249
|
Mon Mar 23 14:12:54 2020 |
Ragan | Runs | E127 experiment stops | | close file : run104 (14:12:44)
E127 experiment finishes on a successful
note )) Congratulations to the entire team |
|
250
|
Mon Mar 23 14:23:01 2020 |
Pierre-Michel | DAQ | Backup recoil files | | Recoil detector files backup to lxg1260:/data.local2/hillenbrand/Laszlo |
|
251
|
Mon Mar 23 14:32:03 2020 |
Jan | Analysis | runs_100to104 NO-SCRAPER RESULTS | | Without scraper, we have very few (p,g) counts,
hard to get numbers quickly, maybe ~50.
|
|
253
|
Mon Mar 23 15:48:15 2020 |
Laszlo | Calibration | run105 - Xray90 calib Am241, distance 167.5mm, wrong? | | Detector: 90
Source: Am241
Distance: 167.5mm
|
|
254
|
Mon Mar 23 16:17:58 2020 |
Laszlo | Calibration | run106 - garbage! | | run106 was clicked by accident, please delete! |
|
255
|
Mon Mar 23 16:21:33 2020 |
Laszlo | Calibration | run107 - Xray90 calib Am241, distance 167.5mm | | Detector: 90
Source: Am241
Distance: 167.5mm
|
|
256
|
Mon Mar 23 17:32:42 2020 |
Laszlo | Calibration | run108 - Xray90 calib Ba133, distance 167.5mm - high rate - deadtime too high! | | Detector: 90
Source: Ba133 - strong source
Distance: 167.5mm
|
|
257
|
Mon Mar 23 17:40:33 2020 |
Laszlo | Calibration | run109 - Xray90 calib Ba133, distance 167.5mm -low rate | | Detector: 90
Source: Ba133 - weak source
Distance: 167.5mm
|
|
258
|
Mon Mar 23 18:15:55 2020 |
Laszlo | General | noisy 90degree Xray det | | Uwe found out, that there was some common
ground effect between 90 and 145 degree detectors.
after moving the grounds, we could record |
|
259
|
Mon Mar 23 18:17:27 2020 |
Laszlo | Calibration | run110 - Xray90 calib Ba133, distance 167.5mm -low rate - LESS NOISE | | Detector: 90
Source: Ba133 - weak source
Distance: 167.5mm
|
|
260
|
Mon Mar 23 19:32:53 2020 |
Laszlo | Calibration | run112 - Xray90 calib Am241, distance 200mm - LESS NOISE | | Detector: 90
Source: 241Am
Distance: 200mm
|
|
261
|
Mon Mar 23 20:52:37 2020 |
Laszlo | Calibration | run113 - Xray145 calib Am241, distance 305mm - LESS NOISE | | Detector: 145
Source: 241Am
Distance: 305mm
|
|
262
|
Mon Mar 23 21:47:40 2020 |
Laszlo | Calibration | run114 - Xray145 calib Ba133, distance 305mm - LESS NOISE | | Detector: 145
Source: 133Ba - strong source
Distance: 305mm
|
|
263
|
Mon Mar 23 22:25:41 2020 |
Laszlo | Calibration | run115 - Xray35 Ba133 d=334mm, Xray90 Lead source, d=167.5mm | | Simultaneous measurement of Xray35 and Xray90.
The dead-time is <3%
|
|
264
|
Tue Mar 24 00:11:35 2020 |
Laszlo | Calibration | run116 - Xray35 calib AM241, distance 334mm | | Detector: 35
Source: 241Am
Distance: 334mm
|
|
265
|
Tue Mar 24 10:20:26 2020 |
Laszlo | Calibration | Pictures of the xray calibration | | |
|
266
|
Tue Mar 24 10:59:18 2020 |
Jan | Calibration | Sources - Specifications | | We used the following sources:
[B]Am241[/B] (OM666) [GSI - Uwe]
|
|
267
|
Wed Mar 25 14:26:22 2020 |
Laszlo | Calibration | run111 - Xray90 calib Am241, distance 167.5mm - LESS NOISE | | Detector: 90
Source: 241Am
Distance: 167.5mm
|
|
268
|
Wed Mar 25 15:06:09 2020 |
Laszlo | Analysis | quick summary of the run numbers | | |
|
269
|
Thu Mar 26 12:05:22 2020 |
Laszlo | Calibration | measurement of collimators | | The measurements were taken by using a caliper
ruler, with precision of +/-0.05mm |
|