HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018 E0028
  proton-capture on 118Te, Page 24 of 29  Not logged in ELOG logo
New entries since:Thu Jan 1 01:00:00 1970
Entry  Sat Mar 21 07:30:19 2020, Ragan, RuiJiu, Runs, run entry - run056,  
Open file: runstart056

beam: 118Te52+
energy: 10 MeV/u

purpose: data with TARGET ON

Detector position (Si): -24 mm

run start at 7:30 , first file: run056_0001.lmd
run stop  at xx:xx , last file:  run0xx_0xxx.lmd

_______________________________________________________________________________

time: 7:30 am.

ON Rates
Si_X:      0 Hz
Si_Y:      0 Hz
Xray_35:   0 Hz
Xray_90:   32 Hz
Xray_145:  41 Hz
BaF_OR:    - Hz

Si voltage:          60.02 V  
Si leakage current:  4.04 uA


SIS particles before:           3.2e9
ESR particles at injection:     estimated ~1.2e6
ESR particles after decel.:     
Target ON density:              1.61e13


copy and repeat the above (below the line) every 30 minutes
Entry  Sat Mar 21 07:00:18 2020, Ragan, RuiJiu, Tino, Oliver, Runs, run entry - run055,  
beam: 118Te52+
energy: 10 MeV/u

purpose: data with TARGET ON

Detector position (Si): -24 mm

run start at 06:30 , first file: run055_0001.lmd
run stop  at xx:xx , last file:  run0xx_0xxx.lmd

_______________________________________________________________________________

time: 7:00 am.

ON Rates
Si_X:      0 Hz
Si_Y:      0 Hz
Xray_35:   0 Hz
Xray_90:   31 Hz
Xray_145:  42 Hz
BaF_OR:    - Hz

Si voltage:          60.02 V  
Si leakage current:  4.04 uA


SIS particles before:           3.6e9
ESR particles at injection:     estimated ~10^5 - 10^4
ESR particles after decel.:     
Target ON density:              1.49e13


copy and repeat the above (below the line) every 30 minutes
Entry  Sat Mar 21 06:29:36 2020, Tino, Oliver, Runs, run entry - run055,  
Open file: runstart055

beam: 118Te52+
energy: 10 MeV/u

purpose: data with TARGET ON

Detector position (Si): -24 mm

run start at 06:29 , first file: run055_0001.lmd
run stop  at xx:xx , last file:  run0xx_0xxx.lmd

_______________________________________________________________________________

time: 06:29 am.

ON Rates
Si_X:      0 Hz
Si_Y:      0 Hz
Xray_35:   0 Hz
Xray_90:   36 Hz
Xray_145:  25 Hz
BaF_OR:    - Hz

Si voltage:          60.02 V  
Si leakage current:  4.04 uA


SIS particles before:           2.3e9
ESR particles at injection:     estimated ~10^5 - 10^4
ESR particles after decel.:     
Target ON density:              1.36e13


copy and repeat the above (below the line) every 30 minutes
Entry  Sat Mar 21 06:02:22 2020, Tino, Oliver, General, Detector filling,  
Run stopped for detector fillings
Entry  Sat Mar 21 06:00:19 2020, Tino, Oliver, Runs, run entry - run054,  
Open file: runstart054

beam: 118Te52+
energy: 10 MeV/u

purpose: data with TARGET ON

Detector position (Si): -24 mm

run start at 05:00 , first file: run054_0001.lmd
run stop  at xx:xx , last file:  run0xx_0xxx.lmd

_______________________________________________________________________________

time: 06:00 am.

ON Rates
Si_X:      0 Hz
Si_Y:      0 Hz
Xray_35:   0 Hz
Xray_90:   58 Hz
Xray_145:  0 Hz
BaF_OR:    - Hz

Si voltage:          60.02 V  
Si leakage current:  4.01 uA


SIS particles before:           2.2e9
ESR particles at injection:     estimated ~10^5 - 10^4
ESR particles after decel.:     
Target ON density:              1.47e13


copy and repeat the above (below the line) every 30 minutes
Entry  Sat Mar 21 05:29:17 2020, Tino, Oliver, Runs, run entry - run054,  
Open file: runstart054

beam: 118Te52+
energy: 10 MeV/u

purpose: data with TARGET ON

Detector position (Si): -24 mm

run start at 05:00 , first file: run054_0001.lmd
run stop  at xx:xx , last file:  run0xx_0xxx.lmd

_______________________________________________________________________________

time: 05:29 am.

ON Rates
Si_X:      0 Hz
Si_Y:      0 Hz
Xray_35:   1 Hz
Xray_90:   25 Hz
Xray_145:  15 Hz
BaF_OR:    - Hz

Si voltage:          60.02 V  
Si leakage current:  3.98 uA


SIS particles before:           2.4e9
ESR particles at injection:     estimated ~10^5 - 10^4
ESR particles after decel.:     
Target ON density:              1.31e13


copy and repeat the above (below the line) every 30 minutes
Entry  Sat Mar 21 05:01:09 2020, Tino, Oliver, Runs, run entry - run054,  
Open file: runstart054

beam: 118Te52+
energy: 10 MeV/u

purpose: data with TARGET ON

Detector position (Si): -24 mm

run start at 05:00 , first file: run054_0001.lmd
run stop  at xx:xx , last file:  run0xx_0xxx.lmd

_______________________________________________________________________________

time: 05:00 am.

