.2021
_______________________________________________________________________________
ON Rates
Si_ON_bdt: ~ 0 Hz
Xray_ON_bdt: ~ 100 Hz
BaF_OR: ---
Si voltage: 100.0 V
Si leakage current: 410 nA
SIS particles before: 2.5e9 (one shoot)
ESR current at injection: 2.75e6 (stacked)
ESR particles after decel.: not measurable
Target ON density: 8.0E13 |
Sun May 30 02:26:07 2021, Oliver, Diego, Runs, Run 0206, 2021
|
No. of stacks: 10 00:52:43
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm
open file : e127b_run0206.lmd
beam: 118Te52+
energy: 6 MeV/u
cooler voltage:
purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)
Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +6.0 mm
run start: 02:25:37 30.05.2021
run stop: 03:30:03 30.05.2021
_______________________________________________________________________________
ON Rates
Si_ON_bdt: ~ 0 Hz
Xray_ON_bdt: ~ 130 Hz
BaF_OR: ---
Si voltage: 100.0 V
Si leakage current: 420 nA
SIS particles before: 2.7e9 (one shoot)
ESR current at injection: 2.75e6 (stacked)
ESR particles after decel.: not measurable
Target ON density: 8.0E13 |
Sun May 30 03:30:37 2021, Oliver, Diego, Runs, Run 0207, 2021
|
No. of stacks: 10 00:52:43
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm
open file : e127b_run0207.lmd
beam: 118Te52+
energy: 6 MeV/u
cooler voltage:
purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)
Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +6.0 mm
run start: 03:30:06 30.05.2021
run stop: 04:27:27 30.05.2021
_______________________________________________________________________________
ON Rates
Si_ON_bdt: ~ 0 Hz
Xray_ON_bdt: ~ 120 Hz
BaF_OR: ---
Si voltage: 100.0 V
Si leakage current: 420 nA
SIS particles before: 2.6e9 (one shoot)
ESR current at injection: 2.9e6 (stacked)
ESR particles after decel.: not measurable
Target ON density: 8.0E13 |
Sun May 30 04:27:58 2021, Oliver, Diego, Runs, Run 0208, 2021
|
No. of stacks: 10 00:52:43
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm
open file : e127b_run0208.lmd
beam: 118Te52+
energy: 6 MeV/u
cooler voltage:
purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)
Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +6.0 mm
run start: 04:27:30 30.05.2021
run stop: 05:32:04 30.05.2021
_______________________________________________________________________________
ON Rates
Si_ON_bdt: ~ 0 Hz
Xray_ON_bdt: ~ 100 Hz
BaF_OR: ---
Si voltage: 100.0 V
Si leakage current: 420 nA
SIS particles before: 2.5e9 (one shoot)
ESR current at injection: 2.5e6 (stacked)
ESR particles after decel.: not measurable
Target ON density: 8.0E13 |
Sun May 30 05:32:42 2021, Oliver, Diego, Runs, Run 0209, 2021
|
No. of stacks: 10 00:52:43
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm
open file : e127b_run0209.lmd
beam: 118Te52+
energy: 6 MeV/u
cooler voltage:
purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)
Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +6.0 mm
run start: 05:32:09 30.05.2021
run stop: 06:32:13 30.05.2021
_______________________________________________________________________________
ON Rates
Si_ON_bdt: ~ 0 Hz
Xray_ON_bdt: ~ 110 Hz
BaF_OR: ---
Si voltage: 100.0 V
Si leakage current: 410 nA
SIS particles before: 2.5e9 (one shoot)
ESR current at injection: 2.5e6 (stacked)
ESR particles after decel.: not measurable
Target ON density: 8.0E13 |
Sun May 30 06:32:43 2021, Oliver, Diego, Jan, Runs, Run 0210, 2021
|
No. of stacks: 10 00:52:43
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm
open file : e127b_run0210.lmd
beam: 118Te52+
energy: 6 MeV/u
cooler voltage:
purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)
Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +6.0 mm
run start: 06:32:16 30.05.2021
run stop: 07:12:04 30.05.2021
_______________________________________________________________________________
ON Rates
Si_ON_bdt: ~ 0 Hz
Xray_ON_bdt: ~ 110 Hz
BaF_OR: ---
Si voltage: 100.0 V
Si leakage current: 400 nA
SIS particles before: 2.7e9 (one shoot)
ESR current at injection: 2.7e6 (stacked)
ESR particles after decel.: not measurable
Target ON density: 8.0E13 |
Sun May 30 07:26:14 2021, Oliver, Diego, Jan, Runs, Run 0211, 2021
|
No. of stacks: 10 00:52:43
length SC10: 2x 5.5s
bump ge01sq2f (sc10): -20 mm
bump target (sc10): -12.8 mm
open file : e127b_run0211.lmd
beam: 118Te52+
energy: 6 MeV/u
cooler voltage:
purpose:
Fragment beam measurement (beating the record of rad. beam pg-measurement)
Detector position (Si): -25 mm
GE01DD1IG position (Scraper): +6.0 mm
run start: 07:25:54 30.05.2021
run stop: 08:04:19 30.05.2021
_______________________________________________________________________________
ON Rates
Si_ON_bdt: ~ 0 Hz
Xray_ON_bdt: ~ 110 Hz
BaF_OR: ---
Si voltage: 100.0 V
Si leakage current: 400 nA
SIS particles before: 2.7e9 (one shoot)
ESR current at injection: 2.7e6 (stacked)
ESR particles after decel.: not measurable
Target ON density: 8.0E13 |
Mon May 31 10:08:42 2021, Jan, Sophia, Runs, E127b runsheet, 2021
|
|
Wed Feb 6 10:16:09 2019, Laszlo Varga, General, Study on scraping,
|
A publication on scraping systems suggested by Siegbert.
In the study they have used cylindrical "high-quality" slits. "High-quality" stands for well polished surface
with surface roughness <50nm. The material of the slit is Tungsten Carbide. |