ID |
Date |
Author |
Category |
Subject |
Year |
Text |
|
45
|
Sun Mar 15 14:41:34 2020 |
Jan | Calibration | run034 - Xray145 calib B133 high rate -SOURCE MOVED! | | measurement with old tennelec specAmps, do
not use for efficiency!
|
|
46
|
Sun Mar 15 15:08:44 2020 |
Jan | Calibration | run035 - Xray145 calib B133 high rate | | measurement with old tennelec specAmps, do
not use for efficiency!
|
|
59
|
Mon Mar 16 15:15:22 2020 |
Jan | Calibration | Gamma- and X-ray intensities for efficiency calib | | Intensities for special gamma or x-ray lines
needed for efficiency calibration can be
found here:
|
|
60
|
Tue Mar 17 08:15:49 2020 |
Jan | General | current grids & screen in TE | | measured beam diagnosis with 124Xe48+ at 326
MeV/u |
|
61
|
Tue Mar 17 11:44:58 2020 |
Jan | Runs | run entry - template | | beam: 118Te52+
energy: 7.4 MeV/u
|
|
64
|
Thu Mar 19 15:09:29 2020 |
Jan | General | starting off with DSSSD | | here is my short guide how to start with the
measurement as soon as we have a beam available.
|
|
71
|
Fri Mar 20 04:29:45 2020 |
Jan | DAQ | DAQ settings for the first night | | |
|
85
|
Fri Mar 20 16:54:53 2020 |
Jan | DAQ | BaF OFF downscaled red=32 | | I introduced a downscaling by a factor of
2^5=32 for the BaF2 during target OFF.
The reason is, that we get a lot reasonable |
|
86
|
Fri Mar 20 16:58:02 2020 |
Jan | DAQ | Xray2 & Xray3 cabling | | We found that Xray2 and Xray3 the cabling
is not consistent:
|
|
87
|
Fri Mar 20 19:37:43 2020 |
Jan | Calibration | Xray energies - rough 1. calibration | | -- E calibration seems to have changed! ---
This calibration is based on the source data |
|
88
|
Fri Mar 20 22:26:42 2020 |
Jan | Detectors | BaF2 deactivated | | BaF2 detectors are deactivated nor for the
following runs. |
|
147
|
Sat Mar 21 21:36:06 2020 |
Jan | Runs | target incident | | in run 66 or 67 we had a single ESR cycle
for which the target didn't switch off for
new injection and deceleration etc.
|
|
152
|
Sat Mar 21 23:12:38 2020 |
Jan | Detectors | removed BaF detectors | | After run 67 and before run 68, the BaF detectors
have been removed in order to reduce the
Compton background in the x-ray detectors. |
|
153
|
Sat Mar 21 23:19:14 2020 |
Jan | Analysis | 118Te52+ on H2-target - Xray spectrum 90 degree | | The attached spectrum is for the 90 degree
X-ray detector taken for about 3 hours with
118Te52+.
|
|
244
|
Mon Mar 23 13:26:59 2020 |
Jan | Analysis | runs_90to99 SCRAPER RESULTS | | With scraper, we have roughly 150-200 (p,g)
counts now.
|
|
251
|
Mon Mar 23 14:32:03 2020 |
Jan | Analysis | runs_100to104 NO-SCRAPER RESULTS | | Without scraper, we have very few (p,g) counts,
hard to get numbers quickly, maybe ~50.
|
|
266
|
Tue Mar 24 10:59:18 2020 |
Jan | Calibration | Sources - Specifications | | We used the following sources:
[B]Am241[/B] (OM666) [GSI - Uwe]
|
|
Draft
|
Tue Apr 21 22:25:00 2020 |
Jan | Analysis | | | |
|
284
|
Fri Oct 23 12:58:12 2020 |
Jan | Detectors | Test of 2nd DSSSD (gen2) | 2021 | This is the documentation of the source tests
with the 2nd micron DSSSD of 2nd generation
(label 3288-17, thickness 529um)
|
|
285
|
Fri Nov 20 15:53:37 2020 |
Jan | Analysis | Test of 2nd DSSSD (gen2) | 2021 | Here is the analysis of the test runs with
the 2nd DSSD (gen2) using the quick_si_plots.c
script attached.
|
|