ON Rates
Si_X:      0 Hz
Si_Y:      0 Hz
Xray_35:   0 Hz
Xray_90:   19 Hz
Xray_145:  23 Hz
BaF_OR:    - Hz

Si voltage:          60.01 V  
Si leakage current:  3.98 uA


SIS particles before:           2.4e9
ESR particles at injection:     estimated ~10^5 - 10^4
ESR particles after decel.:     
Target ON density:              1.34e13


copy and repeat the above (below the line) every 30 minutes
Entry  Sat Mar 21 04:29:59 2020, Tino, Oliver, Runs, run entry - run053,  
Open file: runstart053

beam: 118Te52+
energy: 10 MeV/u

purpose: data with TARGET ON

Detector position (Si): -24 mm

run start at 03:03 , first file: run053_0001.lmd
run stop  at xx:xx , last file:  run0xx_0xxx.lmd

_______________________________________________________________________________

time: 04:30 am.

ON Rates
Si_X:      0 Hz
Si_Y:      0 Hz
Xray_35:   0 Hz
Xray_90:   27 Hz
Xray_145:  17 Hz
BaF_OR:    - Hz

Si voltage:          60.01 V  
Si leakage current:  4.03 uA


SIS particles before:           2.4e9
ESR particles at injection:     estimated ~10^5 - 10^4
ESR particles after decel.:     
Target ON density:              1.40e13


copy and repeat the above (below the line) every 30 minutes
Entry  Sat Mar 21 03:58:56 2020, Tino, Oliver, Runs, run entry - run053,  
Open file: runstart053

beam: 118Te52+
energy: 10 MeV/u

purpose: data with TARGET ON

Detector position (Si): -24 mm

run start at 03:03 , first file: run053_0001.lmd
run stop  at xx:xx , last file:  run0xx_0xxx.lmd

_______________________________________________________________________________

time: 03:58 am.

ON Rates
Si_X:      0 Hz
Si_Y:      0 Hz
Xray_35:   1 Hz
Xray_90:   21 Hz
Xray_145:  21 Hz
BaF_OR:    - Hz

Si voltage:          60.02 V  
Si leakage current:  3.96 uA


SIS particles before:           2.5e9
ESR particles at injection:     estimated ~10^5 - 10^4
ESR particles after decel.:     
Target ON density:              1.16e13


copy and repeat the above (below the line) every 30 minutes
Entry  Sat Mar 21 03:30:21 2020, Tino, Oliver, Runs, run entry - run052,  
Open file: runstart052

beam: 118Te52+
energy: 10 MeV/u

purpose: data with TARGET ON

Detector position (Si): -24 mm

run start at 03:03 , first file: run052_0001.lmd
run stop  at xx:xx , last file:  run0xx_0xxx.lmd

_______________________________________________________________________________

time: 03:30 am.

ON Rates
Si_X:      0 Hz
Si_Y:      0 Hz
Xray_35:   0 Hz
Xray_90:   22 Hz
Xray_145:  14 Hz
BaF_OR:    - Hz

Si voltage:          60.02 V  
Si leakage current:  3.98 uA


SIS particles before:           2.5e9
ESR particles at injection:     estimated ~10^5 - 10^4
ESR particles after decel.:     
Target ON density:              1.23e13


copy and repeat the above (below the line) every 30 minutes
Entry  Sat Mar 21 03:03:53 2020, Tino, Oliver, Runs, run entry - run052,  
Open file: runstart052

beam: 118Te52+
energy: 10 MeV/u

purpose: data with TARGET ON

Detector position (Si): -24 mm

run start at 03:03 , first file: run052_0001.lmd
run stop  at xx:xx , last file:  run0xx_0xxx.lmd

_______________________________________________________________________________

time: 03:03 am.

ON Rates
Si_X:      0 Hz
Si_Y:      0 Hz
Xray_35:   1 Hz
Xray_90:   19 Hz
Xray_145:  26 Hz
BaF_OR:    - Hz

Si voltage:          60.02 V  
Si leakage current:  3.99 uA


SIS particles before:           2.3e9
ESR particles at injection:     estimated ~10^5 - 10^4
ESR particles after decel.:     
Target ON density:              1.22e13


copy and repeat the above (below the line) every 30 minutes
Entry  Sat Mar 21 03:01:22 2020, Tino, Oliver, Runs, run entry - run051,  
Open file: runstart051

beam: 118Te52+
energy: 10 MeV/u

purpose: data with TARGET ON

Detector position (Si): -24 mm

run start at 01:59 , first file: run051_0001.lmd
run stop  at xx:xx , last file:  run0xx_0xxx.lmd

_______________________________________________________________________________

time: 03:01 am.

