Entry  Fri Apr 30 09:57:57 2021, Jan Glorius, Calibration, run0005 - Xray90 calib Pb210 d=167.5mm , 2021 
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
Source: 210Pb
Distance: 167.5mm
Start time: 9:58:00 - 30.04.2021
Stop time:  10:01:09 - 30.04.2021

file name: e127b_run0005.lmd
avrg. rate: 50Hz
dead-time:  1%

Still rate peaks, probably the ESR kicker or some other external influence. 
So we have double triggers occasionally. This might slightly affect deadtime determination.
Entry  Fri Apr 30 11:04:31 2021, Jan Glorius, Calibration, run0006 - Xray90 calib Pb210 d=167.5mm , 2021 
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
Source: 210Pb
Distance: 167.5mm
Start time: 11:04:03 - 30.04.2021
Stop time:  11:53:59 - 30.04.2021

file name: e127b_run0006.lmd
avrg. rate: 50Hz
dead-time:  1%

Now with blocking gate to get rid of multiple triggers.
Dead time determination using: adt/bdt

Threshold at ~20keV

Last seconds of run may not be usable: DAQ node2 has been deactivated shortly by mistake.
Entry  Fri Apr 30 13:07:23 2021, Jan Glorius, Calibration, run0008 - Xray90 calib Pb210 d=167.5mm , 2021 
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
Source: 210Pb
Distance: 167.5mm
Start time: 13:08:35 - 30.04.2021
Stop time:  13:49:56 - 30.04.2021

file name: e127b_run0008.lmd
avrg. rate: 60Hz
dead-time:  1%

Now with blocking gate to get rid of multiple triggers.
Dead time determination using: adt/bdt

Threshold at ~10keV

We have some external noise on the signal. 
Entry  Fri Apr 30 13:10:44 2021, Jan Glorius, Calibration, run0007 - crap, 2021 
crap run
Entry  Fri Apr 30 13:56:59 2021, Jan Glorius, Calibration, run0009 - Xray90 calib Am241 d=167.5mm , 2021 
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
Source: 241Am_high
Distance: 167.5mm
Start time: 13:56:33 - 30.04.2021
Stop time:  14:28:53 - 30.04.2021

file name: e127b_run0009.lmd
avrg. rate: 600Hz
dead-time:  4%

Now with blocking gate to get rid of multiple triggers.
Dead time determination using: adt/bdt

Threshold at ~10keV

We have some external noise on the signal. 
Entry  Fri Apr 30 14:35:11 2021, Jan Glorius, Calibration, run0010 - Xray90 calib Ba133 d=167.5mm , 2021 
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
Source: 133Ba_high
Distance: 167.5mm
Start time: 14:34:51 - 30.04.2021
Stop time:  14:45:42 - 30.04.2021

file name: e127b_run0010.lmd
avrg. rate: 2600 Hz
dead-time:  16%

Now with blocking gate to get rid of multiple triggers.
Dead time determination using: adt/bdt

Threshold at ~10keV

We have some external noise on the signal. 
Entry  Fri Apr 30 14:48:01 2021, Jan Glorius, Calibration, run0011 - Xray90 calib Ba133_low d=167.5mm , 2021 
Efficiency calibration in the lab

Detector: GEM1800 - 90 deg
Source: 133Ba_low
Distance: 167.5mm
Start time: 14:47:40 - 30.04.2021
Stop time:  15:50:12 - 30.04.2021

file name: e127b_run0011.lmd
avrg. rate: 310 Hz
dead-time:  3%

Now with blocking gate to get rid of multiple triggers.
Dead time determination using: adt/bdt

Threshold at ~10keV

We have some external noise on the signal. 
Entry  Sat May 1 22:34:16 2021, Jan Glorius, Calibration, efficiency xray 90deg, 2021 90deg_eff.png90deg_eff.dat
the attached efficiency curve for the 90 deg. xray detector is made with the runs
e127b_run0008.lmd
e127b_run0009.lmd
e127b_run0010.lmd

All of them were taken in the lab (not in ESR).

This is a very preliminary analysis!
E.g. deadtime has been estimated and was not cleanly determined.
Entry  Mon May 3 14:55:44 2021, Jan Glorius, Calibration, Xray energy calib. table, 2021 
Here is the table of energy calibration lines from Am241 for all Xray detectors for several measurements.
Also the linear fit parameters [ E(ch) = A*ch + B ] are given at the bottom of the table.

