  45   Sun Mar 15 14:41:34 2020 JanCalibrationrun034 - Xray145 calib B133 high rate -SOURCE MOVED! measurement with old tennelec specAmps, do
not use for efficiency!
  
  46   Sun Mar 15 15:08:44 2020 JanCalibrationrun035 - Xray145 calib B133 high rate measurement with old tennelec specAmps, do
not use for efficiency!
  
  59   Mon Mar 16 15:15:22 2020 JanCalibrationGamma- and X-ray intensities for efficiency calib Intensities for special gamma or x-ray lines
needed for efficiency calibration can be
found here:
  
  60   Tue Mar 17 08:15:49 2020 JanGeneralcurrent grids & screen in TE  measured beam diagnosis with 124Xe48+ at 326
MeV/u
 2020-03-16_18-51-50-170.pngIMG_0694.jpg 
  61   Tue Mar 17 11:44:58 2020 JanRunsrun entry - template  beam: 118Te52+
energy: 7.4 MeV/u
  
  64   Thu Mar 19 15:09:29 2020 JanGeneralstarting off with DSSSD here is my short guide how to start with the
measurement as soon as we have a beam available.
  
  71   Fri Mar 20 04:29:45 2020 JanDAQDAQ settings for the first night    e127.trlomain.cfgr3bfuser.cfgsetting.1584673412 
  85   Fri Mar 20 16:54:53 2020 JanDAQBaF OFF downscaled red=32 I introduced a downscaling by a factor of
2^5=32 for the BaF2 during target OFF.
The reason is, that we get a lot reasonable
  
  86   Fri Mar 20 16:58:02 2020 JanDAQXray2 & Xray3 cabling We found that Xray2 and Xray3 the cabling
is not consistent:
  
  87   Fri Mar 20 19:37:43 2020 JanCalibrationXray energies - rough 1. calibration  -- E calibration seems to have changed! ---

This calibration is based on the source data
  
  88   Fri Mar 20 22:26:42 2020 JanDetectorsBaF2 deactivated BaF2 detectors are deactivated nor for the
following runs.
  
  147   Sat Mar 21 21:36:06 2020 JanRunstarget incident in run 66 or 67 we had a single ESR cycle
for which the target didn't switch off for
new injection and deceleration etc.
 run67_target_incident.png 
  152   Sat Mar 21 23:12:38 2020 JanDetectorsremoved BaF detectors After run 67 and before run 68, the BaF detectors
have been removed in order to reduce the
Compton background in the x-ray detectors.
  
  153   Sat Mar 21 23:19:14 2020 JanAnalysis118Te52+ on H2-target - Xray spectrum 90 degree The attached spectrum is for the 90 degree
X-ray detector taken for about 3 hours with
118Te52+.
 90_deg_118Te_runs64to66.png 
  244   Mon Mar 23 13:26:59 2020 JanAnalysisruns_90to99 SCRAPER RESULTS With scraper, we have roughly 150-200 (p,g)
counts now.
 124Xe_Si_map.png124Xe_x-ray_35.png124Xe_x-ray_90.png124Xe_x-ray_145.png124Xe_Si_projection.png 
  251   Mon Mar 23 14:32:03 2020 JanAnalysisruns_100to104 NO-SCRAPER RESULTS Without scraper, we have very few (p,g) counts,
hard to get numbers quickly, maybe ~50.
 124Xe_Si_map.png124Xe_Si_projection.png124Xe_x-ray_35.png124Xe_x-ray_90.png124Xe_x-ray_145.png 
  266   Tue Mar 24 10:59:18 2020 JanCalibrationSources - Specifications We used the following sources:

[B]Am241[/B] (OM666) [GSI - Uwe]
 Pb-210-1.pdf241Am_Uwe.pdf133Ba_low.pdf133Ba_hi.pdf 
  Draft   Tue Apr 21 22:25:00 2020 JanAnalysis      
  284   Fri Oct 23 12:58:12 2020 JanDetectorsTest of 2nd DSSSD (gen2)2021This is the documentation of the source tests
with the 2nd micron DSSSD of 2nd generation
(label 3288-17, thickness 529um)
 Si2_run001.jpgSi2_run002.jpg 
  285   Fri Nov 20 15:53:37 2020 JanAnalysisTest of 2nd DSSSD (gen2)2021Here is the analysis of the test runs with
the 2nd DSSD (gen2) using the quick_si_plots.c
script attached.
 quick_si_plots.ce127.hSi2_run001.ana.rootSi2_run002.ana.root 