HOME ESR EXPERIMENTS LABS etc
E121 E127 E132 E125 E143 laser_cooling_2021 E142 Ê128 E146 E0052 E0018 E0028 E0038
  Nuclear hyperfine mixing in 229Th89+, Page 2 of 12  Not logged in ELOG logo
New entries since:Thu Jan 1 01:00:00 1970
ID Date Author Categorydown Subject Text Attachments
  223   Wed Apr 13 14:29:04 2022 Carsten, Rodolfo, Sarper, David, Pierre-Michel, Max, SebastianRunsAfternoon - morning shift runs 13.+14.04.Changed the wavelength scan range from [B]321.62
- 321.10 nm[/B] to [B][B]321.12 - 320.60[/B]
nm[/B] after one scan.
 coolerProblemsAndFragments.JPGfragments.JPGBroaderIonBeam.JPGagainNormalWidth.JPG 
  230   Thu Apr 14 17:47:55 2022 Carsten, Rodolfo, Konstantin, SarperRunsAfternoon shift runs 14.04 changing the wavelength scan rangeThe new scan range is changed from [B]321.12
- 320.60 nm to 320.62 - 320.10 nm[/B]
  
  232   Thu Apr 14 22:44:40 2022 Patrick, Simon, Danyal, MaxRunsRuns night shift 14./15.04.Runs of the night shift. The wavelength range
is 320.62 - 320.10 nm.
  
  236   Sat Apr 16 04:19:05 2022 Patrick, Simon, Alex, SarperRunsRuns night + morning shifts 16.04.Runs of the night shift 16.04. after changing
the wavelength range. The new range is 320.12
- 319.859 nm. The step size was decreased
  
  1   Mon Jul 26 17:15:47 2021 RodolfoLaserQuanta Ray - Water leakOn Thursday afternoon (22.07.2021) a water
connector has broken and water has been spilled
in the laser head. 
 QRAY_AnschlussGeplatzt.jpg 
  2   Mon Jul 26 17:58:52 2021 RodolfoLaserWater FiltersThe water filters (deioninzing and particle)
have been replaced. 
 QRAY_DeionizingFilter.jpg 
  4   Mon Aug 2 10:54:44 2021 RodolfoLaserQuanta Ray - Water leak in the power supplyAfter fixing the broken water connector in
the laser head a new water leak appeared
in the Quanta-Ray power supply. 
 QRAY_WaterPump.jpg 
  11   Thu Mar 3 13:50:45 2022 RodolfoLaserDye cell cracketThe "20 mm" dye cell used for the Oscillator
and Pramplifier got damage. I supposed it
was produced by the pump beam hitting on
 Cobra_Farbstoffzelle_08-02-2022.jpg 
  13   Thu Mar 3 17:34:42 2022 RodolfoLaserPump Laser We have issues with our Pump Laser (Spectra
Physics - Quanta Ray Pro). After the service
made on mid Dezember (2021) the laser does
 QRAY_Simmern.jpg 
  14   Thu Mar 3 21:39:21 2022 RodolfoLaserDye solution, DCM in Ethanol and DMSOI have prepared a new dye solution of DCM
in Ethanol and DMSO (1:1). The concentration
of the "oscillator + preamplifier" is 0,3
 FarbstoffVorrat_Thorium.jpg 
  33   Thu Mar 24 16:59:35 2022 RodolfoLaserPhotodiode signal SOFast Photodiode signal at the "SO-Bogen" (South-east
arc).
 FastPD_SO_Signal_24-03-2022.jpgFastPD_SO_24-03-2022.jpg 
  35   Fri Mar 25 14:31:24 2022 RodolfoLaserPhotodiode signal SO in the MesshütteMeasured at the Panel "Messhütte - Laserhütte",
BNC "Fast PD / SO".
 PD_SO_Signal_MH_25-03-2022.PNG 
  36   Tue Mar 29 12:01:08 2022 RodolfoLaserLaser targetsI have installed two targets* (size: 2 inches)
to monitor the position of the laser beam.
 TargetIN_SO.pngTargetOUT_NO.pngTargetControlBox.jpg 
  38   Tue Mar 29 21:37:13 2022 RodolfoLaserPump LaserThe gold "reflectors and end plates" of our
pump laser (Quanta-Ray PRO) have been exchanged
by new ones.
 QRay_GoldReflektoren_2010-2022.jpg 
  80   Sun Apr 3 13:05:08 2022 CarstenLaserDistance of Partilce detecotrs wr.t. to electron coolerAs of Yuris post in the elog the positions
(Info from H. Weick):
  
  82   Sun Apr 3 18:41:11 2022 Konstantin and RodolfoLaserBeam Position StabilizationWe (Konstantin and I) have optimized the TEM
"laser-beam position stabilization".
 MRC_Konfiguration_MitMainAmp_03-04-2022.pngMRC-ControllerSetting_03-04-2022.jpgFG_SettingsMRC_03-04-2022.jpgTriggerSignals_f_MRC_04-04-2022.PNG 
  101   Tue Apr 5 20:16:13 2022 RodolfoLaserDiscriminator for HF reference signalFor the laser trigger we need to generate
a few ns NIM-puls from the HF reference signal.
Therefore we (Wilfried, Konstantin and me)
 CFD_für_HFSignal.jpgDisc_HFSignal_05-04-2022.PNG 
  102   Tue Apr 5 20:58:53 2022 WilfriedLaserTrigger for LaserElectronic for the timing of the laser puls
optimized. The problem with the broad photodiode
histograms is solved.
 TriggerForLaser_05-04-2022.PNGTriggerLaser_Cabling.jpg 
  118   Thu Apr 7 00:15:11 2022 PatrickLaserLaser timing adjustedWe have adjusted the laser timing using PMT-middle.
Bunching freq. is 3.90362 MHz.
 220407-Laser-ion-timing.png220407-Laser-ion-timing-full-period.png220407-ParDetSO_PMT-middle.png 
  120   Thu Apr 7 01:19:05 2022 PatrickLaserLaser scanningWe did some laser scans. The wavemeter shows
spikes during scans. This may be just while
the frequency changes. See picture.
 2022-04-07_Wavemeter_during_scan.png2022-04-07_Wavemeter_during_scan2.png2022-04-07_Wavemeter_during_scan3.png2022-04-07_Wavemeter_during_scan4.png 
ELOG V3.1.5-fc6679b