ON Rates
Si_X:      0 Hz
Si_Y:      0 Hz
Xray_35:   3 Hz
Xray_90:   23 Hz
Xray_145:  18 Hz
BaF_OR:    - Hz

Si voltage:          60.02 V  
Si leakage current:  3.99 uA


SIS particles before:           2.3e9
ESR particles at injection:     estimated ~10^5 - 10^4
ESR particles after decel.:     
Target ON density:              1.24e13


copy and repeat the above (below the line) every 30 minutes
Entry  Sat Mar 21 02:36:44 2020, Laszlo, General, detector and scraper position,  IMG_20200321_025644.jpg
For the 124Xe primary beam measurement:
-with the DSSSD we scraped the beam at position -40
-position of the detector is set to -25. (1.5cm away from the beam)
-with the scraper (at Eggelhof 1 position) we scraped the beam at ~(-15)-(-13)
-position of the scraper is set to +20 (3.5cm away from beam axis (sollbahn))

For the 118Te fragment measurement:
-with the DSSSD we scraped the beam at position -39
-position of the detector is set to -24 (1.5cm away from the beam)
-with the scraper (at Eggelhof 1 position) we scraped the beam at ~ -14
-position of the scraper is set to +19 (3.5cm away from beam axis (sollbahn))
Entry  Sat Mar 21 02:31:48 2020, Laszlo, Tino, Oliver, Runs, run entry - run051,  
Open file: runstart051

beam: 118Te52+
energy: 10 MeV/u

purpose: data with TARGET ON

Detector position (Si): -24 mm

run start at 01:59 , first file: run051_0001.lmd
run stop  at xx:xx , last file:  run0xx_0xxx.lmd

_______________________________________________________________________________

time: 02:31 am.

ON Rates
Si_X:      0-1 Hz
Si_Y:       0-2 Hz
Xray_35:   1-3 Hz
Xray_90:   39 Hz
Xray_145:  21 Hz
BaF_OR:    - Hz

Si voltage:          60.01 V  
Si leakage current:  3.98 uA


SIS particles before:           2.3e9
ESR particles at injection:     estimated ~10^5 - 10^4
ESR particles after decel.:     
Target ON density:              1.46e13


copy and repeat the above (below the line) every 30 minutes
Entry  Sat Mar 21 02:15:19 2020, Laszlo, Tino, Oliver, Runs, run entry - run051,  
Open file: runstart051

beam: 118Te52+
energy: 10 MeV/u

purpose: data with TARGET ON

Detector position (Si): -24 mm

run start at 01:59 , first file: run051_0001.lmd
run stop  at xx:xx , last file:  run0xx_0xxx.lmd

_______________________________________________________________________________

time: 01:59 am.

ON Rates
Si_X:      0-1 Hz
Si_Y:       0-1 Hz
Xray_35:   2 Hz
Xray_90:   33 Hz
Xray_145:  33 Hz
BaF_OR:    - Hz

Si voltage:          60.02 V  
Si leakage current:  3.98 uA


SIS particles before:           2.3e9
ESR particles at injection:     estimated ~10^5 - 10^4
ESR particles after decel.:     
Target ON density:              1.46e13


copy and repeat the above (below the line) every 30 minutes
Entry  Sat Mar 21 02:00:08 2020, Laszlo, Tino, Oliver, Runs, run entry - run051,  
Beam is back at 01:59 am.
Open file: runstart051

beam: 118Te52+
energy: 10 MeV/u

purpose: data with TARGET ON

Detector position (Si): -24 mm

run start at 01:59 , first file: run051_0001.lmd
run stop  at xx:xx , last file:  run0xx_0xxx.lmd

_______________________________________________________________________________

time: 10:30

ON Rates
Si_X:      0-1 Hz
Si_Y:       0-1 Hz
Xray_35:   2 Hz
Xray_90:   33 Hz
Xray_145:  33 Hz
BaF_OR:    - Hz

Si voltage:          60.02 V  
Si leakage current:  3.98 uA


SIS particles before:           2.3e9
ESR particles at injection:     estimated ~10^5 - 10^4
ESR particles after decel.:     
Target ON density:              1.46e13


copy and repeat the above (below the line) every 30 minutes
Entry  Sat Mar 21 01:44:57 2020, Laszlo, General, ,  
from the SIS side they could optimize the intensity. we gained a factor ~3. Now the SIS intensity is at 
Entry  Sat Mar 21 01:25:07 2020, Laszlo, General, Vacuum while moving in the detector for fragment measurement,  DSSSD_moving_vacuum_change1.JPG
Stayed below E-10
Start:2.42E-11
Top after moving: 4.144E-11
Entry  Fri Mar 20 23:43:26 2020, Laszlo, General, Vacuum while moving in the detector at yesterday evening,  DSSSD_moving_vacuum_change.JPGIMG_20200320_021131.jpg
Vacuum before DSSSD movement: 1.5844E-11
Top value after movement: 3.64378E-11

This oscillating effect is due to magnets?
Entry  Fri Mar 20 22:54:30 2020, Laszlo, General, First view of the beam,  IMG_20200320_222748.jpgIMG_20200320_223424.jpg
Intensity from Sis 1.7E9
The intensity is too.low at ESR to be visible. However we could.catch it somehow with Shottky (picture below)
ELOG V3.1.5-fc6679b