HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018 E0028 E0038
  Nuclear hyperfine mixing in 229Th89+, Page 2 of 12  Not logged in ELOG logo
Entry  Tue Apr 12 13:24:40 2022, Alexandre, Sebastian, Detectors, Sideband in Schottky IMG_20220412_125530_resized_20220412_125736155.jpgIMG_20220412_125643_resized_20220412_125736537.jpg
Sometimes we see a broad sideband at lesser schottky frequencies to the high intensity thorium beam.

Pic 1 shows the relative broad sideband on the left.
    Reply  Tue Apr 12 14:11:52 2022, Ragan, Detectors, Sideband in Schottky 
> Sometimes we see a broad sideband at lesser Schottky frequencies to the high-intensity thorium beam.

> Pic 1 shows the relative broad sideband on the left.
Entry  Tue Apr 12 12:41:20 2022, Alexandre, Sebastian, Accelerator, no beam in SIS18 
 
Entry  Tue Apr 12 04:43:29 2022, Alexandre, Kristian, Sebastian, Runs, 322.01 - 321.60 nm  DriftingVoltage14-40.PNG
We scan the wavelength scan range is  322.01 - 321.60 nm 

before the first scan of the new wavelength range we changed the high voltage divider ratio in Labview from 248514 to 248517, which is the correct value
Entry  Mon Apr 11 22:33:16 2022, Carsten, General, Backup of data to the external RAID disks 
 
Entry  Mon Apr 11 22:14:51 2022, Konstantin, Sarper, Laser, Changing of the wavelength range 
It is planned to change the scan range of the laser wavelength from 322.41 - 322.0 nm to 322.01 - 321.6 nm @ 4 a.m. 
Entry  Mon Apr 11 18:57:18 2022, Rodolfo, Laser, New flashlamps Blitzlampe_26Mio_Schüsse_11-04-2022.jpg
We (Sarper, Konstantin and me) have changed the old flashlamps (with ~ 26 Mio. shots) by new "fresh" ones. We have also exchange the dye. Here are some
parameters of the laser we measured with the new flashlamps and dye*:
Entry  Mon Apr 11 18:46:31 2022, Rodolfo, Sarper and Konstantin, Laser, Alignment-check MRC_Konfiguration_MitMainAmp_04-11-2022.png
We cross-checked the laser-ion alignment. The horizontal position was optimized. However, 
since the laser beam is much larger than the ion beam, this was not an issue for measurements, which
have been performed during the last shifts.
Entry  Mon Apr 11 14:45:11 2022, Paul, Konstantin, Sarper, David, Runs, Afternoon shift runs 11.04 
We continue scan the new wavelength scan range is  322.41 - 322.00 nm 

Entry  Mon Apr 11 12:50:22 2022, Kristian, Accelerator, Th lifetime ThLifetime.JPG
We used the counter to measure the lifetime. During the 30 min scan time dropped only slowly. 

During a longer break, e.g., when the laser flash lamps are changed, one could take data over a longer time.
Entry  Mon Apr 11 11:05:35 2022, Sebastian, Kristian, Laser, Check laser power 
we checked the UV laser power at 10:15 am. It was only 8.5 - 9 mJ. Pumping power was 460 mJ, which is the maximum output of the current system. We should
change the flashlamps soon
Entry  Mon Apr 11 10:31:07 2022, Kristian, Sebastian, Alexandre, Danyal, Paul, General, new laser wavelength scan range 322.41 - 322.00 nm 
new wavelength scan range is  322.41 - 322.00 nm 

at 10:31 
Entry  Mon Apr 11 10:17:01 2022, Danyal, Accelerator, Interlock - reset  
10:17 
Wie heutemorgen in der Sitzung (morning briefing) vereinbart, 
führt Herr Florenkowski einen Interlock und danach einen Reset von E02MK durch. 
Entry  Mon Apr 11 07:54:37 2022, Danyal, Alexandre, Kristian, Sebastian, Accelerator, Jump in E-cooler voltage 
We saw a jump in the E-cooler voltage at about 7:50. 
jump from 209619 to 209610 V 
Entry  Mon Apr 11 04:26:53 2022, Patrick, DAQ, Next wavelength range 
Measurements with the next wavelength scan range should be started roughly around 10:00 am.
The next wavelength scan range is planned to be 322.41 - 322.00 nm, again with a step size of 0.002 nm, 5s dwell time and 2 loops.
This range includes roughly a third of the first scan range.
Entry  Sun Apr 10 22:47:05 2022, Patrick, Volker, Prannay, Danyal, Alexandre, Sebastian, Kristian, Runs, Runs night shift 10.04 + morning shift 11.04 
These are the measurement runs of the night shift (10/11.04.).

Wavelength range is still 322.81 - 322.40 nm.
Entry  Sun Apr 10 15:29:53 2022, Rodolfo, Laser, Status - Laser system 
We have checked the laser system.

- The pump laser has ~ 23Mio shots. The total output energy has dropped to about 475 mJ. Maybe it is time to install new fresh flashlamps. In case the
Entry  Sun Apr 10 15:29:37 2022, Carsten,Sarper, Accelerator, Ecooler failed 
Ecooler failed @15:23 Carsten is fixing it
Fixed @16:00
Entry  Sun Apr 10 15:16:19 2022, Carsten, Rodolfo, Sarper, Runs, Runs afternoon shifts 10.04 
These are the measurement runs of the afternoon shift (10.04.).

Wavelength range is changed from 323.2 - 322.8 nm to 322.81 - 322.40 nm.
Entry  Sun Apr 10 15:08:42 2022, Carsten, DAQ, Backup of Data from 1rst Scanrange to extrenal RAID1 HDDs 
I just performed a backup of all data (+log-Files) that were collected until now to the two RAID HDDs.

For DAQ1: /media/sdc2 - last file 229Thor7889.lmd. Please note, that the naming sceme on DAQ1 is inconsistent - please look for the filenumnbering
Entry  Sun Apr 10 13:17:53 2022, Kristian, Max, DAQ, Wavemeter readout 
We changed the Labview code in order to reduce the amount of data that was overfilling the WM data queue. We removed the temperature and intensity values
and added an additional 250ms wait.
ELOG V3.1.5-fc6679b