E [keV] xray90 (GEM1800) xray35 (GLP) xray145 (GEM3000)
lab (28.04.21) lab (04.05.21) lab (11.05.21)
59.54 2402.45 ch 2399.07 ch 2423.24 ch
26.34 1107.38 ch 1116.65 ch 1127.94 ch
13.95 624.33 ch 638.38 ch 645.39 ch
slope A 0.0256387 keV/ch 0.0258923 keV/ch 0.0256408 keV/ch
offset B -2.05477 keV -2.57638 keV -2.59114 keV
Entry  Tue May 4 14:31:51 2021, Jan Glorius, Calibration, run0012 - Xray35 calib Am241_hi d=334mm, 2021 
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 241Am_hi
Distance: 334mm
Start time: 14:31:24 - 04.05.2021
Stop time:  14:42:13 - 04.05.2021

file name: e127b_run0012.lmd
avrg. rate: 25 Hz
dead-time:  0%


Threshold < 10keV

The scale serial communication is off.

Stopped due to noise rate...
Entry  Tue May 4 14:46:15 2021, Jan Glorius, Calibration, run0013 - Xray35 calib Am241_hi d=334mm, 2021 
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 241Am_hi
Distance: 334mm
Start time: 14:45:51 - 04.05.2021
Stop time:  15:47:48 - 04.05.2021

file name: e127b_run0013.lmd
avrg. rate: 25 Hz
dead-time:  0%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
Entry  Tue May 4 16:04:42 2021, Jan Glorius, Calibration, run0014 - Xray35 calib Am241_hi d=334mm, 2021 
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 241Am_hi
Distance: 334mm
Start time: 16:04:21 - 04.05.2021
Stop time:  20:41:45 - 04.05.2021

file name: e127b_run0014.lmd
avrg. rate: 25 Hz
dead-time:  0%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
Entry  Tue May 4 21:18:15 2021, Jan Glorius, Calibration, run0015 - Xray35 calib Am241_hi d=334mm, 2021 
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 241Am_hi
Distance: 334mm
Start time: 21:17:44 - 04.05.2021
Stop time:   8:42:09 - 05.05.2021

file name: e127b_run0015.lmd
avrg. rate: 25 Hz
dead-time:  0%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
Entry  Wed May 5 10:02:27 2021, Jan Glorius, Calibration, run0016 - Xray35 calib Ba133_hi d=334mm, 2021 
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 133Ba_hi
Distance: 334mm
Start time: 10:02:07 - 05.05.2021
Stop time:  14:13:23 - 05.05.2021

file name: e127b_run0016.lmd
avrg. rate: 90 Hz
dead-time:  1%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
Entry  Wed May 5 14:24:57 2021, Jan Glorius, Calibration, run0018 - Xray35 calib Pb210 d=334mm, 2021 
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 210Pb
Distance: 334mm
Start time: 14:28:26 - 05.05.2021
Stop time:  21:05:00 - 05.05.2021

file name: e127b_run0018.lmd
avrg. rate: 5 Hz
dead-time:  0%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.

There are many rate spikes in the first hour of measurement.
Entry  Wed May 5 14:29:02 2021, Jan Glorius, Calibration, run0017 crap run, 2021 
 
Entry  Wed May 5 21:22:27 2021, Jan Glorius, Calibration, run0019 - Xray35 calib Pb210 d=334mm, 2021 
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 210Pb
Distance: 334mm
Start time: 21:22:06 - 05.05.2021
Stop time:   8:55:21 - 06.05.2021

file name: e127b_run0019.lmd
avrg. rate: 5 Hz
dead-time:  0%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.

Many rates spikes, check carefully!
Entry  Thu May 6 09:40:37 2021, Jan Glorius, Calibration, run0020 - Xray35 calib Ba133_hi d=334mm, 2021 
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 133Ba_hi
Distance: 334mm
Start time: 9:42:24 - 06.05.2021
Stop time: 14:03:44 - 06.05.2021

file name: e127b_run0020.lmd
avrg. rate: 90 Hz
dead-time:  1%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.
Entry  Thu May 6 13:39:02 2021, Jan Glorius, Calibration, efficiency xray 35deg, 2021 35deg_eff.png35deg_eff.dat
the attached efficiency curve for the 35 deg. xray detector is made with the runs
e127b_run0015.lmd
e127b_run0016.lmd
e127b_run0017.lmd + e127b_run0019.lmd

All of them were taken in the lab (not in ESR).

This is a very preliminary analysis!
E.g. deadtime has been estimated and was not cleanly determined.
Entry  Thu May 6 14:30:35 2021, Jan Glorius, Calibration, run0021 - Xray35 calib Pb210 d=334mm, 2021 
Efficiency calibration in the lab

Detector: GLP2000 - 35 deg
Source: 210Pb
Distance: 334mm
Start time: 14:21:09 - 06.05.2021
Stop time:  21:06:10 - 06.05.2021

file name: e127b_run0021.lmd
avrg. rate: 5 Hz
dead-time:  0%

Blocking gate 2.5 us in vulom.

Threshold < 10keV

The scale serial communication is off.